CT verification of isocentre relocatability using stereotactic mask fixation system.

V Y W Wong1, S Y Tung, T W Leung, K H S Ho.   

Abstract

AIMS: This study introduces a non-invasive method based on computed tomography (CT) verification to ensure patients are accurately positioned before fractionated stereotactic radiotherapy. It enables quality control of mask positioning with reference to the CT images of the treatment plan.
MATERIALS AND METHODS: A mask system, together with a dental impression moulded mouth bite, was used for patient immobilisation. In order to facilitate relevant image comparison, special alignment during CT localisation was discussed in the study. The accuracy of patient set-up was studied by assessing the isocentre position in relation to the patient's anatomical structure. The planning CT images were applied as a reference and the study was applied to 261 cranial applications.
RESULTS: The results show that the mean and the maximum overall displacements at the isocentre were 0.7 and 2.5 mm, respectively. The mean and the maximum rotational displacement in the axial plane were 0.56 degrees and 2 degrees, respectively. The mean translational displacement and rotational displacement were close to zero when considering the direction of movement.
CONCLUSIONS: The results indicate that the systematic error of the mask system and the verification method are minimal. Advantages of this technique include the simple set-up, three-dimensional quantification and short study time (10-15 min). It is therefore practical to implement on a routine basis. Investigation of the ability to relocate the mask is also recommended to justify the required safety margin between the clinical and planning target volumes.

Entities:  

Mesh:

Year:  2003        PMID: 12924459     DOI: 10.1016/s0936-6555(03)00091-8

Source DB:  PubMed          Journal:  Clin Oncol (R Coll Radiol)        ISSN: 0936-6555            Impact factor:   4.126


  5 in total

1.  Frameless linac-based stereotactic radiosurgery (SRS) for brain metastases: analysis of patient repositioning using a mask fixation system and clinical outcomes.

Authors:  Giuseppe Minniti; Claudia Scaringi; Enrico Clarke; Maurizio Valeriani; Mattia Osti; Riccardo Maurizi Enrici
Journal:  Radiat Oncol       Date:  2011-11-16       Impact factor: 3.481

2.  Analysis of patient repositioning accuracy in precision radiation therapy using automated image fusion.

Authors:  James L Robar; Brenda G Clark; Jason W Schella; Chang Seon Kim
Journal:  J Appl Clin Med Phys       Date:  2005-01-12       Impact factor: 2.102

3.  CBCT Verification of SRT for Patients With Brain Metastases.

Authors:  Judit Papp; Mihály Simon; Emese Csiki; Árpád Kovács
Journal:  Front Oncol       Date:  2022-01-19       Impact factor: 6.244

4.  Fractionated stereotactic radiotherapy for skull base tumors: analysis of treatment accuracy using a stereotactic mask fixation system.

Authors:  Giuseppe Minniti; Maurizio Valeriani; Enrico Clarke; Marco D'Arienzo; Michelangelo Ciotti; Roberto Montagnoli; Francesca Saporetti; Riccardo Maurizi Enrici
Journal:  Radiat Oncol       Date:  2010-01-13       Impact factor: 3.481

5.  Intra-fraction and Inter-fraction analysis of a dedicated immobilization device for intracranial radiation treatment.

Authors:  Chin Loon Ong; Niccolò Giaj-Levra; Luca Nicosia; Vanessa Figlia; Davide Tomasini; Eric M Franken; Filippo Alongi
Journal:  Radiat Oncol       Date:  2020-08-18       Impact factor: 3.481

  5 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.