Determination of cardiac output during positive end-expiratory pressure--noninvasive electrical bioimpedance compared with standard thermodilution.

G Castor1, G Molter, J Helms, I Niedermark, P Altmayer.   

Abstract

Many investigators have demonstrated the accuracy and reliability of thoracic electrical bioimpedance (TEB) in spontaneously breathing patients and under mechanical intermittent positive-pressure ventilation. Most of these studies showed a good correlation between TEB and invasive methods, such as thermodilution (TD) or the Fick method. But during PEEP, contrary results occur when comparing TEB and TD. In six patients undergoing neurosurgical interventions, TEB cardiac output measurements were compared during zero end-expiratory pressure (ZEEP) and during PEEP at 8 cm H2O with a low respiratory rate. The data revealed a good correlation during ZEEP (r = .93) and during PEEP (r = .91). There was no significant statistical difference when measuring cardiac output by TEB during ventilation with PEEP. During normal or decreased cardiac output, TEB overestimated cardiac output compared with TD, whereas TEB underestimated cardiac output compared with TD during increased cardiac output, especially during PEEP.

Entities:  

Mesh:

Year:  1990        PMID: 2183967     DOI: 10.1097/00003246-199005000-00017

Source DB:  PubMed          Journal:  Crit Care Med        ISSN: 0090-3493            Impact factor:   7.598


  6 in total

1.  Comparison of impedance cardiography and dye dilution method for measuring cardiac output.

Authors:  W Spiering; P N van Es; P W de Leeuw
Journal:  Heart       Date:  1998-05       Impact factor: 5.994

2.  A comparison of the haemodynamic effects of intrathecal meperidine, meperidine-bupivacaine mixture and hyperbaric bupivacaine.

Authors:  F Conway; L A Critchley; J C Stuart; R C Freebairn
Journal:  Can J Anaesth       Date:  1996-01       Impact factor: 5.063

3.  Impedance cardiography. Importance of the equation and the electrode configuration.

Authors:  B J van der Meer; H H Woltjer; A M Sousman; W O Schreuder; E R Bulder; M A Huybregts; P M de Vries
Journal:  Intensive Care Med       Date:  1996-10       Impact factor: 17.440

4.  Non-invasive measurement of left ventricular contractility by impedance cardiography.

Authors:  G Castor
Journal:  Med Biol Eng Comput       Date:  1994-07       Impact factor: 2.602

5.  Physiologic and Clinical Principles behind Noninvasive Resuscitation Techniques and Cardiac Output Monitoring.

Authors:  Anthony M Napoli
Journal:  Cardiol Res Pract       Date:  2011-08-16       Impact factor: 1.866

6.  Non-invasive stroke volume estimation by transthoracic electrical bioimpedance versus Doppler echocardiography in healthy volunteers.

Authors:  Mirae Harford; Samuel H Clark; Jodie F Smythe; Stephen Gerry; Mauricio Villarroel; Joao Jorge; Sitthichok Chaichulee; Lionel Tarassenko; Duncan Young; Peter Watkinson
Journal:  J Med Eng Technol       Date:  2019-04-15
  6 in total