Agreement of white-to-white measurements with the IOLMaster 700, Atlas 9000, and Sirius systems.

Teresa Ferrer-Blasco1, José J Esteve-Taboada1, Noelia Martínez-Albert1, José F Alfonso2,3, Robert Montés-Micó1.   

Abstract

BACKGROUND: Current advances in ocular surgery require the evaluation of differences in automated white-to-white (WTW) measurements between systems. We aim to analyze three different commonly used devices in clinical practice: the IOLMaster 700, the Atlas 9000, and the Sirius.
METHODS: Ninety right eyes of ninety healthy subjects were included. Measurements of WTW distances were taken with each device for all subjects in the same session and by the same observer. Statistical comparison was carried out by means of the rANOVA procedure and the Bonferroni post-hoc test. In addition, the Bland-Altman analysis was applied in order to perform pairwise comparisons.
RESULTS: Average WTW values obtained by the IOLMaster 700, the Atlas 9000, and the Sirius devices were 12.18 ± 0.40-mm, 12.32 ± 0.37-mm and 11.90 ± 0.37-mm, respectively. There were statistically significant differences in all pairwise comparisons. Mean differences revealed that the Atlas 9000 provided the highest WTW values, followed by the IOLMaster 700 values and finally by the Sirius results, which had the lowest values. The limits of agreement obtained in all pairwise comparisons were wide, with a range between 0.64 and 0.75 mm.
CONCLUSIONS: According to our results, these three devices should not be used interchangeably for WTW diameter measurements during the clinical practice.

Keywords:  Atlas 9000; IOLMaster 700; Sirius; White-to-White; agreement studies

Mesh:

Year:  2018        PMID: 29806949     DOI: 10.1080/17434440.2018.1481745

Source DB:  PubMed          Journal:  Expert Rev Med Devices        ISSN: 1743-4440            Impact factor:   3.166


  4 in total

1.  Comparison of Vault Measurements Using a Swept-Source OCT-Based Optical Biometer and Anterior Segment OCT.

Authors:  Yali Du; Chuang Jin; Shengjie Yin; Geng Wang; Qian Ma; Yuancun Li; Binyao Chen; Hongxi Wang; Kunliang Qiu; Mingzhi Zhang
Journal:  Front Med (Lausanne)       Date:  2022-06-23

2.  Inter-device measurement variability of vital data parameters for keratorefractive and cataract refractive surgery.

Authors:  Piotr Kanclerz; Katarzyna Przewłócka; Xiaogang Wang
Journal:  Ther Adv Ophthalmol       Date:  2021-09-20

3.  Agreement of Anterior Segment Parameter Measurements With CASIA 2 and IOLMaster 700.

Authors:  Xiaoting Ruan; GuangYao Yang; Zhaoxia Xia; Jiaqing Zhang; Xiaoxun Gu; Yuan Tan; Zhenzhen Liu; Lixia Luo
Journal:  Front Med (Lausanne)       Date:  2022-02-11

4.  Daily Variations of Corneal White-to-White Diameter Measured with Different Methods.

Authors:  Stefan Bandlitz; Makram Nakhoul; Konstantin Kotliar
Journal:  Clin Optom (Auckl)       Date:  2022-09-20