| Literature DB >> 34458197 |
Mohammad Kiapour1, Kourosh Ebrahimnejad Gorji2, Rahele Mehraeen3, Naser Ghaemian4, Fatemeh Niksirat Sustani5, Razzagh Abedi-Firouzjah5, Ali Shabestani Monfared6.
Abstract
BACKGROUND: Computed tomography (CT) is a routine procedure for diagnosing using ionization radiation which has hazardous effects especially on sensitive organs.Entities:
Keywords: Chest CT Scan; Computed Tomography; Lead Apron; Radiation Protection; Testis; Thermoluminescent Dosimetry
Year: 2021 PMID: 34458197 PMCID: PMC8385221 DOI: 10.31661/jbpe.v0i0.2104-1307
Source DB: PubMed Journal: J Biomed Phys Eng ISSN: 2251-7200
Computed tomography (CT) scan parameters used for chest examination for all participants
| Parameter | Value |
|---|---|
| Tube voltage (kVp) | 130 |
| Effective tube current-time product (mAs) | 80 |
| Rotation time (s) | 1 |
| Image slice thickness (mm) | 5 |
| Pitch | 1 |
| CT dose index (CTDIvol, mGy) | 5.6 ± 1.5 |
| Dose length product (DLP, mGy.cm) | 192 ± 30 |
| Total scan time (s) | 9.5 ± 1.1 |
| Scan field of view | Large |
CT: Computed tomography, CTD: Computed tomography dose index, DLP: Dose length product
Figure 1The locations of the thermoluminescent dosimeters (TLDs). Without lead apron (A) and with lead apron (B).
Figure 2The mean surface dose (± SD) at the testicular region with and without the lead apron.
Comparison of several studies reporting the details of computed tomography (CT) technique and dose reduction along with the current study method and results.
| Testicular dose (mSv) | ||||||
|---|---|---|---|---|---|---|
| Study | Study Type | Scanning Protocol | Year | Without shielding | With shielding | Reduction (%) |
| Present study | Patient study, 0.5-mm apron shield | 16-slice spiral, 16×1.2-mm collimation, 80 mAseff, 10 kV | 2020 | 0.09 | 0.04 | 56 |
| Price et al. [ | Phantom study, 1-mm wraparound apron | Single-slice spiral, 10-mm slice thickness, 220 mA, 120 kV | 1999 | 0.82 | 0.19-0.05 | 77-93 |
| Hidajat et al. [ | Phantom study, 1-mm testicular capsule | Single-slice sequential, 10-mm slice thickness, 250 mAs/slice, 120 kV | 1996 | 1.46 | 0.07 | 95 |
| Hohl et al. [ | Patient study, 1-mm testicular capsule | 16-slice spiral, 16 × 1.5-mm collimation, 150 mAseff, 120 kV | 2005 | 2.40 | 0.32 | 87 |
| Grobe et al. [ | Patient study, 1-mm wraparound circular | 16-slice spiral, 16 × 1.5-mm collimation, 150 mAs, 120 kV | 2009 | 19.2 | 0.7 | 96 |
| Iball et al. [ | Phantom study, shield designed and 0.7-mm lead | 64-slice spiral, 64 × 0.6-mm, 150 mAseff, 120 kV | 2011 | 0.018 | 0.001 | 93 |
| Sancaktutar et al. [ | Patient study, 0.7-mm wraparound circular | 64-slice spiral, 16 × 1.25-mm collimation, 250 mAseff, 120 kV | 2012 | 69 | 6.8 | 90.2 |