Cardiac output determination from the arterial pressure wave: clinical testing of a novel algorithm that does not require calibration.

Gerard R Manecke1, William R Auger.   

Abstract

OBJECTIVE: The purpose of this study was to evaluate the accuracy and precision of a novel algorithm that evaluates cardiac output by using arterial pressure waveform characteristics.
DESIGN: Prospective, observational study comparing the cardiac output values of intermittent thermodilution, continuous thermodilution, and continuous arterial pressure wave assessment.
SETTING: The intensive care unit in a tertiary care university hospital. PARTICIPANTS: Fifty postoperative cardiac surgical patients, within the first 12 hours after surgery.
INTERVENTIONS: All patients received a pulmonary artery catheter (PAC) and at least 1 systemic arterial pressure catheter. The data from the arterial catheter were processed by using a new arterial pressure cardiac output (APCO) algorithm. The data from the PAC (continuous and intermittent assessments) were collected for comparison with the APCO. MEASUREMENTS: Two hundred ninety-five cardiac output measurements using intermittent thermodilution (ICO), continuous thermodilution (CCO), and arterial pressure-based output (APCO) were obtained during various times during the first 12 postoperative hours. The measurements of each method at each time point were compared by using Bland-Altman analysis.
RESULTS: The mean cardiac output ranged from 2.77 to 9.60 L/min. APCO, compared with ICO, revealed a bias of 0.55 L/min and precision of 0.98 L/min. APCO, compared with CCO, revealed a bias of 0.06 L/min and precision of 1.06 L/min. The APCO agreement between femoral and radial arterial catheters was close; the bias was -0.15 L/min, and the precision was 0.56 L/min.
CONCLUSIONS: This novel arterial pressure cardiac output algorithm provides cardiac output assessments that agree satisfactorily for clinical purposes with intermittent and continuous thermodilution techniques in postoperative cardiac surgical patients. Further study is required for other patient populations and clinical situations.

Entities:  

Mesh:

Year:  2006        PMID: 17289472     DOI: 10.1053/j.jvca.2006.08.004

Source DB:  PubMed          Journal:  J Cardiothorac Vasc Anesth        ISSN: 1053-0770            Impact factor:   2.628


  29 in total

1.  Comparison of an advanced minimally invasive cardiac output monitoring with a continuous invasive cardiac output monitoring during lung transplantation.

Authors:  Roland Tomasi; Stephan Prueckner; Stephan Czerner; Renè Schramm; Gerhard Preissler; Bernhard Zwißler; Vera von Dossow-Hanfstingl
Journal:  J Clin Monit Comput       Date:  2015-07-30       Impact factor: 2.502

2.  Cardiac index value measurement by invasive, semi-invasive and non invasive techniques: a prospective study in postoperative off pump coronary artery bypass surgery patients.

Authors:  Murali Chakravarthy; Subramanyam Rajeev; Vivek Jawali
Journal:  J Clin Monit Comput       Date:  2009-05-02       Impact factor: 2.502

3.  Non-invasive automated measurement of cardiac output during stable cardiac surgery using a fully integrated differential CO(2) Fick method.

Authors:  Philip J Peyton; Daniel Thompson; Paul Junor
Journal:  J Clin Monit Comput       Date:  2008-07-12       Impact factor: 2.502

Review 4.  Pressure pulse contour-derived stroke volume and cardiac output in the morbidly obese patient.

Authors:  Donald P Bernstein
Journal:  Obes Surg       Date:  2008-04-29       Impact factor: 4.129

5.  Cardiac output monitoring in septic shock: evaluation of the third-generation Flotrac-Vigileo.

Authors:  Sophie Marqué; Antoine Gros; Loic Chimot; Arnaud Gacouin; Sylvain Lavoué; Christophe Camus; Yves Le Tulzo
Journal:  J Clin Monit Comput       Date:  2013-01-30       Impact factor: 2.502

6.  Influence of tidal volume for stroke volume variation to predict fluid responsiveness in patients undergoing one-lung ventilation.

Authors:  Koichi Suehiro; Ryu Okutani
Journal:  J Anesth       Date:  2011-07-09       Impact factor: 2.078

7.  Continuous arterial pressure waveform monitoring in pediatric cardiac transplant, cardiomyopathy and pulmonary hypertension patients.

Authors:  Sarena Teng; Jon Kaufman; Zhaoxing Pan; Angela Czaja; Heather Shockley; Eduardo da Cruz
Journal:  Intensive Care Med       Date:  2011-05-31       Impact factor: 17.440

8.  Accuracy of arterial pressure waveform analysis for cardiac output measurement in comparison with thermodilution methods in patients undergoing living donor liver transplantation.

Authors:  Kozaburo Akiyoshi; Tadashi Kandabashi; Junko Kaji; Ken Yamaura; Hayashi Yoshimura; Kazuo Irita; Sumio Hoka
Journal:  J Anesth       Date:  2011-01-19       Impact factor: 2.078

9.  Intraoperative fluid optimization using stroke volume variation in high risk surgical patients: results of prospective randomized study.

Authors:  Jan Benes; Ivan Chytra; Pavel Altmann; Marek Hluchy; Eduard Kasal; Roman Svitak; Richard Pradl; Martin Stepan
Journal:  Crit Care       Date:  2010-06-16       Impact factor: 9.097

10.  Clinical evaluation of the flotrac/Vigileo system for continuous cardiac output monitoring in patients undergoing regional anesthesia for elective cesarean section: a pilot study.

Authors:  José Otavio Costa Auler; Marcelo L A Torres; Mônica M Cardoso; Thais C Tebaldi; André P Schmidt; Mario M Kondo; Marcelo Zugaib
Journal:  Clinics (Sao Paulo)       Date:  2010-06       Impact factor: 2.365

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.