Efficacy of auto-CPAP in the treatment of obstructive sleep apnea/hypopnea syndrome.

J C Meurice1, I Marc, F Sériès.   

Abstract

The auto-CPAP (Morphée Plus) is characterized by its ability to modify the positive-pressure level applied during the night for the presence or absence of sleep-induced respiratory disorders. The aim of the study was to compare the efficacy of this new mode of CPAP therapy with that of conventional constant-CPAP in the treatment of sleep apnea/hypopnea syndrome (SAHS). Sixteen patients with SAHS were randomly allocated to two groups that were paired for age, apnea/hypopnea index, and mean sleep latency. In the auto-CPAP group, the pressure level could change within fixed limits in both directions (+2 to -4 cm H2O) of the previously determined effective pressure level (Peff). In the constant-CPAP group, patients used the same apparatus (Morphée Plus) in a constant mode at Peff level. At the beginning of the study, the Peff level was determined during a polysomnographic recording. Day-time vigilance was measured subjectively by a standardized questionnaire and objectively by the maintenance of wakefulness test (MWT); Trailmaking tests (TMT) were used to evaluate cognitive functions. After 3 wk of home CPAP therapy, a control sleep study was done with the CPAP machine used in the protocol, and daytime vigilance and cognitive function tests were obtained. Baseline sleep and nocturnal breathing disorders characteristics did not differ between the two groups, and daytime vigilance and cognitive function abnormalities were similarily altered. In both groups, the apnea/hypopnea index was within normal range at the final CPAP sleep study. In the auto-CPAP group, 49.3 +/- 14.9% (mean +/- SD) of home treatment time was spent at a pressure < or = Peff. Home amount of use estimated by the number of sleeping hours with a positive pressure applied was 6.5 +/- 1.0 h in the auto-CPAP group and 5.1 +/- 1.1 h in the constant-CPAP group (p = 0.02). During the control CPAP sleep study, the positive pressure level was significantly lower during Stage III-IV than during the other sleep stages (p = 0.004). The improvement in the MWT and the TMT observed with CPAP therapy was identical in both groups. We conclude that (1) the amount of use during CPAP treatment is higher with auto-CPAP than with constant-CPAP, and (2) Morphée+auto-CPAP is an efficient as conventional CPAP in correcting nocturnal breathing disorders, daytime sleepiness, and cognitive impairment in SAHS.

Entities:  

Mesh:

Year:  1996        PMID: 8564134     DOI: 10.1164/ajrccm.153.2.8564134

Source DB:  PubMed          Journal:  Am J Respir Crit Care Med        ISSN: 1073-449X            Impact factor:   21.405


  25 in total

1.  Application of telemedicine to sleep medicine.

Authors:  D Kristo; A H Eliasson; N C Netzer; T Bigott
Journal:  Sleep Breath       Date:  2001-06       Impact factor: 2.816

2.  Improvement of CPAP Therapy by a Self-Adjusting System.

Authors:  Katrin Pilz; Stefan Thalhofer; Philipp Meissner; Peter Dorow
Journal:  Sleep Breath       Date:  2000       Impact factor: 2.816

3.  CPAP compliance in sleep apnea patients with and without laboratory CPAP titration.

Authors:  Melanie K Means; Jack D Edinger; Aatif M Husain
Journal:  Sleep Breath       Date:  2004-03       Impact factor: 2.816

Review 4.  Sleep.7: positive airway pressure therapy for obstructive sleep apnoea/hypopnoea syndrome.

Authors:  P Gordon; M H Sanders
Journal:  Thorax       Date:  2005-01       Impact factor: 9.139

5.  Executive dysfunction in OSA before and after treatment: a meta-analysis.

Authors:  Michelle Olaithe; Romola S Bucks
Journal:  Sleep       Date:  2013-09-01       Impact factor: 5.849

6.  Evaluation of an auto-CPAP device for treatment of obstructive sleep apnoea.

Authors:  J H Ficker; G H Wiest; G Lehnert; B Wiest; E G Hahn
Journal:  Thorax       Date:  1998-08       Impact factor: 9.139

7.  Automatic nasal continuous positive airway pressure titration in the laboratory: patient outcomes.

Authors:  J R Stradling; C Barbour; D J Pitson; R J Davies
Journal:  Thorax       Date:  1997-01       Impact factor: 9.139

8.  Treatment of Adult Obstructive Sleep Apnea With Positive Airway Pressure: An American Academy of Sleep Medicine Systematic Review, Meta-Analysis, and GRADE Assessment.

Authors:  Susheel P Patil; Indu A Ayappa; Sean M Caples; R Joh Kimoff; Sanjay R Patel; Christopher G Harrod
Journal:  J Clin Sleep Med       Date:  2019-02-15       Impact factor: 4.062

9.  Is automatic CPAP titration as effective as manual CPAP titration in OSAHS patients? A meta-analysis.

Authors:  Weijie Gao; Yinghui Jin; Yan Wang; Mei Sun; Baoyuan Chen; Ning Zhou; Yuan Deng
Journal:  Sleep Breath       Date:  2011-02-24       Impact factor: 2.816

10.  Effect of air leak on the performance of auto-PAP devices: a bench study.

Authors:  Dale Coller; Dawn Stanley; Sairam Parthasarathy
Journal:  Sleep Breath       Date:  2005-12       Impact factor: 2.816

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.