Dose reduction in orthodontic lateral cephalography: dosimetric evaluation of a novel cephalographic thyroid protector (CTP) and anatomical cranial collimation (ACC).

R C Hoogeveen1, D Rottke, P F van der Stelt, W E R Berkhout.   

Abstract

OBJECTIVES: To test the dose-reducing capabilities of a novel thyroid protection device and a recently introduced cranial collimator to be used in orthodontic lateral cephalography.
METHODS: Cephalographic thyroid protector (CTP) was designed to shield the thyroid while leaving the cervical vertebrae depicted. Using a RANDO(®) head phantom (The Phantom Laboratory, Salem, NY) equipped with dosemeters and a Proline XC (Planmeca, Helsinki, Finland) cephalograph, lateral cephalograms were taken, and the effective dose (ED) was calculated for four protocols: (1) without shielding; (2) with CTP; (3) with CTP and anatomical cranial collimator (ACC); and (4) with a thyroid collar (TC).
RESULTS: The ED for the respective protocols was (1) 8.51; (2) 5.39; (3) 3.50; and (4) 4.97 µSv. The organ dose for the thyroid was reduced from 30.17 to 4.50 µSv in Protocols 2 and 3 and to 3.33 µSv in Protocol 4.
CONCLUSIONS: The use of just the CTP (Protocol 2) resulted in a 36.8% reduction of the ED of a lateral cephalogram. This was comparable to the classical TC (Protocol 4). A 58.8% reduction of the ED was obtained when combining CTP and ACC (Protocol 3). The dose to the radiosensitive thyroid gland was reduced by 85% in Protocols 2 and 3 and by 89% in Protocol 4.

Entities:  

Keywords:  cephalometry; dosimetry; orthodontics; radiation protection; thyroid gland

Mesh:

Year:  2015        PMID: 25564885      PMCID: PMC4628428          DOI: 10.1259/dmfr.20140260

Source DB:  PubMed          Journal:  Dentomaxillofac Radiol        ISSN: 0250-832X            Impact factor:   2.419


  22 in total

Review 1.  Exposing the thyroid to radiation: a review of its current extent, risks, and implications.

Authors:  Bridget Sinnott; Elaine Ron; Arthur B Schneider
Journal:  Endocr Rev       Date:  2010-07-21       Impact factor: 19.871

2.  Utility of thyroid collars in cephalometric radiography.

Authors:  K P Sansare; V Khanna; F Karjodkar
Journal:  Dentomaxillofac Radiol       Date:  2011-12       Impact factor: 2.419

3.  The 2007 Recommendations of the International Commission on Radiological Protection. ICRP publication 103.

Authors: 
Journal:  Ann ICRP       Date:  2007

4.  Dose evaluation of selective collimation effect in cephalography by measurement and Monte Carlo simulation.

Authors:  Boram Lee; Gwisoon Shin; Sunjung Kang; Boram Shin; Ilhong Back; Hyok Park; Changseo Park; Jeongwoo Lee; Wonho Lee; Jonghak Choi; Ryeonghwang Park; Youhyun Kim
Journal:  Radiat Prot Dosimetry       Date:  2011-02-17       Impact factor: 0.972

5.  The influence of lead thyroid collars on cephalometric landmark identification.

Authors:  Dirk Wiechmann; Alain Decker; Ariane Hohoff; Johannes Kleinheinz; Thomas Stamm
Journal:  Oral Surg Oral Med Oral Pathol Oral Radiol Endod       Date:  2007-01-09

6.  Skeletal maturation evaluation using cervical vertebrae.

Authors:  B Hassel; A G Farman
Journal:  Am J Orthod Dentofacial Orthop       Date:  1995-01       Impact factor: 2.650

7.  Thyroid cancer after exposure to external radiation: a pooled analysis of seven studies.

Authors:  E Ron; J H Lubin; R E Shore; K Mabuchi; B Modan; L M Pottern; A B Schneider; M A Tucker; J D Boice
Journal:  Radiat Res       Date:  1995-03       Impact factor: 2.841

8.  Radiation doses of collimated vs non-collimated cephalometric exposures.

Authors:  F Gijbels; G Sanderink; J Wyatt; J Van Dam; B Nowak; R Jacobs
Journal:  Dentomaxillofac Radiol       Date:  2003-03       Impact factor: 2.419

9.  Collimator with filtration compensator: clinical adaptation to meet European Union recommendation 4F on radiological protection for dental radiography.

Authors:  M Alcaraz; M C García-Vera; L A Bravo; Y Martínez-Beneyto; D Armero; J J Morant; M Canteras
Journal:  Dentomaxillofac Radiol       Date:  2009-09       Impact factor: 2.419

10.  Anatomically shaped cranial collimation (ACC) for lateral cephalometric radiography: a technical report.

Authors:  R C Hoogeveen; P F van der Stelt; W E R Berkhout
Journal:  Dentomaxillofac Radiol       Date:  2013-10-29       Impact factor: 2.419

View more
  5 in total

1.  Reducing an already low dental diagnostic X-ray dose: does it make sense? Comparison of three cost-utility analysis methods used to assess two dental dose-reduction measures.

Authors:  R C Hoogeveen; G C H Sanderink; P F van der Stelt; W E R Berkhout
Journal:  Dentomaxillofac Radiol       Date:  2015-07-24       Impact factor: 2.419

2.  Ultrasonographic localization of the thyroid gland for its optimal shielding prior to lateral cephalometric radiography: a pilot study.

Authors:  Elmira Pakbaznejad Esmaeili; Kirsti Hurmerinta; David Rice; Anni Suomalainen
Journal:  Dentomaxillofac Radiol       Date:  2016-02-03       Impact factor: 2.419

3.  European consensus on patient contact shielding.

Authors:  Peter Hiles; Patrick Gilligan; John Damilakis; Eric Briers; Cristian Candela-Juan; Dario Faj; Shane Foley; Guy Frija; Claudio Granata; Hugo de Las Heras Gala; Ruben Pauwels; Marta Sans Merce; Georgios Simantirakis; Eliseo Vano
Journal:  Insights Imaging       Date:  2021-12-23

4.  Assessment of Random Error in Phantom Dosimetry with the Use of Error Simulation in Statistical Software.

Authors:  R C Hoogeveen; E P Martens; P F van der Stelt; W E R Berkhout
Journal:  Biomed Res Int       Date:  2015-12-31       Impact factor: 3.411

5.  Orthodontic radiology: development of a clinical practice guideline.

Authors:  Aldin Kapetanović; Barbara C M Oosterkamp; Antoon A Lamberts; Jan G J H Schols
Journal:  Radiol Med       Date:  2020-05-27       Impact factor: 3.469

  5 in total