Comparison of impedance cardiography and dye dilution method for measuring cardiac output.

W Spiering1, P N van Es, P W de Leeuw.   

Abstract

OBJECTIVE: To assess the degree of agreement between impedance cardiography, using the NCCOM3-R7 device, and the gold standard--the dye dilution method--both under basal conditions and after stimulation of cardiac output. PATIENTS: 35 paired measurements in five healthy male volunteers.
INTERVENTIONS: To obtain higher levels of cardiac output, cardiac performance was stimulated with a dopamine infusion.
RESULTS: In 35 paired measurements, the mean of all the impedance values was higher than that of the dye dilution values, at 10.2 v 7.4 l/min (p < 0.0001). The mean discrepancy between the two methods was 3.3 l/min, and the mean bias -2.9 l/min, with limits of agreement of -9.0 and 3.2 l/min. A change in cardiac output could not adequately be predicted by the NCCOM3-R7. In 20 of 25 measurements obtained during continuous intravenous dopamine infusions there was a rise in dye dilution cardiac output (range 0.2 to 5.9 l/min). Neither the magnitude nor the direction of the change in dye dilution values corresponded with the change measured by impedance cardiography. The mean discrepancy here between the two methods was 1.8 l/min, and the mean bias -0.8 l/min, with limits of agreement of -4.9 and 3.3 l/min.
CONCLUSIONS: In healthy volunteers, impedance cardiography with NCCOM3-R7 is inadequate for assessing cardiac output when compared with the dye dilution method.

Entities:  

Mesh:

Year:  1998        PMID: 9659188      PMCID: PMC1728690          DOI: 10.1136/hrt.79.5.437

Source DB:  PubMed          Journal:  Heart        ISSN: 1355-6037            Impact factor:   5.994


  40 in total

1.  Noninvasive assessment of cardiac output in children using impedance cardiography.

Authors:  D S Braden; L Leatherbury; F A Treiber; W B Strong
Journal:  Am Heart J       Date:  1990-11       Impact factor: 4.749

2.  Bioimpedance versus thermodilution cardiac output measurement: the Bomed NCCOM3 after coronary bypass surgery.

Authors:  A N Thomas; J Ryan; B R Doran; B J Pollard
Journal:  Intensive Care Med       Date:  1991       Impact factor: 17.440

3.  Determination of cardiac output during positive end-expiratory pressure--noninvasive electrical bioimpedance compared with standard thermodilution.

Authors:  G Castor; G Molter; J Helms; I Niedermark; P Altmayer
Journal:  Crit Care Med       Date:  1990-05       Impact factor: 7.598

4.  [Cardiac output monitoring by impedance cardiography in cardiac surgery].

Authors:  H Shimizu; S Seki; A Mizuguchi; H Tsuchida; H Watanabe; A Namiki
Journal:  Masui       Date:  1990-04

5.  Comparison of two impedance cardiographic techniques for measuring cardiac output in critically ill patients.

Authors:  R W Gotshall; V C Wood; D S Miles
Journal:  Crit Care Med       Date:  1989-08       Impact factor: 7.598

6.  Automated non-invasive measurement of cardiac output: comparison of electrical bioimpedance and carbon dioxide rebreathing techniques.

Authors:  S A Smith; A E Russell; M J West; J Chalmers
Journal:  Br Heart J       Date:  1988-03

Review 7.  A review of impedance cardiography.

Authors:  L Jensen; J Yakimets; K K Teo
Journal:  Heart Lung       Date:  1995 May-Jun       Impact factor: 2.210

8.  The dye dilution method for measurement of cardiac output.

Authors:  P Lund-Johansen
Journal:  Eur Heart J       Date:  1990-12       Impact factor: 29.983

9.  Thoracic electrical bioimpedance measurement of cardiac output in postaortocoronary bypass patients.

Authors:  W S Sageman; D E Amundson
Journal:  Crit Care Med       Date:  1993-08       Impact factor: 7.598

10.  Cardiac output measurement in critical care patients: Thoracic Electrical Bioimpedance versus thermodilution.

Authors:  T V Clancy; K Norman; R Reynolds; D Covington; J G Maxwell
Journal:  J Trauma       Date:  1991-08
View more
  5 in total

1.  A novel approach to measure cardiac output noninvasively: a comparison with the thermodilution method on critical care patients.

Authors:  John Kabal; Bruce K Lagerman
Journal:  J Clin Monit Comput       Date:  2004-06       Impact factor: 2.502

Review 2.  Cardiac output in 1998.

Authors:  M Singer
Journal:  Heart       Date:  1998-05       Impact factor: 5.994

3.  Modulation of left ventricular diastolic filling during exercise in persons with cervical motor incomplete spinal cord injury.

Authors:  Monira I Aldhahi; Andrew A Guccione; Lisa M K Chin; Joshua Woolstenhulme; Randall E Keyser
Journal:  Eur J Appl Physiol       Date:  2019-11-07       Impact factor: 3.078

4.  Comparative study of cardiac output measurement by regional impedance cardiography and thermodilution method in patients undergoing off pump coronary artery bypass graft surgery.

Authors:  Amrita Guha; Dheeraj Arora; Yatin Mehta
Journal:  Ann Card Anaesth       Date:  2022 Jul-Sep

5.  The PhysioFlow thoracic impedancemeter is not valid for the measurements of cardiac hemodynamic parameters in chronic anemic patients.

Authors:  Pascal Bogui; Edwige Balayssac-Siransy; Philippe Connes; Nalourgo Tuo; Soualiho Ouattara; Aurélien Pichon; Cyrille Serges Dah
Journal:  PLoS One       Date:  2013-10-22       Impact factor: 3.240

  5 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.