Estimation of the minimum detectable activity of preclinical PET imaging systems with an analytical method.

Qinan Bao1, Arion F Chatziioannou.   

Abstract

PURPOSE: The traditional figures of merit used in the evaluation of positron emission tomography (PET) systems, including system sensitivity and spatial resolution, do not directly reflect the minimum detectable activity (MDA) performance, despite the fact that it is one of the most important tasks for a PET system. MDA, as a combination of the more traditional PET system parameters, is directly related to lesion detection. However, MDA evaluation is task specific and cannot be done by a single measurement. Therefore, a simple method to evaluate system detectability needs to be developed.
METHODS: In this work, an analytical method of MDA estimation was developed, taking into account system sensitivity, spatial resolution, source properties, and noise propagation in image reconstruction by using the Rose criterion and/or the Curie equation as the detection standard. In the implementation, the source background, as well as the intrinsic activity background from the scintillation material of the system, was also taken into consideration. The accuracy of this method was evaluated in two commercially available preclinical PET systems, with phantom experiments that were designed to closely mimic in vivo tumor uptake without introducing finite boundaries between the source and the background.
RESULTS: The lesion contrast-to-noise ratio calculated by the analytical evaluation showed good agreement with that obtained from the experiments. Visual assessment of the reconstructed images at the detection limit (based on analytical evaluation) also was in agreement with the Rose criterion. The MDA performance was quantitatively compared between the two preclinical PET systems and showed different detection limits under different imaging conditions, suggesting that the detection limit of a PET system strongly depends on the lesion properties and acquisition settings.
CONCLUSIONS: An analytical method of evaluating the PET system detectability was developed and validated by experiments. Overall, the analytical MDA calculation provides a simple way to evaluate the signal detectability of a PET system and can be used for comparing different systems. It also provides guidelines for designing new PET tomographs as well as optimizing data acquisition protocols.

Entities:  

Mesh:

Substances:

Year:  2010        PMID: 21158319      PMCID: PMC2994933          DOI: 10.1118/1.3495817

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  27 in total

1.  Comparison of 3-D maximum a posteriori and filtered backprojection algorithms for high-resolution animal imaging with microPET.

Authors:  A Chatziioannou; J Qi; A Moore; A Annala; K Nguyen; R Leahy; S R Cherry
Journal:  IEEE Trans Med Imaging       Date:  2000-05       Impact factor: 10.048

Review 2.  Molecular imaging of small animals with dedicated PET tomographs.

Authors:  Arion F Chatziioannou
Journal:  Eur J Nucl Med Mol Imaging       Date:  2001-11-22       Impact factor: 9.236

Review 3.  Use of positron emission tomography in animal research.

Authors:  S R Cherry; S S Gambhir
Journal:  ILAR J       Date:  2001

4.  A non-parametric bootstrap approach for analysing the statistical properties of SPECT and PET images.

Authors:  Irène Buvat
Journal:  Phys Med Biol       Date:  2002-05-21       Impact factor: 3.609

5.  NEMA NU 2 performance tests for scanners with intrinsic radioactivity.

Authors:  Charles C Watson; Michael E Casey; Lars Eriksson; Tim Mulnix; Doug Adams; Bernard Bendriem
Journal:  J Nucl Med       Date:  2004-05       Impact factor: 10.057

6.  Performance evaluation of the microPET focus: a third-generation microPET scanner dedicated to animal imaging.

Authors:  Yuan-Chuan Tai; Ananya Ruangma; Douglas Rowland; Stefan Siegel; Danny F Newport; Patrick L Chow; Richard Laforest
Journal:  J Nucl Med       Date:  2005-03       Impact factor: 10.057

7.  Attenuation correction for small animal PET tomographs.

Authors:  Patrick L Chow; Fernando R Rannou; Arion F Chatziioannou
Journal:  Phys Med Biol       Date:  2005-04-06       Impact factor: 3.609

8.  Radioactive spheres without inactive wall for lesion simulation in PET.

Authors:  Marisa Bazañez-Borgert; Ralph A Bundschuh; Michael Herz; Maria-Jose Martínez; Markus Schwaiger; Sibylle I Ziegler
Journal:  Z Med Phys       Date:  2008       Impact factor: 4.820

