Comparison of horizontal corneal diameter measurements using Galilei, EyeSys and Orbscan II systems.

Ramin Salouti1, Mohammad H Nowroozzadeh, Mohammad Zamani, Maryam Ghoreyshi, Rosa Salouti.   

Abstract

BACKGROUND: White-to-white corneal measurement is used frequently in cataract and refractive surgery as well as in diagnosing various corneal diseases and monitoring congenital glaucoma. In this article, we determine agreement in horizontal corneal diameter measurements in normal eyes among the Galilei V4.01, the EyeSys and the Orbscan IIz.
METHODS: In this prospective study, 37 consecutive volunteers were enrolled. All 74 eyes were measured with all three devices. Galilei, EyeSys and Orbscan II measurements were compared with one another. In addition, the data from the right and left eyes were compared for each device.
RESULTS: The mean horizontal corneal diameter measurements with the Galilei, the EyeSys and the Orbscan II were 12.01 +/- 0.61 mm, 12.09 +/- 0.87 mm and 11.67 +/- 0.29 mm, respectively. The difference between the Galilei and EyeSys systems was not statistically significant (p = 0.76), however, the Orbscan II showed smaller measurements (p < 0.001). The best 95% limits of agreement between devices were for the Galilei and the Orbscan II (-0.72, 1.48; r = 0.4, p < 0.01). The best 95% limits of agreement between two eyes for each device were found with the Orbscan II (-0.15, 0.17; r = 0.996, p < 0.01). There was no correlation between measurements of right and left eyes for the Galilei or the EyeSys (r = 0.274, p = 0.176 and r = 0.31, p = 0.065, respectively).
CONCLUSIONS: These results suggest that measurements made with the Orbscan II are smaller than those obtained with the EyeSys Corneal Analysis system and the Galilei. Among the three devices, the Galilei and the Orbscan II showed the best agreement, however, it is inadvisable to use the three devices interchangeably in every clinical situation.

Entities:  

Mesh:

Year:  2009        PMID: 19681922     DOI: 10.1111/j.1444-0938.2009.00407.x

Source DB:  PubMed          Journal:  Clin Exp Optom        ISSN: 0816-4622            Impact factor:   2.742


  16 in total

1.  Corneal white-to-white distance and mesopic pupil diameter.

Authors:  Hasan Basri Cakmak; Nurullah Cagil; Huseyin Simavli; Sabri Raza
Journal:  Int J Ophthalmol       Date:  2012-08-18       Impact factor: 1.779

2.  White-to-white corneal diameter: normal values in healthy Iranian population obtained with the Orbscan II.

Authors:  Hamid Gharaee; Mojtaba Abrishami; Masoud Shafiee; Asieh Ehsaei
Journal:  Int J Ophthalmol       Date:  2014-04-18       Impact factor: 1.779

Review 3.  Device interchangeability on anterior chamber depth and white-to-white measurements: a thorough literature review.

Authors:  Alberto Domínguez-Vicent; Cari Pérez-Vives; Teresa Ferrer-Blasco; Santiago García-Lázaro; Robert Montés-Micó
Journal:  Int J Ophthalmol       Date:  2016-07-18       Impact factor: 1.779

4.  Analysis of blink dynamics in patients with blepharoptosis.

Authors:  Felix H W Mak; Anthony Harker; Kyung-Ah Kwon; Mohan Edirisinghe; Geoffrey E Rose; Fabiola Murta; Daniel G Ezra
Journal:  J R Soc Interface       Date:  2016-03       Impact factor: 4.118

5.  Analysis of Age, Gender, and Refractive Error-Related Changes of the Anterior Corneal Surface Parameters Using Oculus Keratograph Topography.

Authors:  Nasrin Moghadas Sharif; Negareh Yazdani; Leila Shahkarami; Hadi Ostadi Moghaddam; Asieh Ehsaei
Journal:  J Curr Ophthalmol       Date:  2020-07-04

6.  Comparison of anterior segment measurements obtained by three different devices in healthy eyes.

Authors:  Carmen Lopez de la Fuente; Ana Sanchez-Cano; Francisco Segura; Isabel Pinilla
Journal:  Biomed Res Int       Date:  2014-06-02       Impact factor: 3.411

7.  White-to-white corneal diameter distribution in an adult population.

Authors:  Hassan Hashemi; Mehdi Khabazkhoob; Mohammad Hassan Emamian; Mohammad Shariati; Abbasali Yekta; Akbar Fotouhi
Journal:  J Curr Ophthalmol       Date:  2015-10-19

8.  Periorbital Biometric Measurements using ImageJ Software: Standardisation of Technique and Assessment Of Intra- and Interobserver Variability.

Authors:  R Rajyalakshmi; Winston D Prakash; Mohammad Javed Ali; Milind N Naik
Journal:  J Cutan Aesthet Surg       Date:  2017 Jul-Sep

9.  Three-dimensional non-parametric method for limbus detection.

Authors:  Ahmed Abass; Bernardo T Lopes; Ashkan Eliasy; Richard Wu; Steve Jones; John Clamp; Renato Ambrósio; Ahmed Elsheikh
Journal:  PLoS One       Date:  2018-11-26       Impact factor: 3.240

10.  Confounding sizing in posterior chamber phakic lens selection due to white-to-white measurement bias.

Authors:  JoaquIn Fernández; Manuel Rodríguez-Vallejo; Javier Martínez; Ana Tauste; Elisa Hueso; David P Piñero
Journal:  Indian J Ophthalmol       Date:  2019-03       Impact factor: 1.848

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.