Performance characterization of the Inveon preclinical small-animal PET/SPECT/CT system for multimodality imaging.

Keiichi Magota1, Naoki Kubo, Yuji Kuge, Ken-Ichi Nishijima, Songji Zhao, Nagara Tamaki.   

Abstract

PURPOSE: We investigated the performance of the Inveon small-animal PET/SPECT/CT system and compared the imaging capabilities of the SPECT and PET components.
METHODS: For SPECT, the energy resolution, tomographic spatial resolution and system sensitivity were evaluated with a (99m)Tc solution using a single pinhole collimator. For PET, the spatial resolution, absolute sensitivity, scatter fraction and peak noise equivalent count were evaluated. Phantoms and a normal rat were scanned to compare the imaging capabilities of SPECT and PET.
RESULTS: The SPECT spatial resolution was 0.84 mm full-width at half-maximum (FWHM) at a radius of rotation of 25 mm using a 0.5-mm pinhole aperture collimator, while the PET spatial resolution was 1.63 mm FWHM at the centre. The SPECT system sensitivity at a radius of rotation of 25 mm was 35.3 cps/MBq (4 × 10(-3)%) using the 0.5-mm pinhole aperture, while the PET absolute sensitivity was 3.2% for 350-650 keV and 3.432 ns. Accordingly, the volume sensitivity of PET was three orders of magnitude higher than that of SPECT.
CONCLUSION: This integrated PET/SPECT/CT system showed high performance with excellent spatial resolution for SPECT and sensitivity for PET. Based on the tracer availability and system performance, SPECT and PET have complementary roles in multimodality small-animal imaging.

Entities:  

Mesh:

Year:  2010        PMID: 21153410     DOI: 10.1007/s00259-010-1683-y

Source DB:  PubMed          Journal:  Eur J Nucl Med Mol Imaging        ISSN: 1619-7070            Impact factor:   9.236


  21 in total

1.  NEMA NU 2 performance tests for scanners with intrinsic radioactivity.

Authors:  Charles C Watson; Michael E Casey; Lars Eriksson; Tim Mulnix; Doug Adams; Bernard Bendriem
Journal:  J Nucl Med       Date:  2004-05       Impact factor: 10.057

2.  Image-quality assessment for several positron emitters using the NEMA NU 4-2008 standards in the Siemens Inveon small-animal PET scanner.

Authors:  Jonathan A Disselhorst; Maarten Brom; Peter Laverman; Cornelius H Slump; Otto C Boerman; Wim J G Oyen; Martin Gotthardt; Eric P Visser
Journal:  J Nucl Med       Date:  2010-03-17       Impact factor: 10.057

3.  Performance evaluation of the microPET focus: a third-generation microPET scanner dedicated to animal imaging.

Authors:  Yuan-Chuan Tai; Ananya Ruangma; Douglas Rowland; Stefan Siegel; Danny F Newport; Patrick L Chow; Richard Laforest
Journal:  J Nucl Med       Date:  2005-03       Impact factor: 10.057

4.  Evaluating performance of a pixel array semiconductor SPECT system for small animal imaging.

Authors:  Naoki Kubo; Songji Zhao; Yutaka Fujiki; Akiyoshi Kinda; Nobutoku Motomura; Chietsugu Katoh; Tohru Shiga; Hidekazu Kawashima; Yuji Kuge; Nagara Tamaki
Journal:  Ann Nucl Med       Date:  2005-10       Impact factor: 2.668

5.  Physical performance evaluation of the Toshiba GCA-9300A triple-headed system.

Authors:  K Kouris; G A Clarke; P H Jarritt; C E Townsend; S N Thomas
Journal:  J Nucl Med       Date:  1993-10       Impact factor: 10.057

6.  Comparison of 99mTc-annexin A5 with 18F-FDG for the detection of atherosclerosis in ApoE-/- mice.

Authors:  Yan Zhao; Yuji Kuge; Songji Zhao; Koichi Morita; Masayuki Inubushi; H William Strauss; Francis G Blankenberg; Nagara Tamaki
Journal:  Eur J Nucl Med Mol Imaging       Date:  2007-04-17       Impact factor: 9.236

7.  Positron-emission tomography reporter gene expression imaging in rat myocardium.

Authors:  Masayuki Inubushi; Joseph C Wu; Sanjiv S Gambhir; Gobalakrishnan Sundaresan; Nagichettiar Satyamurthy; Mohammad Namavari; Simon Yee; Jorge R Barrio; David Stout; Arion F Chatziioannou; Lily Wu; Heinrich R Schelbert
Journal:  Circulation       Date:  2003-01-21       Impact factor: 29.690

8.  Performance evaluation of an Inveon PET preclinical scanner.

Authors:  Cristian C Constantinescu; Jogeshwar Mukherjee
Journal:  Phys Med Biol       Date:  2009-04-21       Impact factor: 3.609

9.  Quantitative analysis of acute myocardial infarct in rat hearts with ischemia-reperfusion using a high-resolution stationary SPECT system.

