A cone-shaped phantom for assessment of small animal PET scatter fraction and count rate performance.

Rameshwar Prasad1, Habib Zaidi.   

Abstract

PURPOSE: Positron emission tomography (PET) image quality deteriorates as the object size increases owing to increased detection of scattered and random events. The characterization of the scatter component in small animal PET imaging has received little attention owing to the small scatter fraction (SF) when imaging rodents. The purpose of this study is first to design and fabricate a cone-shaped phantom which can be used for measurement of object size-dependent SF and noise equivalent count rates (NECR), and second, to assess these parameters for two small animal PET scanners as function of radial offset, object size and lower energy threshold (LET).
METHODS: The X-PET™ and LabPET-8™ scanners were modeled as realistically as possible using GATE Monte Carlo simulation platform. The simulation models were validated against experimental measurements in terms of sensitivity, SF and NECR. The dedicated phantom was fabricated in-house using high-density polyethylene. The optimized dimensions of the cone-shaped phantom are 158 mm (length), 20 mm (minimum diameter), 70 mm (maximum diameter) and taper angle of 9°.
RESULTS: The relative difference between simulated and experimental results for the LabPET-8™ scanner varied between 0.7% and 10% except for a few results where it was below 16%. Depending on the radial offset from the center of the central axial field-of-view (3-6 cm diameter), the SF for the cone-shaped phantom varied from 26.3% to 18.2%, 18.6 to 13.1% and 10.1 to 7.6% for the X-PET™, whereas it varied from 34.4% to 26.9%, 19.1 to 17.0% and 9.1 to 7.3% for the LabPET-8™, for LETs of 250, 350 and 425 keV, respectively. The SF increases as the radial offset decreases, LET decreases and object size increases. The SF is higher for the LabPET-8™ compared with the X-PET™ scanner. The NECR increases as the radial offset increases and object size decreases. The maximum NECR was obtained at a LET of 350 keV for the LabPET-8™ and 250 keV for the X-PET™. High correlation coefficients for SF and NECR were observed between the cone-shaped phantom and an equivalent volume cylindrical phantom for the three considered axial fields of view.
CONCLUSIONS: A single cone-shaped phantom enables the assessment of the impact of three factors, namely radial offset, LET and object size on PET SF and count rate estimates. This phantom is more realistic owing to the non-uniform shape of rodents' bodies compared to cylindrical uniform phantoms and seems to be well suited for evaluation of object size-dependent SF and NECR.

Entities:  

Mesh:

Year:  2012        PMID: 22311541      PMCID: PMC3443338          DOI: 10.1007/s11307-012-0546-2

Source DB:  PubMed          Journal:  Mol Imaging Biol        ISSN: 1536-1632            Impact factor:   3.488


  16 in total

1.  Feasibility of a high-speed gamma-camera design using the high-yield-pileup-event-recovery method.

Authors:  W H Wong; H Li; J Uribe; H Baghaei; Y Wang; S Yokoyama
Journal:  J Nucl Med       Date:  2001-04       Impact factor: 10.057

Review 2.  Scatter modelling and compensation in emission tomography.

Authors:  Habib Zaidi; Kenneth F Koral
Journal:  Eur J Nucl Med Mol Imaging       Date:  2004-03-31       Impact factor: 9.236

3.  NEMA NU-04-based performance characteristics of the LabPET-8™ small animal PET scanner.

Authors:  Rameshwar Prasad; Osman Ratib; Habib Zaidi
Journal:  Phys Med Biol       Date:  2011-09-23       Impact factor: 3.609

4.  Performance evaluation of the microPET focus: a third-generation microPET scanner dedicated to animal imaging.

Authors:  Yuan-Chuan Tai; Ananya Ruangma; Douglas Rowland; Stefan Siegel; Danny F Newport; Patrick L Chow; Richard Laforest
Journal:  J Nucl Med       Date:  2005-03       Impact factor: 10.057

5.  Performance measurement of the microPET focus 120 scanner.

Authors:  Jin Su Kim; Jae Sung Lee; Ki Chun Im; Su Jin Kim; Seog-Young Kim; Dong Soo Lee; Dae Hyuk Moon
Journal:  J Nucl Med       Date:  2007-08-17       Impact factor: 10.057

6.  Treatment of axial data in three-dimensional PET.

Authors:  M E Daube-Witherspoon; G Muehllehner
Journal:  J Nucl Med       Date:  1987-11       Impact factor: 10.057

7.  Validation of the octamouse for simultaneous 18F-fallypride small-animal PET recordings from 8 mice.

Authors:  Axel Rominger; Erik Mille; Shanshan Zhang; Guido Böning; Stefan Förster; Sebastian Nowak; Franz Josef Gildehaus; Björn Wängler; Peter Bartenstein; Paul Cumming
Journal:  J Nucl Med       Date:  2010-09-16       Impact factor: 10.057

8.  Performance evaluation of the GE healthcare eXplore VISTA dual-ring small-animal PET scanner.

Authors:  Yuchuan Wang; Jurgen Seidel; Benjamin M W Tsui; Juan J Vaquero; Martin G Pomper
Journal:  J Nucl Med       Date:  2006-11       Impact factor: 10.057

9.  Development of a 4-D digital mouse phantom for molecular imaging research.

Authors:  William P Segars; Benjamin M W Tsui; Eric C Frey; G Allan Johnson; Stuart S Berr
Journal:  Mol Imaging Biol       Date:  2004 May-Jun       Impact factor: 3.488

10.  Performance evaluation of the inveon dedicated PET preclinical tomograph based on the NEMA NU-4 standards.

Authors:  Qinan Bao; Danny Newport; Mu Chen; David B Stout; Arion F Chatziioannou
Journal:  J Nucl Med       Date:  2009-02-17       Impact factor: 10.057

View more
  2 in total

1.  Scatter characterization and correction for simultaneous multiple small-animal PET imaging.

Authors:  Rameshwar Prasad; Habib Zaidi
Journal:  Mol Imaging Biol       Date:  2014-04       Impact factor: 3.488

2.  Guide to Plant-PET Imaging Using 11CO2.

Authors:  Jens Mincke; Jan Courtyn; Christian Vanhove; Stefaan Vandenberghe; Kathy Steppe
Journal:  Front Plant Sci       Date:  2021-06-02       Impact factor: 5.753

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.