Kusuma Halemani1, Lakshmi Kumar1, Bhadrinath Narayanan1, Sunil Rajan1, Pavithra Ramamurthi1, Abish Sudhakar2.
Abstract
OBJECTIVE: Arterial pulse-derived cardiac output monitors are routinely employed to guide hemodynamic management during liver transplant surgery. In this study, we sought to assess the reliability by evaluating the agreement of the cardiac output measured by the FloTrac Vigileo versus pulmonary artery catheter (continuous cardiac output) at specified times during liver transplant.Entities:
Year: 2022 PMID: 35544253 PMCID: PMC9361166 DOI: 10.5152/TJAR.2021.1356
Source DB: PubMed Journal: Turk J Anaesthesiol Reanim ISSN: 2149-276X
Figure 1.Flow diagram of selection of patients.
Demographics of Patients
| Variable | Mean ± SD |
|---|---|
| Age (years) | 46.62 ± 8.47 |
| Height (cm) | 167.8 ±7.27 |
| Weight (kg) | 73.24 ± 12.15 |
| MELD score | 23.2 ± 5.56 |
| Male: female (n) | 54 : 6 |
| Child–Pugh score “C” (n(%)) | 54 (90%) |
MELD, model for end-stage liver disease; SD, standard deviation.
Bland–Altman Analysis of Cardiac Output and Concordance Correlation Coefficient
| Time Points | n | Bias (L min-1) | SD/Precision | 95% CI of the Difference (L min-1) | % Error (1.96*SD Bias/Mean CO) | Concordance Correlation Coefficient | ||
|---|---|---|---|---|---|---|---|---|
| Lower Limit | Upper Limit | |||||||
| T0 | 60 | 0.2 | 2.09 | −3.9 | 4.3 | 61.03 | 0.65 | |
| T1 | 60 | 0.53 | 3 | −5.4 | 6.4 | 68.08 | 0.39 | |
| T2 | 60 | 0.47 | 2.1 | −3.7 | 4.6 | 50.28 | 0.59 | |
| T3 | 59 | 0.31 | 1.9 | −3.4 | 4.0 | 46.91 | 0.63 | |
| T4 | 59 | 0.44 | 2.15 | −3.8 | 4.7 | 54.15 | 0.43 | |
| T5 | 56 | 0.69 | 1.9 | −2.9 | 4.3 | 44.67 | 0.52 | |
T0, Baseline; T1, 1 hour in dissection phase; T2, anhepatic phase; T3, portosystemic shunt; T4, reperfusion; T5, 1 hour after reperfusion; T6, skin closure.
Figure 2.Bland–Altman graphs comparing CO from CCO and FloTrac. CO, cardiac output; CCO, continuous cardiac output.
Figure 3.Four quadrant plots comparing CO from CCO and FloTrac. CO, cardiac output; CCO, continuous cardiac output.
Hemodynamic Parameters Measured from CCO and FloTrac
| Parameter | T0 | T1 | T2 | T3 | T4 | T5 | T6 | Overall Mean ± SD |
|---|---|---|---|---|---|---|---|---|
| Mean CO by CCO (L min-1) | 8.61 ± 2.7 | 8.93 ± 2.9 | 8.46 ± 2.6 | 8.1 ± 2.4 | 8.01 ± 2.1 | 8.46 ± 2.0 | 8.37 ± 1.9 | 8.42 ± 2.41 |
| Mean CO by FloTrac (L min-1) | 8.41 ± 2.3 | 8.41 ± 2.5 | 8.0 ± 2.19 | 7.79 ± 2.1 | 7.57 ± 1.9 | 7.77 ± 1.9 | 7.94 ± 2.3 | 7.98 ± 2.20 |
| SVR by CCO (dynes s cm-5) | 672.1 ± 243 | 658.7 ± 252 | 635.9 ± 175 | 647.5 ± 174 | 666 ± 190 | 660.5 ± 183 | 621 ± 157 | 653.7 ± 200 |
| SVR by FloTrac (dynes s cm-5) | 646.5 ± 196 | 674 ± 229 | 643.8 ± 168 | 662 ± 187 | 674.9 ± 192 | 675.7 ± 187 | 614.3 ± 167 | 660.4 ± 192 |
| CVP (mmHg) | 14.33 ± 4.2 | 11.71 ± 4 | 11.65 ± 3.8 | 12.18 ± 4.7 | 11.62 ± 4.9 | 12.44 ± 4.4 | 12.67 ± 5.5 | 12.2 ± 4.6 |
| SVV | 6.42 ± 2.6 | 7.07 ± 3 | 6.85 ± 2.8 | 6.56 ± 3.3 | 6.91 ± 3.7 | 6.51 ± 3.57 | 6.16 ± 3.1 | 6.6 ± 3.2 |
| Heart Rate (beats min-1) | 85.4 ± 8.4 | 85.3 ± 8.5 | 95.14 ± 8.7 | 99.1 ± 7.8 | 87.3 ± 10.8 | 86.9 ± 9.3 | 87.7 ± 10.5 | 89.5 ± 5.4 |
| MBP (mm Hg) | 79 ± 12 | 80 ± 13.9 | 75.4 ± 10.1 | 73.1 ± 12.2 | 75.6 ± 9.7 | 77 ± 9.9 | 76.3 ± 11.5 | 76.6 ± 11.5 |
| T0, Baseline; T1, 1 hour in dissection phase; T2, anhepatic phase; T3, portosystemic shunt; T4, reperfusion; T5, 1 hour after reperfusion; T6, skin closure; CO, cardiac output; CCO, continuous cardiac output; SVR, systemic vascular resistance; CVP, central venous pressure; SVV, stroke volume variation; MBP, mean blood pressure. | ||||||||