New Tonopen XL: comparison with the Goldmann tonometer.

M Iester1, A Mermoud, F Achache, S Roy.   

Abstract

PURPOSE: To compare the intraocular pressure (IOP) values obtained using a Goldmann tonometer (Haag-Streit) with those obtained with the new Tonopen XL (Mentor), which has certain differences compared with first- and second-generation models.
METHODS: The IOPs of 104 patients were assessed by Goldmann tonometer and Tonopen XL tonometer. Goldmann measurements was done first in 145 eyes and Tonopen measurements were done first in 53 eyes. Four observers measured the IOP. Observers A, B and C used the Goldmann tonometer first and then the Tonopen XL, while observer D used the Tonopen XL first and then the Goldmann tonometer. The results were analysed by descriptive analysis and, when the distribution of the data was normal, paired t-test and Pearson's r coefficient were used to compare and correlate IOP measurements between Goldmann and Tonopen measurements. When the distribution of the data was non-normal, the Wilcoxon matched-pair test and Spearman coefficient were used. The agreement between Goldmann and Tonopen values was also calculated. ANOVA test was used to compare the difference obtained by 'Goldmann minus Tonopen' measurements among the three different observers.
RESULTS: A statistically significant difference (p < 0.0001) was found between the IOP readings obtained by Goldmann tonometer and the Tonopen XL and a significant correlation was found between the Goldmann values and Tonopen XL values (p < 0.001). When the Goldmann IOP was more than 20 mmHg the Tonopen XL measurements were lower than the Goldmann values. Also in this group this difference was statistically significant. No significant difference was found between Goldmann values and Tonopen values among the three observers, even though a significant difference was found between Goldmann values and Tonopen values for observer B. When the values obtained by first the Goldmann tonometer and then the Tonopen XL were compared with those obtained by first the Tonopen XL and then the Goldmann tonometer, no significant difference was found between the two groups.
CONCLUSION: The new Tonopen XL provides similar results to the Goldmann tomometer in 62% of the cases and was slightly less accurate than the Goldmann tonometer for extreme values, just like the previous Tonopen. Nevertheless the precision is good enough for the purpose of adequate screening.

Entities:  

Mesh:

Year:  2001        PMID: 11318296     DOI: 10.1038/eye.2001.13

Source DB:  PubMed          Journal:  Eye (Lond)        ISSN: 0950-222X            Impact factor:   3.775


  24 in total

1.  Deviations between transpalpebral tonometry using TGDc-01 and Goldmann applanation tonometry depending on the IOP level.

Authors:  Annette Troost; Kathrin Specht; Frank Krummenauer; Sung Hyun Yun; Oliver Schwenn
Journal:  Graefes Arch Clin Exp Ophthalmol       Date:  2005-03-15       Impact factor: 3.117

2.  A comparison of four methods of tonometry: method agreement and interobserver variability.

Authors:  P-A Tonnu; T Ho; K Sharma; E White; C Bunce; D Garway-Heath
Journal:  Br J Ophthalmol       Date:  2005-07       Impact factor: 4.638

3.  Comparison of simultaneous readings of intraocular pressure in rabbits using Perkins handheld, Tono-Pen XL, and TonoVet tonometers.

Authors:  Giedrius Kalesnykas; Hannu Uusitalo
Journal:  Graefes Arch Clin Exp Ophthalmol       Date:  2006-11-21       Impact factor: 3.117

4.  Evaluation of the effectiveness of ophthalmic assistants as screeners for glaucoma in North India.

Authors:  S K Sinha; N Astbury
Journal:  Eye (Lond)       Date:  2011-07-01       Impact factor: 3.775

5.  Agreement between diurnal variations of intraocular pressure by Tono-Pen and Goldmann applanation tonometer in patients on topical anti-glaucoma medication.

Authors:  Shikha Gupta; Gautam Sinha; Reetika Sharma; Bhagabat Nayak; Bharat Patil; Bibhuti Kashyap; Abdul Shameer; Tanuj Dada
Journal:  Int Ophthalmol       Date:  2015-03-28       Impact factor: 2.031

6.  Tono-pen measurement of intraocular pressure under topical anaesthesia in full term normal newborns.

Authors:  Sagili Chandrasekhara Reddy; Rosnita Alias
Journal:  Int J Ophthalmol       Date:  2014-02-18       Impact factor: 1.779

7.  Transpalpebral tonometry: reliability and comparison with Goldmann applanation tonometry and palpation in healthy volunteers.

Authors:  A Troost; S H Yun; K Specht; F Krummenauer; O Schwenn
Journal:  Br J Ophthalmol       Date:  2005-03       Impact factor: 4.638

8.  Effectiveness of the ICare rebound tonometer in patients with overestimated intraocular pressure due to tight orbit syndrome.

Authors:  You Kyung Lee; Y K Lee; Ji Young Lee; J Y Lee; Jung Il Moon; J I Moon; Myoung Hee Park; M H Park
Journal:  Jpn J Ophthalmol       Date:  2014-08-29       Impact factor: 2.447

9.  Intraocular Pressure Changes Following Intravitreal Melphalan and Topotecan for the Treatment of Retinoblastoma With Vitreous Seeding.

Authors:  Matthew D Karl; Jasmine H Francis; Saipriya Iyer; Brian Marr; David H Abramson
Journal:  J Pediatr Ophthalmol Strabismus       Date:  2017-01-17       Impact factor: 1.402

10.  Comparative evaluation of TonoPen AVIA, Goldmann applanation tonometry and non-contact tonometry.

Authors:  Shibal Bhartiya; Shveta Jindal Bali; Reetika Sharma; Neha Chaturvedi; Tanuj Dada
Journal:  Int Ophthalmol       Date:  2011-07-15       Impact factor: 2.031

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.