The accuracy of autotitrating CPAP-determined residual apnea-hypopnea index.

Aykut Cilli1, Rusen Uzun, Ugur Bilge.   

Abstract

PURPOSE: Autotitrating continuous positive airway pressure (auto-CPAP) devices not only titrate CPAP pressures but also measure residual respiratory events. The aim of the present study was to determine the accuracy of auto-CPAP-derived residual apnea-hypopnea index (AHI).
METHODS: We studied 137 consecutive patients (72.3% men) with obstructive sleep apnea from January 2008 to December 2010 who underwent in-laboratory overnight polysomnography (PSG) using auto-CPAP. We excluded patients with comorbidities like congestive heart disease, chronic obstructive pulmonary disease, or hypoventilation syndromes and patients with central sleep apnea. Residual AHI obtained from the auto-CPAP device by smart card (CPAP-AHI) was compared simultaneously with AHI from an overnight PSG on auto-CPAP (PSG-AHI) using Bland-Altman analysis and Wilcoxon signed-rank test.
RESULTS: The mean AHI on the diagnostic study was 45.08 ± 1.8. During the titration, auto-CPAP markedly suppressed the respiratory events (PSG-AHI, 3.40 ± 0.20). On the other hand, CPAP-AHI was 3.35 ± 0.17. Bland-Altman analysis showed good agreement between auto-CPAP-AHI and PSG-AHI (AHI mean difference of 0.05, and the limits of agreement for the AHI were from +4.9 to -4.8). Two methods have also been compared with paired samples t test and no statistically significant difference was found (p > 0.05).
CONCLUSION: Auto-CPAP can identify residual respiratory events equivalent to the use of PSG in a selected population.

Entities:  

Mesh:

Substances:

Year:  2012        PMID: 22371206     DOI: 10.1007/s11325-012-0670-x

Source DB:  PubMed          Journal:  Sleep Breath        ISSN: 1520-9512            Impact factor:   2.816


  14 in total

1.  A laboratory validation study of the diagnostic mode of the Autoset system for sleep-related respiratory disorders.

Authors:  B Fleury; D Rakotonanahary; C Hausser-Hauw; B Lebeau; C Guilleminault
Journal:  Sleep       Date:  1996-07       Impact factor: 5.849

2.  Comparison of a limited computerized diagnostic system (ResCare Autoset) with polysomnography in the diagnosis of obstructive sleep apnoea syndrome.

Authors:  J L Kiely; C Delahunty; S Matthews; W T McNicholas
Journal:  Eur Respir J       Date:  1996-11       Impact factor: 16.671

3.  Evaluation of the apnea-hypopnea index determined by the S8 auto-CPAP, a continuous positive airway pressure device, in patients with obstructive sleep apnea-hypopnea syndrome.

Authors:  Kanako Ueno; Takatoshi Kasai; Gregory Brewer; Hisashi Takaya; Ken-ichi Maeno; Satoshi Kasagi; Fusae Kawana; Sugao Ishiwata; Koji Narui
Journal:  J Clin Sleep Med       Date:  2010-04-15       Impact factor: 4.062

4.  The occurrence of sleep-disordered breathing among middle-aged adults.

Authors:  T Young; M Palta; J Dempsey; J Skatrud; S Weber; S Badr
Journal:  N Engl J Med       Date:  1993-04-29       Impact factor: 91.245

5.  Accuracy of autotitrating CPAP to estimate the residual Apnea-Hypopnea Index in patients with obstructive sleep apnea on treatment with autotitrating CPAP.

Authors:  Himanshu Desai; Anil Patel; Pinal Patel; Brydon J B Grant; M Jeffery Mador
Journal:  Sleep Breath       Date:  2009-05-01       Impact factor: 2.816

6.  Continuous positive airway pressure device-based automated detection of obstructive sleep apnea compared to standard laboratory polysomnography.

Authors:  Bharati Prasad; David W Carley; James J Herdegen
Journal:  Sleep Breath       Date:  2009-10-14       Impact factor: 2.816

7.  Comparison of polysomnography with ResCare Autoset in the diagnosis of the sleep apnoea/hypopnoea syndrome.

Authors:  P A Bradley; I L Mortimore; N J Douglas
Journal:  Thorax       Date:  1995-11       Impact factor: 9.139

8.  Practice parameters for the use of autotitrating continuous positive airway pressure devices for titrating pressures and treating adult patients with obstructive sleep apnea syndrome: an update for 2007. An American Academy of Sleep Medicine report.

