Repeatability and Agreement of a Swept-Source Optical Coherence Tomography-Based Biometer IOLMaster 700 Versus a Scheimpflug Imaging-Based Biometer AL-Scan in Cataract Patients.

Tommy C Y Chan1, Kelvin H Wan, Fang Yao Tang, Yu Meng Wang, Marco Yu, Carol Cheung.   

Abstract

PURPOSE: To compare the repeatability and agreement between a swept-source biometer and a Scheimpflug biometer in cataract patients.
METHODS: Three consecutive measurements were obtained using a swept-source biometer (IOLMaster 700) and a Scheimpflug biometer (AL-Scan) in 52 eyes of 52 patients. Keratometry, central corneal thickness (CCT), anterior chamber depth (ACD), axial length, and white-to-white (WTW) distance were recorded. Astigmatism values were transformed into vector components of J0 and J45. Intraoperator repeatability was analyzed using intraclass correlation coefficients (ICCs) and reproducibility coefficients (RCs). Agreement of measurements between the two devices was evaluated using the Bland-Altman method.
RESULTS: The IOLMaster 700 showed higher ICCs and lower RCs for the mean keratometry (Km) (P≤0.018), CCT (P≤0.027), and ACD (P≤0.001) measurements, whereas the AL-Scan showed higher ICC and lower RC for the J45 vector component of astigmatism at the 2.4-mm zone (P≤0.034). Both the devices had excellent repeatability (ICC=0.999) in axial length measurement. Systematic differences were found in Km, CCT, ACD, and WTW (P≤0.018) between the devices. The mean difference for Km was -0.196 and -0.144 D measured at the 2.4-mm zone and 3.3-mm zone, respectively. The corresponding mean difference for CCT, ACD, and WTW distance was 14.92 μm, -0.017 mm, and 0.283 mm, respectively. These differences led to a statistically significant but clinically insignificant difference in the prediction of intraocular lens power.
CONCLUSIONS: This study showed significant differences in anterior segment measurement repeatability and agreement between a swept-source biometer and a Scheimpflug biometer in eyes with cataract.

Entities:  

Mesh:

Year:  2020        PMID: 30985487     DOI: 10.1097/ICL.0000000000000603

Source DB:  PubMed          Journal:  Eye Contact Lens        ISSN: 1542-2321            Impact factor:   2.018


  4 in total

1.  Comparison of Vault Measurements Using a Swept-Source OCT-Based Optical Biometer and Anterior Segment OCT.

Authors:  Yali Du; Chuang Jin; Shengjie Yin; Geng Wang; Qian Ma; Yuancun Li; Binyao Chen; Hongxi Wang; Kunliang Qiu; Mingzhi Zhang
Journal:  Front Med (Lausanne)       Date:  2022-06-23

2.  Effect of Implantable Collamer Lens on Anterior Segment Measurement and Intraocular Lens Power Calculation Based on IOLMaster 700 and Sirius.

Authors:  Xiaoyong Chen; Di Zhang; Ziyuan Liu; Yinan Liu; Hongyuan Cai; Qianru Wu; Yu Zhang
Journal:  J Ophthalmol       Date:  2021-12-21       Impact factor: 1.909

3.  Inter-device measurement variability of vital data parameters for keratorefractive and cataract refractive surgery.

Authors:  Piotr Kanclerz; Katarzyna Przewłócka; Xiaogang Wang
Journal:  Ther Adv Ophthalmol       Date:  2021-09-20

4.  Evaluation of 6 biometers based on different optical technologies.

Authors:  Robert Montés-Micó
Journal:  J Cataract Refract Surg       Date:  2022-01-01       Impact factor: 3.351

  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.