[Rebound tonometry after vitreoretinal surgery].

A Rosentreter1,2, R Hoerster3, T Schick3, N Eter4, T S Dietlein3, S Fauser3.   

Abstract

PURPOSE: The aim of this study was to analyze the practicability and comparability of the Icare rebound tonometer (RT) versus the Schiötz indentation tonometer (SIT) and the Goldmann applanation tonometer (GAT) for measuring intraocular pressure (IOP) in patients after pars plana vitrectomy (PPV).
METHODS: A total of 100 eyes from 100 patients who underwent vitreoretinal surgery in the Department of Ophthalmology, University of Cologne were included in this prospective analysis. The IOP was measured using RT preoperatively, on the day of surgery and 2 days after surgery, using SIT on the day of surgery and GAT preoperatively and 2 days after surgery. For the evaluation eyes were divided into subgroups with respect to the endotamponade selected and the IOP level.
RESULTS: The mean preoperative IOP for all enrolled eyes was 15.4 ± 8.0 mmHg for RT and 16.1 ± 7.9 mmHg for GAT. Bland-Altman analysis revealed a bias between RT and GAT of - 0.6 mmHg. Bland-Altman analysis for the postoperative course of all eyes revealed a bias of 3.0 mmHg between RT and SIT on the day of surgery and no bias between RT and GAT in the further postoperative follow-up.
CONCLUSION: Rebound tonometry seems to provide precise IOP values after vitreoretinal surgery. Divergence from SIT values on the day of surgery is presumably due to a general tendency of SIT to underestimate IOP values. Therefore, RT can be used in the clinical routine after vitreoretinal surgery as an alternative to GAT.

Entities:  

Keywords:  Endotamponade; Goldmann applanation tonometry; Intraocular pressure; Pars plana vitrectomy; Schiötz indentation tonometry

Mesh:

Year:  2015        PMID: 26070836     DOI: 10.1007/s00347-015-0076-9

Source DB:  PubMed          Journal:  Ophthalmologe        ISSN: 0941-293X            Impact factor:   1.059


  14 in total

1.  Comparison of the iCare rebound tonometer and the Goldmann applanation tonometer over a wide IOP range.

Authors:  S Munkwitz; A Elkarmouty; E M Hoffmann; N Pfeiffer; H Thieme
Journal:  Graefes Arch Clin Exp Ophthalmol       Date:  2008-01-12       Impact factor: 3.117

2.  The iCare rebound tonometer: comparisons with Goldmann tonometry, and influence of central corneal thickness.

Authors:  Ali Poostchi; Robert Mitchell; Simon Nicholas; Gordon Purdie; Anthony Wells
Journal:  Clin Exp Ophthalmol       Date:  2009-09       Impact factor: 4.207

3.  Intraocular pressure measurement after pars plana vitrectomy.

Authors:  L V Del Priore; R G Michels; M A Nunez; W Smiddy; B M Glaser; S de Bustros
Journal:  Ophthalmology       Date:  1989-09       Impact factor: 12.079

4.  Course of intraocular pressure after vitreoretinal surgery: is early postoperative intraocular pressure elevation predictable?

Authors:  Philipp S Muether; Robert Hoerster; Bernd Kirchhof; Sascha Fauser
Journal:  Retina       Date:  2011-09       Impact factor: 4.256

5.  Statistical methods for assessing agreement between two methods of clinical measurement.

Authors:  J M Bland; D G Altman
Journal:  Lancet       Date:  1986-02-08       Impact factor: 79.321

6.  A new rebound tonometer for home monitoring of intraocular pressure.

Authors:  André Rosentreter; Kerstin S Jablonski; Anne C Mellein; Stergiani Gaki; Arno Hueber; Thomas S Dietlein
Journal:  Graefes Arch Clin Exp Ophthalmol       Date:  2011-08-14       Impact factor: 3.117

7.  Comparison of rebound tonometry with Goldmann applanation tonometry and correlation with central corneal thickness.

Authors:  M E Iliev; D Goldblum; K Katsoulis; C Amstutz; B Frueh
Journal:  Br J Ophthalmol       Date:  2006-05-03       Impact factor: 4.638

8.  Method for the noninvasive measurement of intraocular pressure in mice.

Authors:  John Danias; Antti I Kontiola; Theodoros Filippopoulos; Thom Mittag
Journal:  Invest Ophthalmol Vis Sci       Date:  2003-03       Impact factor: 4.799

9.  Three portable tonometers, the TGDc-01, the ICARE and the Tonopen XL, compared with each other and with Goldmann applanation tonometry*.

Authors:  Liane H van der Jagt; Nomdo M Jansonius
Journal:  Ophthalmic Physiol Opt       Date:  2005-09       Impact factor: 3.117

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.