Comparison of rebound tonometry with Goldmann applanation tonometry and correlation with central corneal thickness.

M E Iliev1, D Goldblum, K Katsoulis, C Amstutz, B Frueh.   

Abstract

BACKGROUND/AIMS: Rebound tonometry (RT) is performed without anaesthesia with a hand held device. The primary aim was to compare RT with Goldmann applanation tonometry (GAT) and to correlate with central corneal thickness (CCT). The secondary aim was to prove tolerability and practicability of RT under "study conditions" and "routine practice conditions."
METHODS: In group 1 (52 eyes/28 patients), all measurements were taken by the same physician, in the same room and order: non-contact optical pachymetry, RT, slit lamp inspection, GAT. Patients were questioned about discomfort or pain. In group 2 (49 eyes/27 patients), tonometry was performed by three other physicians during routine examinations.
RESULTS: RT was well tolerated and safe. Intraocular pressure (IOP) ranged between 6 mm Hg and 48 mm Hg. No different trends were found between the groups. RT tended to give slightly higher readings: n = 101, mean difference 1.0 (SD 2.17) mm Hg; 84.1% of RT readings within plus or minus 3 mm Hg of GAT; 95% confidence interval in the Bland-Altman analysis -3.2 mm Hg to +5.2 mm Hg. Both RT and GAT showed a weak positive correlation with CCT (r2 0.028 and 0.025, respectively).
CONCLUSIONS: RT can be considered a reliable alternative for clinical screening and in cases where positioning of the head at the slit lamp is impossible or topical preparations are to be avoided.

Entities:  

Mesh:

Year:  2006        PMID: 16672330      PMCID: PMC1857178          DOI: 10.1136/bjo.2005.089870

Source DB:  PubMed          Journal:  Br J Ophthalmol        ISSN: 0007-1161            Impact factor:   4.638


  14 in total

1.  A new induction-based impact method for measuring intraocular pressure.

Authors:  A I Kontiola
Journal:  Acta Ophthalmol Scand       Date:  2000-04

2.  New Tonopen XL: comparison with the Goldmann tonometer.

Authors:  M Iester; A Mermoud; F Achache; S Roy
Journal:  Eye (Lond)       Date:  2001-02       Impact factor: 3.775

3.  Clinical evaluation of the new TGDc-01 "PRA" palpebral tonometer: comparison with contact and non-contact tonometry.

Authors:  Carlos García Resúa; Maria J Giráldez Fernández; Alejandro Cerviño Expósito; Javier González Pérez; Eva Yebra-Pimentel
Journal:  Optom Vis Sci       Date:  2005-02       Impact factor: 1.973

4.  Clinical comparison of air-puff and Goldmann tonometers.

Authors:  H L Brencher; P Kohl; A R Reinke; R L Yolton
Journal:  J Am Optom Assoc       Date:  1991-05

5.  Non-invasive determination of intraocular pressure in the rat eye. Comparison of an electronic tonometer (TonoPen), and a rebound (impact probe) tonometer.

Authors:  David Goldblum; Antti Ilmari Kontiola; Thom Mittag; Bin Chen; John Danias
Journal:  Graefes Arch Clin Exp Ophthalmol       Date:  2002-09-26       Impact factor: 3.117

6.  The induction/impact tonometer: a new instrument to measure intraocular pressure in the rat.

Authors:  A I Kontiola; D Goldblum; T Mittag; J Danias
Journal:  Exp Eye Res       Date:  2001-12       Impact factor: 3.467

7.  Statistical methods for assessing agreement between two methods of clinical measurement.

Authors:  J M Bland; D G Altman
Journal:  Lancet       Date:  1986-02-08       Impact factor: 79.321

8.  Clinical comparison of the XPERT non-contact tonometer and the conventional Goldmann applanation tonometer.

Authors:  M K Hansen
Journal:  Acta Ophthalmol Scand       Date:  1995-04

9.  Central corneal thickness measurements in patients with normal tension glaucoma, primary open angle glaucoma, pseudoexfoliation glaucoma, or ocular hypertension.