9.  Imaging of weak-source distributions in LSO-based small-animal PET scanners.

Authors:  Andrew L Goertzen; Joon Young Suk; Christopher J Thompson
Journal:  J Nucl Med       Date:  2007-09-14       Impact factor: 10.057

10.  Performance evaluation of the GE healthcare eXplore VISTA dual-ring small-animal PET scanner.

Authors:  Yuchuan Wang; Jurgen Seidel; Benjamin M W Tsui; Juan J Vaquero; Martin G Pomper
Journal:  J Nucl Med       Date:  2006-11       Impact factor: 10.057

View more
  17 in total

1.  Multifunctional-layered materials for creating membrane-restricted nanodomains and nanoscale imaging.

Authors:  P Srinivasan
Journal:  Appl Phys Lett       Date:  2016-01-21       Impact factor: 3.791

2.  MR simulation for prostate radiation therapy: effect of coil mounting position on image quality.

Authors:  J Sun; P Pichler; J Dowling; F Menk; P Stanwell; J Arm; P B Greer
Journal:  Br J Radiol       Date:  2014-07-25       Impact factor: 3.039

3.  Sodium-glucose transporter 2 is a diagnostic and therapeutic target for early-stage lung adenocarcinoma.

Authors:  Claudio R Scafoglio; Brendon Villegas; Gihad Abdelhady; Sean T Bailey; Jie Liu; Aditya S Shirali; W Dean Wallace; Clara E Magyar; Tristan R Grogan; David Elashoff; Tonya Walser; Jane Yanagawa; Denise R Aberle; Jorge R Barrio; Steven M Dubinett; David B Shackelford
Journal:  Sci Transl Med       Date:  2018-11-14       Impact factor: 17.956

4.  Evaluation of the Genisys4, a bench-top preclinical PET scanner.

Authors:  Ken Herrmann; Magnus Dahlbom; David Nathanson; Liu Wei; Caius Radu; Arion Chatziioannou; Johannes Czernin
Journal:  J Nucl Med       Date:  2013-04-29       Impact factor: 10.057

5.  Single-cell tracking with PET using a novel trajectory reconstruction algorithm.

Authors:  Keum Sil Lee; Tae Jin Kim; Guillem Pratx
Journal:  IEEE Trans Med Imaging       Date:  2014-11-21       Impact factor: 10.048

6.  A NIM PET/CT phantom for evaluating the PET image quality of micro-lesions and the performance parameters of CT.

Authors:  Shujie Lu; Peng Zhang; Chengwei Li; Jie Sun; Wenli Liu; Pu Zhang
Journal:  BMC Med Imaging       Date:  2021-11-08       Impact factor: 1.930

7.  Cerenkov luminescence and PET imaging of 90Y: capabilities and limitations in small animal applications.

Authors:  Gregory S Mitchell; P N Thomas Lloyd; Simon R Cherry
Journal:  Phys Med Biol       Date:  2020-03-20       Impact factor: 3.609

8.  NEMA NU-4 performance evaluation of PETbox4, a high sensitivity dedicated PET preclinical tomograph.

Authors:  Z Gu; R Taschereau; N T Vu; H Wang; D L Prout; R W Silverman; B Bai; D B Stout; M E Phelps; A F Chatziioannou
Journal:  Phys Med Biol       Date:  2013-05-10       Impact factor: 3.609

9.  Deep learning-based attenuation correction in the absence of structural information for whole-body positron emission tomography imaging.

Authors:  Xue Dong; Yang Lei; Tonghe Wang; Kristin Higgins; Tian Liu; Walter J Curran; Hui Mao; Jonathon A Nye; Xiaofeng Yang
Journal:  Phys Med Biol       Date:  2020-03-02       Impact factor: 3.609

10.  Assessment of acquisition protocols for routine imaging of Y-90 using PET/CT.

Authors:  Thomas Carlier; Thomas Eugène; Caroline Bodet-Milin; Etienne Garin; Catherine Ansquer; Caroline Rousseau; Ludovic Ferrer; Jacques Barbet; Frédéric Schoenahl; Françoise Kraeber-Bodéré
Journal:  EJNMMI Res       Date:  2013-02-17       Impact factor: 3.138

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.