Patient position reproducibility in fractionated stereotactically guided conformal radiotherapy using the BrainLab mask system.

H Alheit1, S Dornfeld, M Dawel, M Alheit, B Henzel, K Steckler, H Blank, P Geyer.   

Abstract

PURPOSE: Dedicated mask systems nowadays allow the use of stereotactic radiotherapy in fractionated regimes, therefore combining the advantages of high precision radiotherapy with the biological benefit of fractionation. Therefore the knowledge of institution specific isocenter accuracy is essential for decision-making about margins to be allowed to form the planning target volume. PATIENTS AND
METHOD: Measurements of isocenter deviations during fractionated treatments were performed in 33 patients using the simulator Simulix-xy (Oldelft) in connection with the BrainLab angiographic localizer-box as well as port-films. In both cases repeated images were overlaid by use of anatomical landmarks with a methodical accuracy in the order of 0.5 mm.
RESULTS: Both methods yield random isocenter deviations of less then 2 mm (standard deviation) in all three directions and no significant systematic deviations. These values are in the order of the accuracy of the method, obtained by comparison of two independent investigators, as well as they are comparable with the literature.
CONCLUSIONS: The accuracy of less than 2 mm indicates safety margins of 3-4 mm as sufficient for clinical routine to cover the target in 95.5% of all set-ups (2 SD).

Entities:  

Mesh:

Year:  2001        PMID: 11398613     DOI: 10.1007/pl00002407

Source DB:  PubMed          Journal:  Strahlenther Onkol        ISSN: 0179-7158            Impact factor:   3.621


  15 in total

1.  High symptom improvement and local tumor control using stereotactic radiotherapy when given early after diagnosis of meningioma. A multicentre study.

Authors:  I Compter; K Zaugg; R M A Houben; J T A Dings; G Bosmans; C Buescher; M M H M E Anten; B G Baumert
Journal:  Strahlenther Onkol       Date:  2012-09-09       Impact factor: 3.621

2.  Intrafraction motion during frameless radiosurgery using Varian HyperArcTM and BrainLab ElementsTM immobilization systems.

Authors:  Amish P Shah; Dylan T Meeks; Twyla R Willoughby; Naren Ramakrishna; Christopher J Warner; Cameron W Swanick; Patrick Kelly; Sanford L Meeks
Journal:  J Radiosurg SBRT       Date:  2020

3.  Intensity-modulated fractionated stereotactic radiotherapy with reduced margin for high grade gliomas: dosimetric analysis of sparing optic nerve & chiasm.

Authors:  Hyeon Kang Koh; Chae-Yong Kim; Jung Ho Han; Yu Jung Kim; Hak Jae Kim; Jae Sung Kim; In Ah Kim
Journal:  J Radiosurg SBRT       Date:  2013

4.  Relocatable fixation systems in intracranial stereotactic radiotherapy. Accuracy of serial CT scans and patient acceptance in a randomized design.

Authors:  A Theelen; J Martens; G Bosmans; R Houben; J J Jager; I Rutten; P Lambin; A W Minken; B G Baumert
Journal:  Strahlenther Onkol       Date:  2011-12-24       Impact factor: 3.621

5.  Evaluation of novel 3D-printed and conventional thermoplastic stereotactic high-precision patient fixation masks for radiotherapy.

Authors:  Veronika M Miron; Tanja Etzelstorfer; Raimund Kleiser; Tobias Raffelsberger; Zoltan Major; Hans Geinitz
Journal:  Strahlenther Onkol       Date:  2022-06-13       Impact factor: 4.033

6.  Comparison of stereotactic radiosurgery and fractionated stereotactic radiotherapy of acoustic neurinomas according to 3-D tumor volume shrinkage and quality of life.

Authors:  Martin Henzel; Klaus Hamm; Helmut Sitter; Markus W Gross; Gunnar Surber; Gabriele Kleinert; Rita Engenhart-Cabillic
Journal:  Strahlenther Onkol       Date:  2009-09-12       Impact factor: 3.621

7.  Fractionated stereotactic radiotherapy for skull base tumors: analysis of treatment accuracy using a stereotactic mask fixation system.

Authors:  Giuseppe Minniti; Maurizio Valeriani; Enrico Clarke; Marco D'Arienzo; Michelangelo Ciotti; Roberto Montagnoli; Francesca Saporetti; Riccardo Maurizi Enrici
Journal:  Radiat Oncol       Date:  2010-01-13       Impact factor: 3.481

8.  Efficacy of Stereotactic Conformal Radiotherapy vs Conventional Radiotherapy on Benign and Low-Grade Brain Tumors: A Randomized Clinical Trial.

Authors:  Rakesh Jalali; Tejpal Gupta; Jayant S Goda; Savita Goswami; Nalini Shah; Debnarayan Dutta; Uday Krishna; Jayita Deodhar; Padmavati Menon; Sadhna Kannan; Rajiv Sarin
Journal:  JAMA Oncol       Date:  2017-10-01       Impact factor: 31.777

9.  Quality assurance and commissioning of an infrared marker-based patient positioning system for frameless extracranial stereotactic radiotherapy.

Authors:  Tejpal Gupta; Reena Phurailatpam; Mishra Ajay; Pai Rajeshri; Mohindra Pranshu; Chopra Supriya
Journal:  Int J Biomed Sci       Date:  2007-12

10.  Potential identity of multi-potential cancer stem-like subpopulation after radiation of cultured brain glioma.

Authors:  Mi K Kang; Beong I Hur; Mi H Ko; Cheul H Kim; Seung H Cha; Soo K Kang
Journal:  BMC Neurosci       Date:  2008-01-30       Impact factor: 3.288

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.