The influence of refractive errors on IOP measurement by rebound tonometry (ICare) and Goldmann applanation tonometry.

Teresio Avitabile1, Antonio Longo, Daniela Rocca, Roberta Amato, Caterina Gagliano, Marine Castaing.   

Abstract

PURPOSE: To evaluate the effect of refractive errors and central corneal thickness (CCT) on the measurement of intraocular pressure (IOP) by ICare rebound tonometer (RT), and its agreement with measurements by Goldmann applanation tonometer (GAT). PATIENTS AND METHODS: Two observers measured the IOP by using RT and GAT in four groups of healthy volunteers with emmetropic (n = 78), hyperopic (n = 83), myopic (n = 87) and astigmatic (n = 79) eyes. Refraction was assessed by an autorefractometer. CCT was assessed by ultrasound pachymetry.
RESULTS: In all groups, no significant interobserver difference was seen in IOP values detected by both tonometers (Wilcoxon signed-rank test not significant). In all groups, IOP values were higher as measured by RT than by GAT (paired t-test p = 0.000): mean RT-GAT difference was higher in myopic eyes (+1.6 +/- 1.8 mmHg), and it was less than 1 mmHg in the other groups. RT-GAT difference was correlated to the refraction (p < 0.001), and it was greater when an higher IOP was detected by RT (significant correlation between RT-GAT difference and IOP by RT, p < 0.001). Compared with GAT values, the IOP readings by RT were greater than 2 mmHg in respectively 17.9% (emmetropic), 13.3% (hyperopic), 34.5% (myopic) and 7.6% (astigmatic) of the eyes. With both tonometers, in all groups the IOP values were correlated with CCT (p < 0.05), but the discrepancy between RT and GAT values was not related to CCT.
CONCLUSIONS: In all groups of subjects, higher IOP values were detected by RT; the IOP readings exceed the GAT values usually in a range of less than 1 mmHg, except when RT detects IOP >18 mmHg and generally in myopic eyes; RT-GAT discrepancy is related to the refractive error, but not to CCT.

Entities:  

Mesh:

Year:  2009        PMID: 19727794     DOI: 10.1007/s00417-009-1176-5

Source DB:  PubMed          Journal:  Graefes Arch Clin Exp Ophthalmol        ISSN: 0721-832X            Impact factor:   3.117


  24 in total

1.  Clinical evaluation of rebound tonometer.

Authors:  Leon N Davies; Hannah Bartlett; Edward A H Mallen; James S Wolffsohn
Journal:  Acta Ophthalmol Scand       Date:  2006-04

2.  Agreement of rebound tonometer in measuring intraocular pressure with three types of applanation tonometers.

Authors:  Makoto Nakamura; Urtogtah Darhad; Yasuko Tatsumi; Miyuki Fujioka; Azusa Kusuhara; Hidetaka Maeda; Akira Negi
Journal:  Am J Ophthalmol       Date:  2006-08       Impact factor: 5.258

3.  Comparison of the iCare rebound tonometer and the Goldmann applanation tonometer over a wide IOP range.

Authors:  S Munkwitz; A Elkarmouty; E M Hoffmann; N Pfeiffer; H Thieme
Journal:  Graefes Arch Clin Exp Ophthalmol       Date:  2008-01-12       Impact factor: 3.117

4.  The induction/impact tonometer: a new instrument to measure intraocular pressure in the rat.

Authors:  A I Kontiola; D Goldblum; T Mittag; J Danias
Journal:  Exp Eye Res       Date:  2001-12       Impact factor: 3.467

5.  Comparison of ICare tonometer with Goldmann applanation tonometer in glaucoma patients.

Authors:  Paolo Brusini; Maria Letizia Salvetat; Marco Zeppieri; Claudia Tosoni; Lucia Parisi
Journal:  J Glaucoma       Date:  2006-06       Impact factor: 2.503

6.  Effect of corneal thickness on dynamic contour, rebound, and goldmann tonometry.

Authors:  Jose M Martinez-de-la-Casa; Julian Garcia-Feijoo; Eva Vico; Ana Fernandez-Vidal; Jose M Benitez del Castillo; Mohamed Wasfi; J Garcia-Sanchez
Journal:  Ophthalmology       Date:  2006-09-25       Impact factor: 12.079

7.  The influence of corneal properties on rebound tonometry.

Authors:  Wan-sang Chui; Andrew Lam; Davie Chen; Roger Chiu
Journal:  Ophthalmology       Date:  2007-06-15       Impact factor: 12.079

8.  Measuring intraocular pressure with the Pulsair 3000 and Rebound tonometers in elderly patients without an anesthetic.

