Image reconstruction and image quality evaluation for a 64-slice CT scanner with z-flying focal spot.

T G Flohr1, K Stierstorfer, S Ulzheimer, H Bruder, A N Primak, C H McCollough.   

Abstract

We present a theoretical overview and a performance evaluation of a novel z-sampling technique for multidetector row CT (MDCT), relying on a periodic motion of the focal spot in the longitudinal direction (z-flying focal spot) to double the number of simultaneously acquired slices. The z-flying focal spot technique has been implemented in a recently introduced MDCT scanner. Using 32 x 0.6 mm collimation, this scanner acquires 64 overlapping 0.6 mm slices per rotation in its spiral (helical) mode of operation, with the goal of improved longitudinal resolution and reduction of spiral artifacts. The longitudinal sampling distance at isocenter is 0.3 mm. We discuss in detail the impact of the z-flying focal spot technique on image reconstruction. We present measurements of spiral slice sensitivity profiles (SSPs) and of longitudinal resolution, both in the isocenter and off-center. We evaluate the pitch dependence of the image noise measured in a centered 20 cm water phantom. To investigate spiral image quality we present images of an anthropomorphic thorax phantom and patient scans. The full width at half maximum (FWHM) of the spiral SSPs shows only minor variations as a function of the pitch, measured values differ by less than 0.15 mm from the nominal values 0.6, 0.75, 1, 1.5, and 2 mm. The measured FWHM of the smallest slice ranges between 0.66 and 0.68 mm at isocenter, except for pitch 0.55 (0.72 mm). In a centered z-resolution phantom, bar patterns up to 15 lp/cm can be visualized independent of the pitch, corresponding to 0.33 mm longitudinal resolution. 100 mm off-center, bar patterns up to 14 lp/cm are visible, corresponding to an object size of 0.36 mm that can be resolved in the z direction. Image noise for constant effective mAs is almost independent of the pitch. Measured values show a variation of less than 7% as a function of the pitch, which demonstrates correct utilization of the applied radiation dose at any pitch. The product of image noise and square root of the slice width (FWHM of the respective SSP) is the same constant for all slices except 0.6 mm. For the thinnest slice, relative image noise is increased by 17%. Spiral windmill-type artifacts are effectively suppressed with the z-flying focal spot technique, which has the potential to maintain a low artifact level up to pitch 1.5, in this way increasing the maximum volume coverage speed that can be clinically used.

Entities:  

Mesh:

Year:  2005        PMID: 16193784     DOI: 10.1118/1.1949787

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  68 in total

1.  A strategy to decrease partial scan reconstruction artifacts in myocardial perfusion CT: phantom and in vivo evaluation.

Authors:  Juan C Ramirez-Giraldo; Lifeng Yu; Birgit Kantor; Erik L Ritman; Cynthia H McCollough
Journal:  Med Phys       Date:  2012-01       Impact factor: 4.071

2.  Role of neuroradiology in evaluating cerebral aneurysms.

Authors:  A Stafa; M Leonardi
Journal:  Interv Neuroradiol       Date:  2008-10-09       Impact factor: 1.610

3.  Is metal artefact reduction mandatory in cardiac PET/CT imaging in the presence of pacemaker and implantable cardioverter defibrillator leads?

Authors:  Pardis Ghafarian; S M R Aghamiri; Mohammad R Ay; Arman Rahmim; Thomas H Schindler; Osman Ratib; Habib Zaidi
Journal:  Eur J Nucl Med Mol Imaging       Date:  2010-10-20       Impact factor: 9.236

4.  Technical Note: Development and validation of an open data format for CT projection data.

Authors:  Baiyu Chen; Xinhui Duan; Zhicong Yu; Shuai Leng; Lifeng Yu; Cynthia McCollough
Journal:  Med Phys       Date:  2015-12       Impact factor: 4.071

5.  Validation of a Projection-domain Insertion of Liver Lesions into CT Images.

Authors:  Baiyu Chen; Chi Ma; Shuai Leng; Jeff L Fidler; Shannon P Sheedy; Cynthia H McCollough; Joel G Fletcher; Lifeng Yu
Journal:  Acad Radiol       Date:  2016-07-16       Impact factor: 3.173

6.  MSCT is better than stress perfusion imaging for detecting CAD--For.

Authors:  Filippo Cademartiri; Jeroen J Bax
Journal:  Eur J Nucl Med Mol Imaging       Date:  2006-03       Impact factor: 9.236

7.  Optimal image reconstruction intervals for non-invasive coronary angiography with 64-slice CT.

Authors:  Sebastian Leschka; Lars Husmann; Lotus M Desbiolles; Oliver Gaemperli; Tiziano Schepis; Pascal Koepfli; Thomas Boehm; Borut Marincek; Philipp A Kaufmann; Hatem Alkadhi
Journal:  Eur Radiol       Date:  2006-05-13       Impact factor: 5.315

8.  Multidetector CT (64 Slices) of the liver: examination techniques.

Authors:  Andrea Laghi
Journal:  Eur Radiol       Date:  2006-09-29       Impact factor: 5.315

9.  Stereolithographic reproduction of complex cardiac morphology based on high spatial resolution imaging.

Authors:  G F Greil; I Wolf; A Kuettner; M Fenchel; S Miller; P Martirosian; F Schick; M Oppitz; H-P Meinzer; L Sieverding
Journal:  Clin Res Cardiol       Date:  2007-01-22       Impact factor: 5.460

10.  Super-resolution imaging in a multiple layer EPID.

Authors:  Haijian Chen; Joerg Rottmann; Stephen Sf Yip; Daniel Morf; Rony Füglistaller; Josh Star-Lack; George Zentai; Ross Berbeco
Journal:  Biomed Phys Eng Express       Date:  2017-02-21
View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.