Comparison of ultrasonic pachymetry with Orbscan in corneal haze.

Rana Altan-Yaycioglu1, Aysel Pelit, Yonca Aydin Akova.   

Abstract

BACKGROUND: To compare the results of ultrasonic and Orbscan pachymetry in corneal haze not related to surgery.
METHODS: An institutionally based, prospective, controlled clinical trial included 40 eyes of 40 consecutive patients with corneal haze not related to surgery. Fifty normal eyes were used as a control group to calculate the customized acoustic factor. Corneal haze was graded with slit-lamp by a single examiner. Ultrasonic and Orbscan II pachymetry measurements were obtained. Paired t- and Kruskal-Wallis tests were used to evaluate the difference between ultrasonic and Orbscan pachymetry measurements. Correlation of the two methods and haze grade were determined with Kendall's tau-b correlation analysis.
RESULTS: Eighty-five percent of patients had moderate haze. With the customized acoustic factor, Orbscan values were significantly smaller than ultrasonic pachymetry measurements (p = 0.001, paired samplet-test). Increased haze grade resulted in smaller Orbscan pachymetry measurements however, this difference was insignificant (p = 0.444, Kruskal-Wallis test). A positive correlation was found between two methods, though insignificant (p = 0.064, Kendall's tau-b correlation). When grouped in respect of diagnoses, differences in pachymetry measurements as well as the correlation were not significant (p > 0.05).
CONCLUSION: An insignificant positive correlation was found between the two methods in the presence of corneal haze. Orbscan measurements were inversely related to haze grading, and were significantly lower than ultrasonic pachymetry. According to our results, we recommend that clinicians should not count on Orbscan topography alone in the presence of corneal haze.

Entities:  

Mesh:

Year:  2007        PMID: 17583820     DOI: 10.1007/s00417-007-0578-5

Source DB:  PubMed          Journal:  Graefes Arch Clin Exp Ophthalmol        ISSN: 0721-832X            Impact factor:   3.117


  15 in total

1.  Comparison of corneal thickness measurements using ultrasound and Orbscan slit-scanning topography in normal and post-LASIK eyes.

Authors:  H S Chakrabarti; J P Craig; A Brahma; T Y Malik; C N McGhee
Journal:  J Cataract Refract Surg       Date:  2001-11       Impact factor: 3.351

2.  Comparison of corneal pachymetry using ultrasound and Orbscan II.

Authors:  Mohamed A Fakhry; Alberto Artola; José I Belda; Ma José Ayala; Jorge L Alió
Journal:  J Cataract Refract Surg       Date:  2002-02       Impact factor: 3.351

3.  Accuracy of Orbscan pachymetry measurements and DHG ultrasound pachymetry in primary laser in situ keratomileusis and LASIK enhancement procedures.

Authors:  N G Iskander; E Anderson Penno; N T Peters; H V Gimbel; M Ferensowicz
Journal:  J Cataract Refract Surg       Date:  2001-05       Impact factor: 3.351

4.  Reliability of pachymetric measurements using orbscan after excimer refractive surgery.

Authors:  Olivier Prisant; Nila Calderon; Philippe Chastang; Damien Gatinel; Thanh Hoang-Xuan
Journal:  Ophthalmology       Date:  2003-03       Impact factor: 12.079

5.  Central and peripheral corneal thickness measurement with Orbscan II and topographical ultrasound pachymetry.

Authors:  Jose Manuel González-Méijome; Alejandro Cerviño; Eva Yebra-Pimentel; Manuel A Parafita
Journal:  J Cataract Refract Surg       Date:  2003-01       Impact factor: 3.351

6.  Central corneal thickness measurements with partial coherence interferometry, ultrasound, and the Orbscan system.

Authors:  Georg Rainer; Oliver Findl; Vanessa Petternel; Barbara Kiss; Wolfgang Drexler; Christian Skorpik; Michael Georgopoulos; Leopold Schmetterer
Journal:  Ophthalmology       Date:  2004-05       Impact factor: 12.079

7.  Repeatability and reproducibility of central corneal thickness measurement with Pentacam, Orbscan, and ultrasound.

Authors:  Birgit Lackner; Gerald Schmidinger; Stefan Pieh; Martin A Funovics; Christian Skorpik
Journal:  Optom Vis Sci       Date:  2005-10       Impact factor: 1.973

8.  Corneal thickness measurements with the Orbscan Topography System and ultrasonic pachymetry.

Authors:  V Yaylali; S C Kaufman; H W Thompson
Journal:  J Cataract Refract Surg       Date:  1997-11       Impact factor: 3.351

9.  Accuracy of Orbscan II slit-scanning elevation topography.

Authors:  Gerard Cairns; Charles N J McGhee; Michael J Collins; Helen Owens; Greg D Gamble
Journal:  J Cataract Refract Surg       Date:  2002-12       Impact factor: 3.351

10.  Pachymetry assessment with Orbscan II in postoperative patients with myopic LASIK.

Authors:  Arthur C K Cheng; Srinivas K Rao; Emily Tang; Dennis S C Lam
Journal:  J Refract Surg       Date:  2006-04       Impact factor: 3.573

View more
  2 in total

1.  One-year outcomes of conventional and accelerated collagen crosslinking in progressive keratoconus.

Authors:  Vanissa W S Chow; Tommy C Y Chan; Marco Yu; Victoria W Y Wong; Vishal Jhanji
Journal:  Sci Rep       Date:  2015-09-25       Impact factor: 4.379

2.  The relationship of central corneal thickness with the status of diabetic retinopathy.

Authors:  Handan Canan; Nedime Sahinoglu-Keskek; Rana Altan-Yaycioglu
Journal:  BMC Ophthalmol       Date:  2020-06-08       Impact factor: 2.209

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.