PURPOSE: To assess the accuracy of patient repositioning and clinical outcomes of frameless stereotactic radiosurgery (SRS) for brain metastases using a stereotactic mask fixation system. PATIENTS AND METHODS: One hundred two patients treated consecutively with frameless SRS as primary treatment at University of Rome Sapienza Sant'Andrea Hospital between October 2008 and April 2010 and followed prospectively were involved in the study. A commercial stereotactic mask fixation system (BrainLab) was used for patient immobilization. A computerized tomography (CT) scan obtained immediately before SRS was used to evaluate the accuracy of patient repositioning in the mask by comparing the isocenter position to the isocenter position established in the planning CT. Deviations of isocenter coordinates in each direction and 3D displacement were calculated. Overall survival, brain control, and local control were estimated using the Kaplan-Meier method calculated from the time of SRS.Entities:
Mesh:
Year: 2011 PMID: 22085700 PMCID: PMC3253058 DOI: 10.1186/1748-717X-6-158
Source DB: PubMed Journal: Radiat Oncol ISSN: 1748-717X Impact factor: 3.481
Summary of tumor charaterictiscs and treatment parameters of patients treated with radiosurgery
| No (%) | |
|---|---|
| 102 | |
| 64 | |
| 52/50 | |
| 1 lesion | 64 (63%) |
| 2 lesions | 24 (23%) |
| 3 lesions | 14 (14%) |
| lung | 54 (53.5%) |
| breast | 17 (16.5%) |
| melanoma | 14 (13.5%) |
| others | 17(16.5%) |
| frontal | 31 (20%) |
| parietal | 37 (24%) |
| temporal | 30 (19%) |
| cerebellar | 23 (15%) |
| occipital | 26 (17%) |
| brainstem | 7 (5%) |
| 20 Gy | 86 (56%) |
| 18 Gy | 44 (28%) |
| 15-16 | 24 (16%) |
| median | 1.6 |
| range | 0.03-16.3 |
| median | 2.2 |
| range | 0.2-18.8 |
Accuracy of isocenter relocation at CT verification
| Direction (mm) | Mean | SD | Range |
|---|---|---|---|
| Cranio-caudal | 0.4 | 0.6 | -1,2 - 2.1 |
| Medio-lateral | 0.12 | 0.35 | -0.9 - 1.4 |
| Anterior-posterior | 0.2 | 0.4 | -1,6 - 1.3 |
| 3D-displacement | 0.5 | 0.7 | 0.1-2.9 |
SD, standard deviation
Mean and standard deviation of isocenter displacement between CT verification and post-treatment CT
| Direction (mm) | Mean | SD | Range |
|---|---|---|---|
| Cranio-caudal | 0.08 | 0.2 | -0.3 - 0.2 |
| Medio-lateral | 0.04 | 0.14 | -0.3 - 0.2 |
| Anterior-posterior | 0.06 | 0.15 | -0.5 - 0.4 |
| 3D-displacement | 0.09 | 0.28 | 0-0.6 |
SD, standard deviation
Figure 1Kaplan-Meier analysis of overall survival, brain control, and local control.