Peter Hiles1, Patrick Gilligan2,3, John Damilakis4,5, Eric Briers6, Cristian Candela-Juan2,7, Dario Faj8,9, Shane Foley10,11, Guy Frija4,12, Claudio Granata13,14, Hugo de Las Heras Gala2,15, Ruben Pauwels16, Marta Sans Merce8,17, Georgios Simantirakis8,18, Eliseo Vano4,19.
Abstract
Patient contact shielding has been in use for many years in radiology departments in order to reduce the effects and risks of ionising radiation on certain organs. New technologies in projection imaging and CT scanning such as digital receptors and automatic exposure control systems have reduced doses and improved image consistency. These changes and a greater understanding of both the benefits and the risks from the use of shielding have led to a review of shielding use in radiology. A number of professional bodies have already issued guidance in this regard. This paper represents the current consensus view of the main bodies involved in radiation safety and imaging in Europe: European Federation of Organisations for Medical Physics, European Federation of Radiographer Societies, European Society of Radiology, European Society of Paediatric Radiology, EuroSafe Imaging, European Radiation Dosimetry Group (EURADOS), and European Academy of DentoMaxilloFacial Radiology (EADMFR). It is based on the expert recommendations of the Gonad and Patient Shielding (GAPS) Group formed with the purpose of developing consensus in this area. The recommendations are intended to be clear and easy to use. They are intended as guidance, and they are developed using a multidisciplinary team approach. It is recognised that regulations, custom and practice vary widely on the use of patient shielding in Europe and it is hoped that these recommendations will inform a change management program that will benefit patients and staff.Entities:
Keywords: Consensus; Protective devices; Radiation dosage; Radiology; Tomography (X-ray computed)
Year: 2021 PMID: 34939154 PMCID: PMC8695402 DOI: 10.1186/s13244-021-01085-4
Source DB: PubMed Journal: Insights Imaging ISSN: 1869-4101
Rationale for consensus statements
| Rationale | Consensus Recommendation | Symbol |
|---|---|---|
| Evidence that using patient contact shielding is beneficial and effective | ‘Should use shielding’ |
|
| General agreement favours usefulness of patient contact shielding in some circumstances | ‘May use shielding’ |
|
| Evidence or general agreement not to use patient contact shielding | ‘Not recommended to use shielding’ |
|
Fig. 1Female lifetime risks of cancer incidence by organ and age for a composite Euro-American population (% per Gy), based on data from reference [19]
| Application | Imaging modality | Inside or outside FOV | Recommendation | Symbol |
|---|---|---|---|---|
| Male and female gonad contact shielding | All X-ray | Both | ‘Not recommended to use shielding’ |
|
| Application | Imaging modality | Inside or outside FOV | Recommendation | Symbol |
|---|---|---|---|---|
| Thyroid contact shielding | All X-ray (except Ceph.) | Inside | ‘Not recommended to use shielding’ |
|
| Thyroid contact shielding | Cephalometric radiography | Inside | ‘May use shielding’ |
|
| Thyroid contact shielding | Radiography, Mammography, Fluoroscopy, CT | Outside | ‘Not recommended to use shielding’ |
|
| Thyroid contact shielding | Dental intraoral and cephalometric radiography | Outside | ‘May use shielding’ |
|
| Thyroid contact shielding | CBCT | Outside | ‘May use shielding’ |
|
| Application | Imaging modality | Inside or outside FOV | Recommendation | Symbol |
|---|---|---|---|---|
| Breast contact shielding | All X-ray | Both | ‘Not recommended to use shielding’ |
|
| Application | Imaging modality | Inside or outside FOV | Recommendation | Symbol |
|---|---|---|---|---|
| Eye lens contact shielding | All X-ray | Both | ‘Not recommended to use shielding’ |
|
| Application | Imaging modality | Inside or outside FOV | Recommendation | Symbol |
|---|---|---|---|---|
| Embryo / Fetal contact shielding | All X-ray | Inside | ‘Not recommended to use shielding’ |
|
| Embryo / Fetal contact shielding | Radiography, Mammography, Fluoroscopy, Dental Radiography, CT | Outside | ‘Not recommended to use shielding’ |
|