| Literature DB >> 34568748 |
Piotr Kanclerz1, Katarzyna Przewłócka2, Xiaogang Wang3.
Abstract
INTRODUCTION: The measurements of corneal white-to-white (WTW) diameter and pupil size are critical for decision making in refractive surgery. Currently, automatic measurement of keratometry, corneal WTW, and pupil size are implemented in several ocular devices. The purpose of this study was to examine the agreement between two commonly used devices, an autorefractor and an optical biometer, for these parameters.Entities:
Keywords: automated measurements; corneal size; corneal white-to-white; intraocular lens calculation; pupil size
Year: 2021 PMID: 34568748 PMCID: PMC8458669 DOI: 10.1177/25158414211045750
Source DB: PubMed Journal: Ther Adv Ophthalmol ISSN: 2515-8414
Mean anterior eye distances (expressed as mean ± standard deviation) obtained with Nidek ARK-1 and Lenstar LS-900 systems.
| Device | Result | Mean difference ± SD | 95% limits of agreement | Correlation coefficient | |
|---|---|---|---|---|---|
| Flat keratometry (D) | ARK-1 | 43.36 ± 1.80 | –0.02 ± 0.24 | –0.49 to 0.46 | 0.9911 (0.96) |
| LS–900 | 43.38 ± 1.78 | ||||
| Steep keratometry (D) | ARK–1 | 44.40 ± 1.92 | –0.04 ± 0.29 | –0.61 to 0.53 | 0.9889 (0.90) |
| LS–900 | 44.44 ± 1.88 | ||||
| Mean keratometry (D) | ARK–1 | 43.88 ± 1.82 | –0.03 ± 0.21 | –0.44 to 0.39 | 0.9933 (0.93) |
| LS–900 | 43.91 ± 1.78 | ||||
| Corneal white-to-white (mm) | ARK–1 | 12.01 ± 0.44 | –0.23 ± 0.28 | –0.78 to 0.32 | 0.8071 (<0.01) |
| LS–900 | 12.19 ± 0.42 | ||||
| Pupil diameter (mm) | ARK–1 | 5.15 ± 0.95 | 0.90 ± 0.87 | –0.80 to 2.60 | 0.4890 (<0.01) |
| LS-900 | 4.38 ± 0.79 |
Figure 1.Agreement in measurements of corneal white-to-white between ARK-1 and Lenstar LS-900 presented in Bland-Altman plot.
Figure 2.Agreement in measurements of pupil size between ARK-1 and Lenstar LS-900 presented in Bland-Altman plot.
Differences in automated methods for corneal WTW assessment, presented as mean ± SD, 95% LOA.
| Study | Devices | Mean difference in WTW ± SD (mm) | 95% LOA (mm) |
|---|---|---|---|
| Fernández | Keratograph 5M vs. Orbscan IIz | 0.01 ± 0.19 | –0.18; 0.20 |
| Chan | IOLMaster700 vs. AL-Scan | 0.28 ± 0.3[ | –0.31; 0.88 |
| Ferrer-Blasco | IOLMaser700 vs. Atlas 9000 | –0.14 ± 0.17[ | –0.47; 0.18 |
| IOLMaster 700 vs. Sirius | 0.28 ± 0.19[ | –0.10; 0.65 | |
| Muzyka-Woźniak | IOLMaster 500 vs. Pentacam AXL | 0.4 ± 0.2[ | 0.3; 0.5 |
| Salouti | Galilei vs. Orbscan | 0.38 ± 0.56[ | –0.75; 1.48 |
| Galilei vs. EyeSys | 0.05 ± 0.75[ | –1.52; 1.42 | |
| Buckhurst | IOLMaster 500 vs. Lenstar | 0.06 ± 0.17 | –0.25; 0.39 |
| Baumeister | IOLMaster vs. Orbscan | 0.24 ± 0.42 | –0.61; 1.08 |
LOA, limits of agreement; WTW, white to white.
Difference statistically significant.