Authors:  Antti Kontiola; Päivi Puska
Journal:  Graefes Arch Clin Exp Ophthalmol       Date:  2003-11-21       Impact factor: 3.117

9.  Clinical comparison of the Icare tonometer and Goldmann applanation tonometry.

Authors:  Nima Pakrou; Tim Gray; Richard Mills; John Landers; Jamie Craig
Journal:  J Glaucoma       Date:  2008 Jan-Feb       Impact factor: 2.503

10.  Corneal hysteresis and axial length among Chinese secondary school children: the Xichang Pediatric Refractive Error Study (X-PRES) report no. 4.

Authors:  Yue Song; Nathan Congdon; Liping Li; Zhongxia Zhou; Kai Choi; Dennis S C Lam; Chi Pui Pang; Zhenling Xie; Xueyu Liu; Abhishek Sharma; Weihong Chen; Mingzhi Zhang
Journal:  Am J Ophthalmol       Date:  2008-03-10       Impact factor: 5.258

View more
  8 in total

1.  Comparison of intraocular pressure measurements with different contact tonometers in young healthy persons.

Authors:  Saulius Galgauskas; Rasa Strupaite; Ernesta Strelkauskaite; Rimvydas Asoklis
Journal:  Int J Ophthalmol       Date:  2016-01-18       Impact factor: 1.779

2.  A Comparison of Applanation and Rebound Tonometers in Young Chicks.

Authors:  Lisa A Ostrin; Christine F Wildsoet
Journal:  Curr Eye Res       Date:  2020-07-03       Impact factor: 2.424

3.  The utility of rebound tonometer in late elderly subjects.

Authors:  Rei Sakata; Jiro Numaga
Journal:  Clin Interv Aging       Date:  2014-01-22       Impact factor: 4.458

4.  Comprasion of ICare rebound tonometer and Goldmann applanation tonometer in high myopia.

Authors:  Halil Huseyin Cagatay; Metin Ekinci; Zeliha Yazar; Gokcen Gokce; Erdinc Ceylan
Journal:  ScientificWorldJournal       Date:  2014-10-14

5.  Agreement of patient-measured intraocular pressure using rebound tonometry with Goldmann applanation tonometry (GAT) in glaucoma patients.

Authors:  Shaoying Tan; Marco Yu; Nafees Baig; Linda Hansapinyo; Clement C Tham
Journal:  Sci Rep       Date:  2017-02-06       Impact factor: 4.379

6.  Analysis of differences in intraocular pressure evaluation performed with contact and non-contact devices.

Authors:  Michele Lanza; Michele Rinaldi; Ugo Antonello Gironi Carnevale; Silvio di Staso; Mario Bifani Sconocchia; Ciro Costagliola
Journal:  BMC Ophthalmol       Date:  2018-09-03       Impact factor: 2.209

7.  Association analysis of cigarette smoking with onset of primary open-angle glaucoma and glaucoma-related biometric parameters.

Authors:  Degui Wang; Yuqiang Huang; Chukai Huang; Pengfei Wu; Jianwei Lin; Yuqian Zheng; Yi Peng; Yuanbo Liang; Jian-Huan Chen; Mingzhi Zhang
Journal:  BMC Ophthalmol       Date:  2012-11-27       Impact factor: 2.209

8.  Hypobaric hypoxia: effects on intraocular pressure and corneal thickness.

Authors:  Marcella Nebbioso; Stefano Fazio; Dario Di Blasio; Nicola Pescosolido
Journal:  ScientificWorldJournal       Date:  2014-01-16
  8 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.