| Literature DB >> 18652688 |
Frédéric Sériès1, Julie Plante, Yves Lacasse.
Abstract
BACKGROUND: CPAP titration may be completed by automatic apparatus. However, differences in pressure behaviour could interfere with the reliability of pressure recommendations. Our objective was to compare pressure behaviour and effective pressure recommendations between three Automatic CPAP machines (Autoset Spirit, Remstar Auto, GK 420).Entities:
Mesh:
Year: 2008 PMID: 18652688 PMCID: PMC2515310 DOI: 10.1186/1465-9921-9-56
Source DB: PubMed Journal: Respir Res ISSN: 1465-9921
Characteristics of participating subjects.
| Age (y) | 49 (38 – 65) |
| Sex | 14 M/2F |
| BMI (Kg/m2) | 30.8 ± 4.5 |
| ESS | 12.3 ± 1.1 |
| Co morbidities | HBP (n = 6), CAD (n = 4), diabetes (n = 4), gout (n = 1), GERD (n = 1) |
| Medications | Beta blockers (n = 4), ACE (n = 2), AR blockers (n = 2), diuretics (n = 2), nitrates (n = 2), Allopurinol (n = 1). |
| RDI (n/h) | 38. 5 ± 20.0 |
ESS: Epworth Sleepiness score, RDI: respiratory disturbances index. HBP: high blood pressure, CAD: Coronary artery disease, GERD: Gastro-oesophageal Reflux Disease, Mean (range) or SD
Results of the titration studies according to each Auto CPAP device during the first titration night.
| Auto CPAP device | Autoset Spirit | Remstar Auto | GK 420 |
| CPAP usage (h) | 6.2 ± 1.9 a | 4.6 ± 2.4 a | 5.9 ± 1.6 a |
| AHI (n/h) | 8.0 ± 5.6 a | 4.2 ± 3.1 b | 2.9 ± 3.0 c |
| Max pressure (cm H2O) | 11.1 ± 1.4 a | 10.9 ± 2.4 a | 11.5 ± 3.3 a |
| Median pressure (cm H2O) | 7.4 ± 1.3 a | 7.1 ± 1.8 a | 5.7 ± 1.8 b |
| Recommended pressure level (cm H2O) | 9.9 ± 1.2 a | 9.6 ± 2.0 a | 9.3 ± 2.9 a |
| Pressure variability | 2.6 ± 0.9 a | 1.8 ± 0.9 b | 3.0 ± 1.9 ab |
For a given variable (row), columns assigned with different letters denote statistically significant difference. Conversely, for a given variable, columns that are sharing a same given letter are not significantly different. Mean ± SD.
Results of the titration studies according to each Auto CPAP device during the whole week titration.
| Auto CPAP device | Autoset Spirit | Remstar Auto | GK 420 |
| CPAP usage (h) | 5.6 ± 1.9 a | 4.6 ± 1.7 a | 5.0 ± 2.3 a |
| AHI (n/h) | 7.1 ± 4.1 a | 4.7 ± 3.4 b | 3.1 ± 3.1 c |
| Max pressure (cm H2O) | 10.9 ± 1.0 a | 12.6 ± 2.4 b | 11.0 ± 2.4 a |
| Median pressure (cm H2O) | 7.4 ± 1.3 a | 6.6 ± 1.9 a | 5.9 ± 1.8 b |
| Recommended pressure level (cm H2O) | 9.8 ± 1.1 a | 9.0 ± 2.0 a | 9.2 ± 2.7 a |
| Pressure variability | 2.6 ± 0.5 a | 1.8 ± 1.0 b | 2.9 ± 1.6 a |
Significant differences between columns are indicated by the use of different letters as described in table 2.
Figure 1Individual values of the recommended effective pressure (Peff) level obtained after one night and one week of automatic CPAP titration with the three tested apparatus.
Figure 2Bland and Altman representation of the recommended effective pressure (Peff) level obtained after one night minus Peff at one week of automatic CPAP titration against the corresponding mean Peff value for each tested apparatus. (Mean ± 95% CI).