[Evaluation of the self-tonometer Icare ONE in comparison to Goldmann applanation tonometry].

V Witte1, Ä Glass, R Beck, R Guthoff.   

Abstract

OBJECTIVE: The aim of this clinical study was to demonstrate the practicability of self-measured intraocular pressure and to evaluate the reliability by comparing the data with those obtained by Goldmann applanation tonometry (GAT).
METHODS: A total of 40 patients aged between 44 and 82 years with glaucoma were introduced to the handling of the tonometer. The self-measurements were done for 1-3 days following the medical measurement by GAT. The data were saved as "correct" or in the case of wrong handling as "false". The impressions of the patients were obtained by a questionnaire.
RESULTS: A total number of 191 single measurements were registered by the Icare ONE and of these there were 97 (50.8%) signed "false". Of the patients 45% reached a maximum difference of ±5 mmHg between GAT and self-measurement in every single measurement. In the subgroup of under 60-year-old patients 70% reached this result. There were no indications of a systematic error. Of the probands 60% considered the handling of the Icare ONE as difficult. Nevertheless, 80% could imagine using the self-tonometer at home.
CONCLUSIONS: The differences between the self-measurements and the GAT were highly fluctuating in some cases. In the group of patients younger than 60 years the agreement was better, so problems with the handling of the Icare ONE may be an important factor. However, the acceptance in the patients tested was high and continuous pressure measurements at home could be reasonable. Advancements in the handling and reliability are needed to improve clinical application of the measured values.

Entities:  

Mesh:

Year:  2012        PMID: 22828746     DOI: 10.1007/s00347-012-2526-y

Source DB:  PubMed          Journal:  Ophthalmologe        ISSN: 0941-293X            Impact factor:   1.059


  29 in total

1.  Self-tonometry with the Ocuton S versus Goldmann tonometry.

Authors:  I Theofylaktopoulos; M Diestelhorst; G K Krieglstein
Journal:  Graefes Arch Clin Exp Ophthalmol       Date:  1999-09       Impact factor: 3.117

2.  [Self-measurement of intraocular pressure].

Authors:  R Theisen; V Scherer; K W Ruprecht
Journal:  J Fr Ophtalmol       Date:  2001-04       Impact factor: 0.818

3.  Comparing applanation tonometry and rebound tonometry in glaucomatous and ocular hypertensive eyes.

Authors:  Marino Marini; Stefano Da Pozzo; Agostino Accardo; Tiziana Canziani
Journal:  Eur J Ophthalmol       Date:  2011 May-Jun       Impact factor: 2.597

4.  Comparison of the iCare rebound tonometer and the Goldmann applanation tonometer over a wide IOP range.

Authors:  S Munkwitz; A Elkarmouty; E M Hoffmann; N Pfeiffer; H Thieme
Journal:  Graefes Arch Clin Exp Ophthalmol       Date:  2008-01-12       Impact factor: 3.117

5.  Statistical methods for assessing agreement between two methods of clinical measurement.

Authors:  J M Bland; D G Altman
Journal:  Lancet       Date:  1986-02-08       Impact factor: 79.321

6.  Effect of corneal thickness on dynamic contour, rebound, and goldmann tonometry.

Authors:  Jose M Martinez-de-la-Casa; Julian Garcia-Feijoo; Eva Vico; Ana Fernandez-Vidal; Jose M Benitez del Castillo; Mohamed Wasfi; J Garcia-Sanchez
Journal:  Ophthalmology       Date:  2006-09-25       Impact factor: 12.079

7.  [Intraocular pressure measurements with the Proview self-tonometer in comparison of Goldmann applanation tonometry in healthy and glaucomatous eyes].

Authors:  M W Meyer; V Yevelenko; L Hoy; A Meyer; C Erb
Journal:  Klin Monbl Augenheilkd       Date:  2006-11       Impact factor: 0.700

8.  The influence of corneal properties on rebound tonometry.

Authors:  Wan-sang Chui; Andrew Lam; Davie Chen; Roger Chiu
Journal:  Ophthalmology       Date:  2007-06-15       Impact factor: 12.079

9.  Comparison of the rebound tonometer with the Goldmann applanation tonometer in glaucoma patients.

Authors:  Afsun Sahin; Leyla Niyaz; Nilgun Yildirim
Journal:  Clin Exp Ophthalmol       Date:  2007 May-Jun       Impact factor: 4.207

10.  Clinical comparison of the Icare tonometer and Goldmann applanation tonometry.

Authors:  Nima Pakrou; Tim Gray; Richard Mills; John Landers; Jamie Craig
Journal:  J Glaucoma       Date:  2008 Jan-Feb       Impact factor: 2.503

View more
  4 in total

1.  Twenty-four-hour efficacy of preservative-free tafluprost for open-angle glaucoma patients, assessed by home intraocular pressure (Icare-ONE) and blood-pressure monitoring.

Authors:  Soon Young Cho; Yong Yeon Kim; Chungkwon Yoo; Tae-Eun Lee
Journal:  Jpn J Ophthalmol       Date:  2015-09-28       Impact factor: 2.447

2.  [Ease of handling of first and second generation rebound tonometers].

Authors:  N Mihailovic; J Termühlen; M Alnawaiseh; N Eter; T S Dietlein; A Rosentreter
Journal:  Ophthalmologe       Date:  2016-04       Impact factor: 1.059

3.  Comparison of Icare Pro Tonometry and Icare One Tonometry Measurements in Healthy Eyes

Authors:  Hüseyin Mayalı; Çağlar Sarıgül; Emin Kurt; Özcan Rasim Kayıkçıoğlu; Süleyman Sami İlker
Journal:  Turk J Ophthalmol       Date:  2019-06-27

Review 4.  Teleglaucoma: ready to go?

Authors:  N G Strouthidis; G Chandrasekharan; J P Diamond; I E Murdoch
Journal:  Br J Ophthalmol       Date:  2014-04-10       Impact factor: 4.638

  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.