| Literature DB >> 17367529 |
Lawrence T Dauer1, Kevin A Casciotta, Yusuf E Erdi, Lawrence N Rothenberg.
Abstract
BACKGROUND: It is estimated that 60 million computed tomography (CT) scans were performed during 2006, with approximately 11% of those performed on children age 0-15 years. Various types of gonadal shielding have been evaluated for reducing exposure to the gonads. The purpose of this study was to quantify the radiation dose reduction to the gonads and its effect on image quality when a wrap-around male pediatric gonad shield was used during CT scanning. This information is obtained to assist the attending radiologist in the decision to utilize such male gonadal shields in pediatric imaging practice.Entities:
Mesh:
Year: 2007 PMID: 17367529 PMCID: PMC1831769 DOI: 10.1186/1471-2342-7-5
Source DB: PubMed Journal: BMC Med Imaging ISSN: 1471-2342 Impact factor: 1.930
Figure 1Humanoid abdominal/pelvic phantom.
Figure 2CT image of Humanoid phantom showing the location of the ion chamber at the level of the pelvis. Ion chamber was located to approximate male gonad geometry (window level set at maximum to enhance the view of the chamber).
Figure 31 mm lead wrap-around gonad shields in three sizes. Flexible gonad protector, Dr. Goos-Suprema, Heidelberg, Germany.
CT technique factors utilized in this study
| Technique | Adult – Helical |
| kVp | 120 |
| mA | 260 |
| Time per rotation (s) | 0.7 |
| mAseff | 194 |
| Scan Field of View | Large |
| Z-axis collimation (mm) | 1.25 |
| Table speed (mm/rot) | 18.75 |
| Pitch | 0.938 |
| Reconstructed Scan Width (mm) | 5 |
Measured ion chamber readings (mR) representing gonadal exposure from an abdominal CT scan
| 1 | 69.7 | 29.4 | 33.2 | 49.8 |
| 2 | 68.7 | 29.5 | ||
| 3 | 68.9 | 28.3 | ||
| 4 | 68.3 | 29.0 | ||
| 5 | 67.2 | 29.2 | ||
| Average +/- std.dev. | 68.6 +/- 0.9 | 29.1 +/- 0.5 | ||
| % Reduction | 58% | 52% | 42% | |
| Reduction Factor | 2.4 | 2.1 | 1.7 |
Measured ion chamber readings (mR) representing gonadal exposure from a pelvic CT scan
| 1 | 4820 | 138.2 | 171.4 | 265.4 |
| 2 | 4800 | 138.3 | ||
| 3 | 4800 | 138.9 | ||
| 4 | 4810 | 136.4 | ||
| 5 | 4800 | 134.2 | ||
| Average +/- std.dev. | 4806 +/- 8.9 | 137.2 +/- 1.9 | ||
| % Reduction | 97% | 96% | 94% | |
| Reduction Factor | 35 | 28 | 18 |
Figure 4Comparison of abdominal CT images. Abdominal CT images of Humanoid phantom both (A) with and (B) without the gonad shield in place. An identical region of interest (ROI), as indicated by the circle, was selected on each image to measure noise.
Figure 5Pelvic CT image of Humanoid phantom. Pelvic CT image with gonad shield in place and directly situated in the beam. Note the severe streaking artifacts and general image degradation.
Comparison of literature reports and present study for scattered dose reduction with the use of gonadal shields during abdominal CT scans (with the testes not in the direct beam)
| Without Shield | With Shield | ||||
| Hidajat et al. 1996 [19] | Phantom, 1 mm capsule | Single-slice sequential, 10 mm slice, 250 mAs/slice, 120 kV | 1.46 | 0.07 | 95 |
| Price et al. 1995 [18] | Phantom, 1 mm wrap-around lead rubber | Single-slice spiral, 10 mm slice, 220 mA, 120 kV | 0.821 | 0.191 | 77 |
| Hohl et al. 2005 [7] | Patient, 1 mm capsule | 16-slice spiral, 16 × 1.5 mm collimation,150 mAseff,120 kV | 2.40 | 0.32 | 87 |
| Present study | Phantom, 1 mm wrap-around canvas | 16-slice spiral, 16 × 1.25 mm collimation 194 mAseff, 120 kV | 0.62 | 0.26 | 58 |
1Values initially reported in mSv and were converted to mGy for comparison, assuming 1 mSv = 1 mGy.