Evaluation of coronary stents and stenoses at different heart rates with dual source spiral CT (DSCT).

Michael M Lell1, Christoph Panknin, Roya Saleh, James W Sayre, Bernhard Schmidt, Howard Dinh, Stefan G Ruehm.   

Abstract

OBJECTIVES: Evaluation of coronary arteries at higher heart rates and in the presence of coronary stents remains problematic. The utilization of dual source computed tomography (DSCT) might improve the visualization of the coronary arteries under these conditions by imaging at a temporal resolution of 83 milliseconds, independent of heart rate.
MATERIALS AND METHODS: Vessel phantoms (diameter 2-4 mm) were attached to a robotic device to simulate cardiac motion and scanned with a DSCT system. The phantoms had either inserts leading to 50% stenosis or carried stents. Images were evaluated for motion artifacts and measurements of the normal, stenotic, and in-stent lumen at different heart rates (50-120 bpm) were performed. Quantile regression analysis was performed to investigate heart rate dependence of the measurement errors.
RESULTS: Visualization of the stenoses and stents was possible without motion artifacts at heart rates of up to 120 bpm. Image quality was similar for the static (0 bpm) and the dynamic (50-120 bpm) scans. Errors for diameter measurements of the vessel lumen and the stenotic lumen were low (3-mm vessel: 1-2%), but considerable for in-stent diameter measurements (3-mm stent: 27-32%). A window/level setting of 1500/300 Hounsfield units was more favorable for stent evaluation. No heart rate dependence was found.
CONCLUSIONS: Depiction of coronary stents with DSCT is possible across a large range of simulated heart rates without motion artifacts and with image quality superior to that of previous generations of CT scanners.

Entities:  

Mesh:

Year:  2007        PMID: 17568277     DOI: 10.1097/RLI.0b013e318041f674

Source DB:  PubMed          Journal:  Invest Radiol        ISSN: 0020-9996            Impact factor:   6.016


  4 in total

1.  Banding and Step-Stair Artifacts on the Cardiac-CT Caused By Pseudo-Ectopic Beats.

Authors:  Amolak Singh; Yash Sethi; Sonya Watkins; Angela Youtsey; Angie Thomas
Journal:  J Radiol Case Rep       Date:  2009-06-01

2.  Measurement of temporal resolution in dual source CT.

Authors:  Cynthia H McCollough; Bernhard Schmidt; Lifeng Yu; Andrew Primak; Stefan Ulzheimer; Herbert Bruder; Thomas G Flohr
Journal:  Med Phys       Date:  2008-02       Impact factor: 4.071

3.  Screening for coronary artery disease in respiratory patients: comparison of single- and dual-source CT in patients with a heart rate above 70 bpm.

Authors:  Vittorio Pansini; Martine Remy-Jardin; Nunzia Tacelli; Jean-Baptiste Faivre; Thomas Flohr; Valérie Deken; Alain Duhamel; Jacques Remy
Journal:  Eur Radiol       Date:  2008-05-08       Impact factor: 5.315

4.  The diagnostic evaluation of dual-source CT (DSCT) in the diagnosis of coronary artery stenoses.

Authors:  Ziqiao Lei; Jin Gu; Qing Fu; Heshui Shi; Haibo Xu; Ping Han; Jianming Yu
Journal:  Pak J Med Sci       Date:  2013-01       Impact factor: 1.088

  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.