Literature DB >> 12694848

Isocenter accuracy in frameless stereotactic radiotherapy using implanted fiducials.

Ki-Hwan Kim1, Moon-June Cho, Jun-Sang Kim, Jae-Sung Kim, Chang-Joon Song, Shi-Hun Song, Seon-Hwan Kim, Lee Myers, Yong-Eun Kim.   

Abstract

PURPOSE: The stereotactic radiotherapy (SRT) system verifies isocenter accuracy in patient space. In this study, we evaluate isocenter accuracy in frameless SRT using implanted cranial gold markers.
MATERIALS AND METHODS: We performed frameless SRT on 43 intracranial tumor patients between August 1997 and December 2000. The treatment technique was determined by the tumor shape and volume, and by the location of critical organs. The coordinates of anterior-posterior and lateral port film were inputted to ISOLOC software, which calculated (1) the couch moves translation distance required to bring the target point to the isocenter, and (2) the intermarker distance comparisons between the CT study and the treatment machine films. We evaluated the isocenter deviation based on the error between orthogonal film target coordinates and isocenter coordinates.
RESULTS: The mean treatment isocenter deviations (x, y, z) were -0.03, 0.14, and -0.04 mm, respectively. The systematic component isocenter standard deviations were 0.28, 0.31, and 0.35 mm (1 SD), respectively, and the random component isocenter standard deviations were 0.53, 0.52, and 0.50 mm (1 SD), respectively.
CONCLUSIONS: The isocenter accuracy in the frameless SRT-implanted fiducial system is highly reliable and is comparable to that of other stereotactic radiosurgery systems.

Entities:  

Mesh:

Substances:

Year:  2003        PMID: 12694848     DOI: 10.1016/s0360-3016(03)00088-9

Source DB:  PubMed          Journal:  Int J Radiat Oncol Biol Phys        ISSN: 0360-3016            Impact factor:   7.038


  6 in total

1.  Frameless linac-based stereotactic radiosurgery (SRS) for brain metastases: analysis of patient repositioning using a mask fixation system and clinical outcomes.

Authors:  Giuseppe Minniti; Claudia Scaringi; Enrico Clarke; Maurizio Valeriani; Mattia Osti; Riccardo Maurizi Enrici
Journal:  Radiat Oncol       Date:  2011-11-16       Impact factor: 3.481

2.  Stereotactic Irradiation of GH-Secreting Pituitary Adenomas.

Authors:  G Minniti; C Scaringi; D Amelio; R Maurizi Enrici
Journal:  Int J Endocrinol       Date:  2012-02-22       Impact factor: 3.257

3.  Analysis of inter- and intrafraction accuracy of a commercial thermoplastic mask system used for image-guided particle radiation therapy.

Authors:  Dante Amelio; Marcus Winter; Daniel Habermehl; Oliver Jäkel; Jurgen Debus; Stephanie E Combs
Journal:  J Radiat Res       Date:  2013-07       Impact factor: 2.724

4.  Fractionated stereotactic radiotherapy for skull base tumors: analysis of treatment accuracy using a stereotactic mask fixation system.

Authors:  Giuseppe Minniti; Maurizio Valeriani; Enrico Clarke; Marco D'Arienzo; Michelangelo Ciotti; Roberto Montagnoli; Francesca Saporetti; Riccardo Maurizi Enrici
Journal:  Radiat Oncol       Date:  2010-01-13       Impact factor: 3.481

5.  Intracranial application of IMRT based radiosurgery to treat multiple or large irregular lesions and verification of infra-red frameless localization system.

Authors:  Joshua D Lawson; Jia-Zhu Wang; Sameer K Nath; Roger Rice; Todd Pawlicki; Arno J Mundt; Kevin Murphy
Journal:  J Neurooncol       Date:  2009-08-20       Impact factor: 4.130

6.  Simulation of intrafraction motion and overall geometrical accuracy of a frameless intracranial radiosurgery process.

Authors:  Vladimir Feygelman; Luke Walker; Prakash Chinnaiyan; Kenneth Forster
Journal:  J Appl Clin Med Phys       Date:  2008-10-24       Impact factor: 2.102

  6 in total