Authors:  Timothy I Morgenthaler; R Nisha Aurora; Terry Brown; Rochelle Zak; Cathy Alessi; Brian Boehlecke; Andrew L Chesson; Leah Friedman; Vishesh Kapur; Rama Maganti; Judith Owens; Jeffrey Pancer; Todd J Swick
Journal:  Sleep       Date:  2008-01       Impact factor: 5.849

9.  Simultaneous laboratory-based comparison of ResMed Autoset with polysomnography in the diagnosis of sleep apnoea/hypopnoea syndrome.

Authors:  P Mayer; J C Meurice; F Philip-Joet; A Cornette; D Rakotonanahary; N Meslier; J L Pepin; P Lévy; D Veale
Journal:  Eur Respir J       Date:  1998-10       Impact factor: 16.671

10.  Detection of apnoeas, hypopnoeas and arousals by the AutoSet in the sleep apnoea/hypopnoea syndrome.

Authors:  K Rees; P K Wraith; M Berthon-Jones; N J Douglas
Journal:  Eur Respir J       Date:  1998-10       Impact factor: 16.671

View more
  14 in total

1.  Residual Events during Use of CPAP: Prevalence, Predictors, and Detection Accuracy.

Authors:  Joel Reiter; Bashar Zleik; Mihaela Bazalakova; Pankaj Mehta; Robert Joseph Thomas
Journal:  J Clin Sleep Med       Date:  2016-08-15       Impact factor: 4.062

2.  Urgent Need to Improve PAP Management: The Devil Is in Two (Fixable) Details.

Authors:  Robert J Thomas; Matt T Bianchi
Journal:  J Clin Sleep Med       Date:  2017-05-15       Impact factor: 4.062

3.  Man Versus Machine.

Authors:  Christine H Won
Journal:  J Clin Sleep Med       Date:  2017-02-15       Impact factor: 4.062

4.  Detection of upper airway status and respiratory events by a current generation positive airway pressure device.

Authors:  Qing Yun Li; Richard B Berry; Mark G Goetting; Bethany Staley; Haideliza Soto-Calderon; Sheila C Tsai; Jeffrey G Jasko; Allan I Pack; Samuel T Kuna
Journal:  Sleep       Date:  2015-04-01       Impact factor: 5.849

5.  A longitudinal study of the accuracy of positive airway pressure therapy machine-detected apnea-hypopnea events.

Authors:  Yue-Nan Ni; Robert Joseph Thomas
Journal:  J Clin Sleep Med       Date:  2022-04-01       Impact factor: 4.062

6.  A new characterization of adherence patterns to auto-adjusting positive airway pressure in severe obstructive sleep apnea syndrome: clinical and psychological determinants.

Authors:  Rute Sampaio; M Graça Pereira; João C Winck
Journal:  Sleep Breath       Date:  2013-02-06       Impact factor: 2.816

7.  All APAPs Are Not Equivalent for the Treatment of Sleep Disordered Breathing: A Bench Evaluation of Eleven Commercially Available Devices.

Authors:  Kaixian Zhu; Gabriel Roisman; Sami Aouf; Pierre Escourrou
Journal:  J Clin Sleep Med       Date:  2015-07-15       Impact factor: 4.062

8.  Use of the WatchPAT to detect occult residual sleep-disordered breathing in patients on CPAP for obstructive sleep apnea.

Authors:  Matthew Epstein; Tariq Musa; Stephanie Chiu; Jacquelyn Costanzo; Christine Dunne; Federico Cerrone; Robert Capone
Journal:  J Clin Sleep Med       Date:  2020-07-15       Impact factor: 4.062

9.  Interpreting CPAP device respiratory indices in children.

Authors:  Rebecca Mihai; Kirsten Ellis; Margot J Davey; Gillian M Nixon
Journal:  J Clin Sleep Med       Date:  2020-10-15       Impact factor: 4.062

10.  Does either obesity or OSA severity influence the response of autotitrating CPAP machines in very obese subjects?

Authors:  Chris D Turnbull; Ari R Manuel; John R Stradling
Journal:  Sleep Breath       Date:  2015-10-08       Impact factor: 2.816

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.