Reproducibility of optical quality parameters measured at objective and subjective best focuses in a double-pass system.

Ai-Lian Hu1, Li-Ya Qiao1, Ye Zhang1, Xiao-Gu Cai1, Lei Li1, Xiu-Hua Wan1.   

Abstract

AIM: To evaluate intra-session repeatability and reproducibility of optical quality parameters measured at objective and subjective best focuses in a double-pass system.
METHODS: Thirty Chinese healthy adults (19 to 40 years old) meeting our inclusion criterion were enrolled in the study. After a basic eye examination, two methods of optical quality measurement, based on subjective and objective best focuses were performed using the Optical Quality Analysis System (OQAS) with an artificial pupil diameter of 4.0 mm.
RESULTS: With each method, three consecutive measurements of the following parameters: the modulation transfer function cutoff frequency (MTFcutoff), the Strehl(2D) ratio, the OQAS values (OVs) at contrasts of 100%, 20%, 9% and the objective scatter index (OSI) were performed by an experienced examiner. The repeatability of each method was evaluated by the repeatability limit (RL) and the coefficient of repeatability (COR). Reproducibility of the two methods was evaluated by intra-class correlation coefficient (ICC) and the 95% limits of agreement (Bland and Altman analysis). Thirty subjects, seven females and twenty three males, of whom 15 right eyes and 15 left eyes were selected randomly for recruitment in the study. The RLs (percentage) for the six parameters measured at objective focus and subjective focus ranged from 8.44% to 15.13% and 10.85% to 16.26%, respectively. The CORs for the two measurement methods ranged from 8.27% to 14.83% and 10.63% to 15.93%, respectively. With regard to reproducibility, the ICCs for the six parameters of OQAS ranged from 0.024 to 0.276. The 95% limits of agreement obtained for the six parameters (in comparison of the two methods) ranged from -0.57 to 42.18 (MTFcutoff), -0.01 to 0.23 (Strehl(2D) ratio), -0.02 to 1.40 (OV100%), -0.10 to 1.75 (OV20%), -0.14 to 1.80 (OV9%) and -1.46 to 0.18 (OSI).
CONCLUSION: Measurements provided by OQAS with either method showed a good repeatability. However, the results obtained from the two different measurement methods showed a poor reproducibility. These findings suggest that it might be best to evaluate patients' optical quality by OQAS using the best focus as chosen automatically by the instrument.

Entities:  

Keywords:  best focus; double-pass system; optical quality; repeatability; reproducibility

Year:  2015        PMID: 26558224      PMCID: PMC4630992          DOI: 10.3980/j.issn.2222-3959.2015.05.34

Source DB:  PubMed          Journal:  Int J Ophthalmol        ISSN: 2222-3959            Impact factor:   1.779


  26 in total

1.  Imperfect optics may be the eye's defence against chromatic blur.

Authors:  James S McLellan; Susana Marcos; Pedro M Prieto; Stephen A Burns
Journal:  Nature       Date:  2002-05-09       Impact factor: 49.962

2.  Axial chromatic aberration of the human eye.

Authors:  R E BEDFORD; G WYSZECKI
Journal:  J Opt Soc Am       Date:  1957-06

3.  Repeatability of measurements with a double-pass system.

Authors:  Alain Saad; Marc Saab; Damien Gatinel
Journal:  J Cataract Refract Surg       Date:  2010-01       Impact factor: 3.351

4.  Using the Bland-Altman method to measure agreement with repeated measures.

Authors:  P S Myles; J Cui
Journal:  Br J Anaesth       Date:  2007-09       Impact factor: 9.166

5.  Repeatability of corneal thickness measured using an Oculus Pentacam.

Authors:  Marco A Miranda; Hema Radhakrishnan; Clare O'Donnell
Journal:  Optom Vis Sci       Date:  2009-03       Impact factor: 1.973

6.  Optical and visual performance of diffractive intraocular lens implantation after myopic laser in situ keratomileusis.