Authors:  Zhonglin Liu; George A Kastis; Gail D Stevenson; Harrison H Barrett; Lars R Furenlid; Matthew A Kupinski; Dennis D Patton; Donald W Wilson
Journal:  J Nucl Med       Date:  2002-07       Impact factor: 10.057

10.  Performance evaluation of the inveon dedicated PET preclinical tomograph based on the NEMA NU-4 standards.

Authors:  Qinan Bao; Danny Newport; Mu Chen; David B Stout; Arion F Chatziioannou
Journal:  J Nucl Med       Date:  2009-02-17       Impact factor: 10.057

View more
  34 in total

Review 1.  A shift from cell cultures to creatures: in vivo imaging of small animals in experimental regenerative medicine.

Authors:  Anna J Studwell; Darrell N Kotton
Journal:  Mol Ther       Date:  2011-09-27       Impact factor: 11.454

2.  Performance evaluation of stationary and semi-stationary acquisition with a non-stationary small animal multi-pinhole SPECT system.

Authors:  Catharina Lange; Ivayla Apostolova; Mathias Lukas; Kai P Huang; Frank Hofheinz; Betina Gregor-Mamoudou; Winfried Brenner; Ralph Buchert
Journal:  Mol Imaging Biol       Date:  2013-11-09       Impact factor: 3.488

Review 3.  Imaging technologies and basic considerations for welfare of laboratory rodents.

Authors:  Jordi L Tremoleda; Jane Sosabowski
Journal:  Lab Anim (NY)       Date:  2015-03       Impact factor: 12.625

4.  Investigation of Image Reconstruction Parameters of the Mediso nanoScan PC Small-Animal PET/CT Scanner for Two Different Positron Emitters Under NEMA NU 4-2008 Standards.

Authors:  Anastasios Gaitanis; George A Kastis; Elena Vlastou; Penelope Bouziotis; Panayotis Verginis; Constantinos D Anagnostopoulos
Journal:  Mol Imaging Biol       Date:  2017-08       Impact factor: 3.488

5.  Development of a Germanium Small-Animal SPECT System.

Authors:  Lindsay C Johnson; Oleg Ovchinnikov; Sepideh Shokouhi; Todd E Peterson
Journal:  IEEE Trans Nucl Sci       Date:  2015-10-09       Impact factor: 1.679

6.  Multimodal assessment of SERS nanoparticle biodistribution post ingestion reveals new potential for clinical translation of Raman imaging.

Authors:  Jos L Campbell; Elliott D SoRelle; Ohad Ilovich; Orly Liba; Michelle L James; Zhen Qiu; Valerie Perez; Carmel T Chan; Adam de la Zerda; Cristina Zavaleta
Journal:  Biomaterials       Date:  2017-04-28       Impact factor: 12.479

7.  Comparative analysis of peptide p5 and serum amyloid P component for imaging AA amyloid in mice using dual-isotope SPECT.

Authors:  Jonathan S Wall; Tina Richey; Angela Williams; Alan Stuckey; Dustin Osborne; Emily Martin; Stephen J Kennel
Journal:  Mol Imaging Biol       Date:  2012-08       Impact factor: 3.488

8.  New whole-body multimodality imaging of gastric cancer peritoneal metastasis combining fluorescence imaging with ICG-labeled antibody and MRI in mice.

Authors:  Akihiro Ito; Yuichi Ito; Shigeru Matsushima; Daisuke Tsuchida; Mai Ogasawara; Junichi Hasegawa; Kazunari Misawa; Eisaku Kondo; Norio Kaneda; Hayao Nakanishi
Journal:  Gastric Cancer       Date:  2013-11-28       Impact factor: 7.370

9.  Tc-99m Radiolabeled Peptide p5 + 14 is an Effective Probe for SPECT Imaging of Systemic Amyloidosis.

Authors:  Stephen J Kennel; Alan Stuckey; Helen P McWilliams-Koeppen; Tina Richey; Jonathan S Wall
Journal:  Mol Imaging Biol       Date:  2016-08       Impact factor: 3.488

10.  Performance evaluation of small-animal multipinhole μSPECT scanners for mouse imaging.

Authors:  Steven Deleye; Roel Van Holen; Jeroen Verhaeghe; Stefaan Vandenberghe; Sigrid Stroobants; Steven Staelens
Journal:  Eur J Nucl Med Mol Imaging       Date:  2013-01-24       Impact factor: 9.236

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.