Authors:  A C Ventura; M Böhnke; D S Mojon
Journal:  Br J Ophthalmol       Date:  2001-07       Impact factor: 4.638

10.  Method for the noninvasive measurement of intraocular pressure in mice.

Authors:  John Danias; Antti I Kontiola; Theodoros Filippopoulos; Thom Mittag
Journal:  Invest Ophthalmol Vis Sci       Date:  2003-03       Impact factor: 4.799

View more
  38 in total

1.  Calibration of the TonoLab tonometer in mice with spontaneous or experimental glaucoma.

Authors:  Mary E Pease; Frances E Cone; Scott Gelman; Janice L Son; Harry A Quigley
Journal:  Invest Ophthalmol Vis Sci       Date:  2011-02-22       Impact factor: 4.799

Review 2.  Rebound tonometry: new opportunities and limitations of non-invasive determination of intraocular pressure.

Authors:  A Cervino
Journal:  Br J Ophthalmol       Date:  2006-12       Impact factor: 4.638

3.  Comparison of the iCare rebound tonometer and the Goldmann applanation tonometer over a wide IOP range.

Authors:  S Munkwitz; A Elkarmouty; E M Hoffmann; N Pfeiffer; H Thieme
Journal:  Graefes Arch Clin Exp Ophthalmol       Date:  2008-01-12       Impact factor: 3.117

4.  Early and frequent development of ocular hypertension in children with nephrotic syndrome.

Authors:  Emi Kawaguchi; Kenji Ishikura; Riku Hamada; Yoshinobu Nagaoka; Yoshihiko Morikawa; Tomoyuki Sakai; Yuko Hamasaki; Hiroshi Hataya; Eiichiro Noda; Masaru Miura; Takashi Ando; Masataka Honda
Journal:  Pediatr Nephrol       Date:  2014-05-24       Impact factor: 3.714

5.  Comparison of Goldmann applanation tonometry, rebound tonometry and dynamic contour tonometry in normal and glaucomatous eyes.

Authors:  Fatih Özcura; Nilgün Yildirim; Afsun Şahin; Ertuğrul Çolak
Journal:  Int J Ophthalmol       Date:  2015-04-18       Impact factor: 1.779

6.  The influence of refractive errors on IOP measurement by rebound tonometry (ICare) and Goldmann applanation tonometry.

Authors:  Teresio Avitabile; Antonio Longo; Daniela Rocca; Roberta Amato; Caterina Gagliano; Marine Castaing
Journal:  Graefes Arch Clin Exp Ophthalmol       Date:  2009-08-29       Impact factor: 3.117

7.  IOP agreement between I-Care TA01 rebound tonometer and the Goldmann applanation tonometer in eyes with and without glaucoma.

Authors:  Gautam Sinha; Shikha Gupta; Shreyas Temkar; Veena Pandey; Ramanjit Sihota; Tanuj Dada
Journal:  Int Ophthalmol       Date:  2014-12-16       Impact factor: 2.031

8.  [Rebound tonometry after vitreoretinal surgery].

Authors:  A Rosentreter; R Hoerster; T Schick; N Eter; T S Dietlein; S Fauser
Journal:  Ophthalmologe       Date:  2015-11       Impact factor: 1.059

9.  [Evaluation of the self-tonometer Icare ONE in comparison to Goldmann applanation tonometry].

Authors:  V Witte; Ä Glass; R Beck; R Guthoff
Journal:  Ophthalmologe       Date:  2012-10       Impact factor: 1.059

10.  Central corneal thickness does not correlate with TonoLab-measured IOP in several mouse strains with single transgenic mutations of matricellular proteins.

Authors:  Ayan Chatterjee; Dong-Jin Oh; Min Hyung Kang; Douglas J Rhee
Journal:  Exp Eye Res       Date:  2013-06-24       Impact factor: 3.467

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.