Authors:  Luis Fernández-Vega; David Madrid-Costa; José F Alfonso; Robert Montés-Micó; Arancha Poo-López
Journal:  J Cataract Refract Surg       Date:  2009-05       Impact factor: 3.351

7.  Comparison of the retinal image quality with a Hartmann-Shack wavefront sensor and a double-pass instrument.

Authors:  Fernando Díaz-Doutón; Antonio Benito; Jaume Pujol; Montserrat Arjona; José Luis Güell; Pablo Artal
Journal:  Invest Ophthalmol Vis Sci       Date:  2006-04       Impact factor: 4.799

8.  A new approach to the study of ocular chromatic aberrations.

Authors:  S Marcos; S A Burns; E Moreno-Barriusop; R Navarro
Journal:  Vision Res       Date:  1999-10       Impact factor: 1.886

9.  Objective optical assessment of tear-film quality dynamics in normal and mildly symptomatic dry eyes.

Authors:  Antonio Benito; Guillermo M Pérez; Sandra Mirabet; Meritxell Vilaseca; Jaume Pujol; José M Marín; Pablo Artal
Journal:  J Cataract Refract Surg       Date:  2011-08       Impact factor: 3.351

10.  Optical quality one month after verisyse and Veriflex phakic IOL implantation and Zeiss MEL 80 LASIK for myopia from 5.00 to 16.50 diopters.

Authors:  Meritxell Vilaseca; Adenay Padilla; Jaume Pujol; Juan C Ondategui; Pablo Artal; Jose L Güell
Journal:  J Refract Surg       Date:  2009-08       Impact factor: 3.573

View more
  8 in total

1.  Evaluation of visual quality of spherical and aspherical intraocular lenses by Optical Quality Analysis System.

Authors:  Yan Chen; Xue Wang; Chuan-Di Zhou; Qiang Wu
Journal:  Int J Ophthalmol       Date:  2017-06-18       Impact factor: 1.779

2.  Tear Film Characteristics During Wear of Daily Disposable Contact Lenses.

Authors:  Giancarlo Montani; Maurizio Martino
Journal:  Clin Ophthalmol       Date:  2020-06-04

3.  Objective assessment of optical quality in dry eye disease using a double-pass imaging system.

Authors:  Larissa Gouvea; George Oral Waring; Ashley Brundrett; Michelle Crouse; Karolinne Maia Rocha
Journal:  Clin Ophthalmol       Date:  2019-10-09

4.  Utility of the optical quality analysis system for decision-making in Nd: YAG laser posterior capsulotomy in patients with light posterior capsule opacity.

Authors:  Bo Lu; Weijie Zhu; Yu Fan; Dong Shi; Liwei Ma
Journal:  BMC Ophthalmol       Date:  2021-01-06       Impact factor: 2.209

5.  Study of tear film optics and its impact on quality of vision.

Authors:  Sharon D'Souza; Sriram Annavajjhala; Prashansa Thakur; Ritika Mullick; S J Tejal; Naren Shetty
Journal:  Indian J Ophthalmol       Date:  2020-12       Impact factor: 1.848

6.  Impact of Dry Eye Disease on Vision Quality: An Optical Quality Analysis System Study.

Authors:  Antoine Herbaut; Hong Liang; Ghislaine Rabut; Liem Trinh; Karima Kessal; Christophe Baudouin; Antoine Labbé
Journal:  Transl Vis Sci Technol       Date:  2018-07-12       Impact factor: 3.283

7.  Using Optical Quality Analysis System for predicting surgical parameters in age-related cataract patients.

Authors:  Thibaud Garcin; Damien Grivet; Gilles Thuret; Philippe Gain
Journal:  PLoS One       Date:  2020-10-12       Impact factor: 3.240

8.  Comparison Between an Intraocular Lens With Extended Depth of Focus (Tecnis Symfony ZXR00) and a New Monofocal Intraocular Lens With Enhanced Intermediate Vision (Tecnis Eyhance ICB00).

Authors:  Young Joon Jeon; Yisang Yoon; Tae-Im Kim; Kyungmin Koh
Journal:  Asia Pac J Ophthalmol (Phila)       Date:  2021-10-01
  8 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.