Imaging of weak-source distributions in LSO-based small-animal PET scanners.

Andrew L Goertzen1, Joon Young Suk, Christopher J Thompson.   

Abstract

UNLABELLED: Lutetium oxyorthosilicate (LSO)- or lutetium-yttrium oxyorthosilicate (LYSO)-based PET scanners have intrinsic radioactivity in the scintillator crystals due to the presence of (176)Lu, which decays by beta-emission followed by one or more prompt gamma-ray emissions. This leads to intrinsic true counts that can influence the image when scanning low levels of activity. An evaluation of the effects of this intrinsic activity for low levels of activity and different energy windows is performed on an LSO-based small-animal PET scanner.
METHODS: Intrinsic count rate and sensitivity were measured for a range of lower-level discriminators (LLDs) ranging from 100 to 750 keV. The noise equivalent count rate (NECR) as a function of LLD for activity levels from 100 Bq to 100 kBq was estimated using a combination of measurement and previously published data for this scanner. Phantom imaging was performed using three (68)Ge sources of strength 55, 220, and 940 Bq and LLD levels of 250, 350, and 400 keV. The images were assessed using a contrast-to-noise ratio (CNR) analysis and by comparing the observed ratio of source activities to the true ratio value.
RESULTS: The intrinsic true count rate is reduced from 940 counts per second (cps) for a 250- to 750-keV energy window to <2 cps for a 400- to 750-keV window. There is a corresponding 2-fold drop in sensitivity for detected true events for external positron sources for these 2 energy windows. The NECR versus LLD curves showed a highly peaked shape, with the optimum LLD being approximately 425 keV. The phantom image results were dominated by the intrinsic true counts when an energy window of 250-750 keV was used. The intrinsic true counts were almost completely removed by raising the LLD to 400 keV. The CNR for each of the sources was higher for the narrow energy window and the 55 Bq could be easily visualized in images acquired with LLD levels of 350 and 400 keV but not when the 250-keV LLD was used. Images acquired with an LLD of 400 keV and reconstructed with 2-dimensional filtered backprojection were the most quantitatively accurate.
CONCLUSION: It is possible to visualize sources of <1 kBq in LSO-based animal PET systems by raising the LLD to 400 keV to exclude the majority of the counts due to the intrinsic activity present in the LSO.

Entities:  

Mesh:

Substances:

Year:  2007        PMID: 17873140     DOI: 10.2967/jnumed.107.040584

Source DB:  PubMed          Journal:  J Nucl Med        ISSN: 0161-5505            Impact factor:   10.057


  17 in total

1.  System design of a small OpenPET prototype with 4-layer DOI detectors.

Authors:  Eiji Yoshida; Shoko Kinouchi; Hideaki Tashima; Fumihiko Nishikido; Naoko Inadama; Hideo Murayama; Taiga Yamaya
Journal:  Radiol Phys Technol       Date:  2011-11-29

2.  Estimation of the minimum detectable activity of preclinical PET imaging systems with an analytical method.

Authors:  Qinan Bao; Arion F Chatziioannou
Journal:  Med Phys       Date:  2010-11       Impact factor: 4.071

3.  Reduction method for intrinsic random coincidence events from (176)Lu in low activity PET imaging.

Authors:  Eiji Yoshida; Hideaki Tashima; Fumihiko Nishikido; Hideo Murayama; Taiga Yamaya
Journal:  Radiol Phys Technol       Date:  2014-02-05

4.  Performance evaluation of PETbox: a low cost bench top preclinical PET scanner.

Authors:  Hui Zhang; Qinan Bao; Nam T Vu; Robert W Silverman; Richard Taschereau; Brittany N Berry-Pusey; Ali Douraghy; Fernando R Rannou; David B Stout; Arion F Chatziioannou
Journal:  Mol Imaging Biol       Date:  2011-10       Impact factor: 3.488

5.  A method for measuring the energy spectrum of coincidence events in positron emission tomography.

Authors:  Andrew L Goertzen; David B Stout; Christopher J Thompson
Journal:  Phys Med Biol       Date:  2009-12-21       Impact factor: 3.609

6.  Performance Characteristics of BGO Detectors for a Low Cost Preclinical PET Scanner.

Authors:  H Zhang; N T Vu; Q Bao; R W Silverman; B N Berry-Pusey; A Douraghy; D A Williams; F R Rannou; D B Stout; A F Chatziioannou
Journal:  IEEE Trans Nucl Sci       Date:  2010-06-01       Impact factor: 1.679

7.  Performance and limitations of positron emission tomography (PET) scanners for imaging very low activity sources.

Authors:  Melissa I Freedenberg; Ramsey D Badawi; Alice F Tarantal; Simon R Cherry
Journal:  Phys Med       Date:  2013-05-14       Impact factor: 2.685

8.  Performance evaluation of the inveon dedicated PET preclinical tomograph based on the NEMA NU-4 standards.

Authors:  Qinan Bao; Danny Newport; Mu Chen; David B Stout; Arion F Chatziioannou
Journal:  J Nucl Med       Date:  2009-02-17       Impact factor: 10.057

9.  Performance characteristics of a positron projection imager for mouse whole-body imaging.

Authors:  Jurgen Seidel; Wenze Xi; John W Kakareka; Thomas J Pohida; Elaine M Jagoda; Michael V Green; Peter L Choyke
Journal:  Nucl Med Biol       Date:  2013-02-09       Impact factor: 2.408

10.  NEMA NU 2-2007 performance measurements of the Siemens Inveon preclinical small animal PET system.

Authors:  Brad J Kemp; Carrie B Hruska; Aaron R McFarland; Mark W Lenox; Val J Lowe
Journal:  Phys Med Biol       Date:  2009-03-25       Impact factor: 3.609

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.