Literature DB >> 7612703

Image quality control for digital mammographic systems: initial experience and outlook.

H Roehrig1, T Yu, E Krupinski.   

Abstract

This report presents (1) a broad topical review and a tutorial of the possibilities for image quality control (IQC) with digital systems, and (2) results and initial experience for IQC with two commercial digital imaging systems, but with limited discussion on any particular method. Digital imaging systems used for mammographically guided digital stereotactic breast biopsy were evaluated extensively at the University of Arizona. Measurements were made of linearity, sensitivity, signal-to-noise ratio, and square-wave modulation. Images of phantoms such as the American College of Radiology Accreditation Phantom and the contrast detail mammography Phantom were evaluated as well as images of the x-ray source's focal spot. The evaluation also included the cathode ray tubes for the imaging systems. The data collected show that digital imaging systems have an important advantage over film-screen systems because they provide a digital signal as output that can be used for quantitative analysis. As a result, IQC can become a much more quantitative discipline than presently practiced, providing more information on the imaging systems under evaluation, and providing better control over their properties during actual operation.

Mesh:

Year:  1995        PMID: 7612703     DOI: 10.1007/BF03168128

Source DB:  PubMed          Journal:  J Digit Imaging        ISSN: 0897-1889            Impact factor:   4.056


  19 in total

1.  Performance tests and quality control of cathode ray tube displays.

Authors:  H Roehrig; H Blume; T L Ji; M Browne
Journal:  J Digit Imaging       Date:  1990-08       Impact factor: 4.056

2.  The measurement and routine checking of mammography x-ray tube kV.

Authors:  J Law
Journal:  Phys Med Biol       Date:  1991-08       Impact factor: 3.609

3.  The American College of Radiology Mammography Accreditation Program.

Authors:  R McLelland; R E Hendrick; M D Zinninger; P A Wilcox
Journal:  AJR Am J Roentgenol       Date:  1991-09       Impact factor: 3.959

4.  Standardization of image quality and radiation dose in mammography.

Authors:  R E Hendrick
Journal:  Radiology       Date:  1990-03       Impact factor: 11.105

5.  Signal-to-noise properties of mammographic film-screen systems.

Authors:  R M Nishikawa; M J Yaffe
Journal:  Med Phys       Date:  1985 Jan-Feb       Impact factor: 4.071

6.  Introduction to perceptual linearization of video display systems for medical image presentation.

Authors:  B M Hemminger; R E Johnston; J P Rolland; K E Muller
Journal:  J Digit Imaging       Date:  1995-02       Impact factor: 4.056

7.  Multiformat video and laser cameras: history, design considerations, acceptance testing, and quality control. Report of AAPM Diagnostic X-Ray Imaging Committee Task Group No. 1.

Authors:  J E Gray; W F Anderson; C C Shaw; S J Shepard; L A Zeremba; P J Lin
Journal:  Med Phys       Date:  1993 Mar-Apr       Impact factor: 4.071

8.  The digital radiology department of the future.

Authors:  M P Capp; H Roehrig; G W Seeley; H D Fisher; T W Ovitt
Journal:  Radiol Clin North Am       Date:  1985-06       Impact factor: 2.303

9.  Stripping of X-ray bremsstrahlung spectra up to 300 kVp on a desk type computer.

Authors:  W W Seelentag; W Panzer
Journal:  Phys Med Biol       Date:  1979-07       Impact factor: 3.609

10.  Development and implementation of a quality assurance program for mammography.

Authors:  E S de Paredes; A B Frazier; G D Hartwell; A M Strash; C Scheer; D C Smith; P A Barnette; W Noel; D Kenneweg
Journal:  Radiology       Date:  1987-04       Impact factor: 11.105

View more
  4 in total

1.  Full breast digital mammography with an amorphous silicon-based flat panel detector: physical characteristics of a clinical prototype.

Authors:  S Vedantham; A Karellas; S Suryanarayanan; D Albagli; S Han; E J Tkaczyk; C E Landberg; B Opsahl-Ong; P R Granfors; I Levis; C J D'Orsi; R E Hendrick
Journal:  Med Phys       Date:  2000-03       Impact factor: 4.071

2.  Mammographic imaging with a small format CCD-based digital cassette: physical characteristics of a clinical system.

Authors:  S Vedantham; A Karellas; S Suryanarayanan; I Levis; M Sayag; R Kleehammer; R Heidsieck; C J D'Orsi
Journal:  Med Phys       Date:  2000-08       Impact factor: 4.071

3.  Task-based lens design with application to digital mammography.

Authors:  Liying Chen; Harrison H Barrett
Journal:  J Opt Soc Am A Opt Image Sci Vis       Date:  2005-01       Impact factor: 2.129

4.  Quality assurance and quality control of an intensive care unit picture archiving and communication system.

Authors:  D M Tucker; M McEachern
Journal:  J Digit Imaging       Date:  1995-11       Impact factor: 4.056

  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.