Cynthia H McCollough1, Bernhard Schmidt, Lifeng Yu, Andrew Primak, Stefan Ulzheimer, Herbert Bruder, Thomas G Flohr.
Abstract
This work assesses the temporal resolution of dual-source computed tomography (CT) in a visually intuitive manner. Exploiting the principles of ring artifact creation, a phantom containing a highly attenuating delta function signal (a wire) was rotated at the same speed as the gantry, creating a partial ring artifact where the angular extent of the artifact provides a direct indication of the temporal resolution. A demonstration of the effect of the evaluated 165 and 83 ms nominal temporal resolutions on motion artifacts in cardiac CT is provided using patient data.Entities: