Out-of-field dose in stereotactic radiotherapy for paediatric patients.

Lachlan Garrett1,2, Nicholas Hardcastle2,3, Adam Yeo2, Peta Lonski2, Rick Franich1, Tomas Kron1,2,3.   

Abstract

BACKGROUND AND
PURPOSE: Stereotactic radiotherapy combines image guidance and high precision delivery with small fields to deliver high doses per fraction in short treatment courses. In preparation for extension of these treatment techniques to paediatric patients we characterised and compared doses out-of-field in a paediatric anthropomorphic phantom for small flattened and flattening filter free (FFF) photon beams. METHOD AND MATERIALS: Dose measurements were taken in several organs and structures outside the primary field in an anthropomorphic phantom of a 5 year old child (CIRS) using thermoluminescence dosimetry (LiF:Mg,Cu,P). Out-of-field doses from a medical linear accelerator were assessed for 6 MV flattened and FFF beams of field sizes between 2 × 2 and 10 × 10 cm2.
RESULTS: FFF beams resulted in reduced out-of-field doses for all field sizes when compared to flattened beams. Doses for FFF and flattened beams converged for all field sizes at larger distances (>40 cm) from the central axis as leakage becomes the primary source of out-of-field dose. Rotating the collimator to place the MLC bank in the longitudinal axis of the patient was shown to reduce the peripheral doses measured by up to 50% in Varian linear accelerators.
CONCLUSION: Minimising out-of-field doses by using FFF beams and aligning the couch and collimator to provide tertiary shielding demonstrated advantages of small field, FFF treatments in a paediatric setting.
© 2021 The Authors. Published by Elsevier B.V. on behalf of European Society of Radiotherapy & Oncology.

Entities:  

Keywords:  High dose rate; Out-of-field dose; Paediatric; Peripheral dose; SABR

Year:  2021        PMID: 34307913      PMCID: PMC8295843          DOI: 10.1016/j.phro.2021.05.006

Source DB:  PubMed          Journal:  Phys Imaging Radiat Oncol        ISSN: 2405-6316


Introduction

Globally, cancer is one of the leading causes of death in children aged 0–19 years old, with approximately 300,000 new cases diagnosed each year [1]. The use of radiotherapy in children has fluctuated over the years and at present up to one third of children receive radiation therapy as a part of their cancer treatment [2]. While historically stereotactic radiotherapy procedures were mostly reserved for arterio-venous malformations (AVM) [3] there is an increasing interest to also extend the use of stereotactic ablative body radiotherapy (SABR) to childhood cancers [4], [5]. Out-of-field radiation dose is of particular concern in the context of children who are more prone to radiation induced secondary cancers [6], a fact that has been a major driver for the interest in proton radiotherapy [7]. While areas surrounding the primary beam receive the majority of out-of-field dose, there are structures that receive dose from internal scatter and leakage beyond regions where the treatment planning system (TPS) can calculate dose accurately [8], [9]. In the context of high precision radiotherapy which relies on image guidance such as SABR this effect is compounded by the associated imaging dose [10]. Out-of-field doses from linear accelerators in radiotherapy have been studied by many authors with several reviews providing summaries [11], [12], [13]. However, there is little published specifically on paediatric treatment scenarios with most of the papers specific to clinical treatment scenarios such as for brain lesions [14], [15] including an EURADOS study on out-of-field doses from IMRT and Gammaknife in AVMs [16]. We are aiming to add to this by specifically addressing out-of-field dose to paediatric patients from small fields such as that which may be used in SABR treatments. Of particular interest in this context is the use of flattening filter free (FFF) beams [17], [18]. Therefore, the aim of this study was to measure out-of-field dose in an anthropomorphic paediatric phantom for different target locations and small field sizes, comparing flattened and FFF beams.

Method and materials

Phantom and dosimetry

Thermoluminescence dosimetry measurements were conducted using a phantom of a five-year-old paediatric patient (CIRS, Norfolk, USA) made from tissue equivalent materials including bone, soft tissue, and lung. The phantom featured 5 mm diameter holes spaced in a 3 × 3 cm2 grid in each phantom slice for dosimetry using Thermoluminescence Dosimeters (TLDs). Individually calibrated LiF:Mg,Cu,P TLD-100H chips (3.1 × 3.1 × 0.9 mm3, Harshaw, Ohio, USA) were used due to their high sensitivity and wide dosimetric range [19]. A Harshaw 5500 automatic TLD reader was used for readout using nitrogen gas heating, a pre-read anneal of 10 s at 150 °C and a readout temperature of 250 °C. All TLDs were annealed after each use for 30 min at 240 °C. For the measurements, TLDs were placed in the anatomical locations of the brain, thyroid, heart, lung, abdomen, and gonads as indicated in Table 1. The TLDs were placed in the same coronal and sagittal planes with the exception of the heart and (right) lung which were placed off centre to correspond to organ location.
Table 1

Thermoluminescence dosimeter locations (organs) and distances from the central axis for all four radiotherapy field sites.

Distance from the field centre (cm)
Measurement locationBrainNeckThoraxAbdomen
Brain15.225.239.9
Thyroid15.210.224.7
Heart25.09.92.714.8
Lung25.911.89.116.1
Abdomen39.924.715.1
Gonads52.537.327.612.6
Thermoluminescence dosimeter locations (organs) and distances from the central axis for all four radiotherapy field sites. The reproducibility of the each TLD measurement in reference conditions was better than 2% at 95% confidence interval (CI). Using five TLDs from the same batch as standards irradiated to a known dose (typically 100 mGy) at depth of maximum dose in a 10 × 10 cm2 field of 6 MV X-rays and two TLDs per measurement point resulted in an overall uncertainty of ± 5% at 95% CI for a given dose readout [20], [21].

Irradiation

Out-of-field measurements were taken in the paediatric phantom at six locations as listed in Table 1. The phantom was set up in the head-first supine position and four different field sites were irradiated: brain, neck, thorax, and abdomen. The field sites and measurement locations are listed in Table 1 and illustrated in Fig. 1. All irradiations were performed with a Varian TrueBeam STx (Varian, Palo Alto, USA) with an HD120 MLC. For each field location, beams were delivered with jaw-defined 10 × 10 cm2, 5 × 5 cm2 (MLC retracted), and MLC-defined 2 × 2 cm2 field sizes. For the 2 × 2 cm2 field the jaws were set to 3 × 3 cm2 which reflects typical commissioning conditions for the Varian Eclipse treatment planning system used in our institution. The largest field size was used as reference. The collimator rotation was 90° except where otherwise specified. To quantify the additional shielding provided by the MLC bank, additional measurements were performed for 5 × 5 cm2 abdominal fields with the collimator oriented at both 0° and 90°. All exposures were delivered with 100 MU, operated at 6 MV and 6 MV Flattening Filter Free (FFF). TPR2010 for the two beams was 0.6678 and 0.6315, respectively.
Fig. 1

Schematic representing the four different field locations used in the present study: (a) brain, (b) neck, (c) thorax, and (d) abdomen. Shown are the 2 × 2 cm2 fields. Additional field sizes of 5 × 5 cm2 and 10 × 10 cm2 were also used. TLDs were placed at each labelled anatomical location. (e) shows two measurement points at equal distances from the beam (i) and (ii) but with different radiological path lengths due to inhomogeneities.

Schematic representing the four different field locations used in the present study: (a) brain, (b) neck, (c) thorax, and (d) abdomen. Shown are the 2 × 2 cm2 fields. Additional field sizes of 5 × 5 cm2 and 10 × 10 cm2 were also used. TLDs were placed at each labelled anatomical location. (e) shows two measurement points at equal distances from the beam (i) and (ii) but with different radiological path lengths due to inhomogeneities.

Results

Measured out-of-field dose reduced with distance from the field centre as shown in Fig. 2 for two of the three field sizes for a 6 MV X-ray flattened beam and a 6 MV FFF beam. As can be seen in Fig. 1 there were circumstances where field position and detector locations were exchanged (for example, dose from abdominal field to thyroid and dose from thyroid field to abdomen). This should lead to similar doses in both treatment scenarios as the radiological distance between field centre and TLD position is similar even if scatter conditions vary. In the six exchangeable locations probed over 3 field sizes and for both flattened and FFF beams the average ratio of dose received in the two scenarios where field centre and TLD position were exchanged was 1.12 ± 0.39 (1 SD).
Fig. 2

Out-of-field doses for two field sizes using a 6MV X-ray beam in flattened and flattening filter free (FFF) mode as a function of the distance from the field centre.

Out-of-field doses for two field sizes using a 6MV X-ray beam in flattened and flattening filter free (FFF) mode as a function of the distance from the field centre. The difference between out-of-field dose for the flattened beam and the FFF beam of the same size was found to increase with distance from the field. The variation in out-of-field dose for similar distances from the field centre in the same beam as shown in Fig. 2 can be attributed to the use of an anthropomorphic phantom. This was shown by comparing dose in locations where lung was located between field and measurement location, to dose where the direct path was through tissue only (as illustrated in Fig. 1). Fig. 3 shows this for a 2 × 2 cm2 field. A tendency was found that dose behind lung was larger than behind tissue.
Fig. 3

Comparison of out-of-field dose behind lung and tissue for a 2 × 2 cm2 field of 6MV X-rays with and without flattening filter. The out-of-field doses in the FFF beams were lower than for the flattened beams shown in the left panel.

Comparison of out-of-field dose behind lung and tissue for a 2 × 2 cm2 field of 6MV X-rays with and without flattening filter. The out-of-field doses in the FFF beams were lower than for the flattened beams shown in the left panel. In the smallest field the dose was found to decrease faster with distance from the field in an FFF beam compared to a flattened beam. This is shown in Fig. 4, which shows the ratio of out-of-field doses from FFF beams to flattened beams for all 3 field sizes as a function of distance from the central axis. For the 5 × 5 cm2 and 10 × 10 cm2 fields, the ratio of doses between FFF and flattened fields was found to be largely independent of distance. The ratio was found to be below 1 as the overall number of photons delivered is smaller for FFF beams when both fields are normalised at central axis.
Fig. 4

Variation of the ratio between out-of-field dose in FFF beams compared to flattened beams as a function of field size and distance from the field centre. While the data show a lower out-of-field dose for the FFF beam (ratio < 1), there was a significant reduction of dose ratio with distance from the field centre only in case of the smallest field size.

Variation of the ratio between out-of-field dose in FFF beams compared to flattened beams as a function of field size and distance from the field centre. While the data show a lower out-of-field dose for the FFF beam (ratio < 1), there was a significant reduction of dose ratio with distance from the field centre only in case of the smallest field size. In the smallest field size, head leakage outside of the field dominated over phantom scatter and as such the change with distance reflected the shape of the primary dose profile. The beam profile of the 6MV photon beam for FFF reached approximately 50% 20 cm away from central axis, which was similar to the reduction of the dose ratio in Fig. 4. In the studied linear accelerator the collimator angle was also found to influence out-of-field doses. Fig. 5 shows the out-of-field dose with the collimator angles of 0° and 90° for measurements using a flattened beam. The reduction of leakage dose at a distance of 20 cm was of the order of 50% when the collimator was set to 90°.
Fig. 5

Out-of-field doses from abdominal 5 × 5 cm2 fields at various distances from the field edge using a flattened beam and different collimator angles. The location of the MLC bank in case of 90° collimator rotation is indicated.

Out-of-field doses from abdominal 5 × 5 cm2 fields at various distances from the field edge using a flattened beam and different collimator angles. The location of the MLC bank in case of 90° collimator rotation is indicated.

Discussion

Out-of-field dose is composed of leakage, collimator and patient scatter [13]. The variation in measurements at the same distance from the field edge was related to the location of the field relative to the measurement point in the phantom. While several studies have investigated this in homogenous phantoms [12], [22], [23] we investigated out-of-field doses in a more realistic scenario using an anthropomorphic phantom. There was a lack of full and symmetric scatter conditions, and low-density structures such as lung affected the radiological path length of the scattered radiation. Fig. 1e shows an example of scattered radiation paths that traverse tissue and/or lung. As out-of-field dose is more concerning for late effects in children [10], [24] an anthropomorphic phantom was chosen that mimics a 5 year old in size and body composition. A limitation of our study was that only static fields symmetric around central axis (CAX) were explored. It can be expected that the complexity increases further if modulated arc treatments are used. For both flattened and FFF beams larger field sizes resulted in greater out-of-field doses at small distances from CAX where phantom scatter dominates as illustrated in Fig. 2. However, when the distance from the CAX was increased the dependence on field size is reduced as leakage becomes more important [21], [25]. The data shown in Fig. 4 for the smallest field size demonstrates this as phantom scatter was small and the difference between flattened and FFF beam reflected the beam profiles behind MLC and collimator. Close to the field edge, patient and collimator scatter were found to be dependent on the field size which causes the large differences between the doses for different field sizes [26], [27]. Two components contributed to the reduction in out-of-field dose in FFF beams compared to flattened beams: (i) the removal of a major scattering component source, the flattening filter [28], and (ii) the fact that all beams were normalised at central axis which will lead to less primary photons in the periphery for FFF beams resulting in lower transmission through the collimation system. Fig. 5 shows the ratio of the dose with the collimator at 0° to that at 90° as a function of distance from the field centre. One technique for dose reduction in Varian linear accelerators is to rotate the collimator to align the MLC bank to the longitudinal axis of patient which reduces the dose in a 6MV beam by half for distances greater than 15 cm from the field edge [15]. The work presented in the present study confirmed that the use of FFF beams for SABR treatments can deliver lower out-of-field doses overall compared to their flattening filter beam counterparts [26], [28]. This will not only reduce the probability of developing a second cancer but also other late effects such as cardiac toxicity as a result of the radiation exposure.[29], [30] All this is particularly important for paediatric patients and was shown to hold in the complex dose distributions observed in anthropomorphic phantoms.

Declaration of Competing Interest

The authors declare the following financial interests/personal relationships which may be considered as potential competing interests: Tomas Kron and Nicholas Hardcastle are coinvestigators in an unrelated research collaboration with Varian Medical Systems
  30 in total

1.  Variations in dose response with x-ray energy of LiF:Mg,Cu,P thermoluminescence dosimeters: implications for clinical dosimetry.

Authors:  Lisa Duggan; Claire Hood; Helen Warren-Forward; Mamoon Haque; Tomas Kron
Journal:  Phys Med Biol       Date:  2004-09-07       Impact factor: 3.609

2.  Technical Report: Evaluation of peripheral dose for flattening filter free photon beams.

Authors:  E L Covington; T A Ritter; J M Moran; A M Owrangi; J I Prisciandaro
Journal:  Med Phys       Date:  2016-08       Impact factor: 4.071

3.  Out-of-field in vivo dosimetry using TLD in SABR for primary kidney cancer involving mixed photon fields.

Authors:  P Lonski; S Keehan; S Siva; D Pham; R D Franich; M L Taylor; T Kron
Journal:  Phys Med       Date:  2017-04-07       Impact factor: 2.685

4.  A collimated detection system for assessing leakage dose from medical linear accelerators at the patient plane.

Authors:  P Lonski; M L Taylor; R D Franich; T Kron
Journal:  Australas Phys Eng Sci Med       Date:  2013-12-08       Impact factor: 1.430

5.  OUT-OF-FIELD DOSES IN CHILDREN TREATED FOR LARGE ARTERIOVENOUS MALFORMATIONS USING HYPOFRACTIONATED GAMMA KNIFE RADIOSURGERY AND INTENSITY-MODULATED RADIATION THERAPY.

Authors:  Marijke De Saint-Hubert; Marija Majer; Hrvoje Hršak; Zdravko Heinrich; Željka Kneževic; Saveta Miljanic; Paulina Porwol; Liliana Stolarczyk; Filip Vanhavere; Roger M Harrison
Journal:  Radiat Prot Dosimetry       Date:  2018-10-01       Impact factor: 0.972

Review 6.  A review of dosimetry studies on external-beam radiation treatment with respect to second cancer induction.

Authors:  X George Xu; Bryan Bednarz; Harald Paganetti
Journal:  Phys Med Biol       Date:  2008-06-09       Impact factor: 3.609

7.  Peripheral dose measurements for 6 and 18 MV photon beams on a linear accelerator with multileaf collimator.

Authors:  Michalis Mazonakis; Fotini Zacharopoulou; Haralambos Varveris; John Damilakis
Journal:  Med Phys       Date:  2008-10       Impact factor: 4.071

8.  Stereotactic Body Radiation Therapy for Metastatic and Recurrent Solid Tumors in Children and Young Adults.

Authors:  Christopher L Tinkle; Charu Singh; Shane Lloyd; Yian Guo; Yimei Li; Alberto S Pappo; Steven G DuBois; John T Lucas; Daphne A Haas-Kogan; Stephanie A Terezakis; Steve E Braunstein; Matthew J Krasin
Journal:  Int J Radiat Oncol Biol Phys       Date:  2020-11-28       Impact factor: 7.038

9.  International incidence of childhood cancer, 2001-10: a population-based registry study.

Authors:  Eva Steliarova-Foucher; Murielle Colombet; Lynn A G Ries; Florencia Moreno; Anastasia Dolya; Freddie Bray; Peter Hesseling; Hee Young Shin; Charles A Stiller
Journal:  Lancet Oncol       Date:  2017-04-11       Impact factor: 41.316

10.  Assessment of leakage dose in vivo in patients undergoing radiotherapy for breast cancer.

Authors:  Peta Lonski; Tomas Kron; Michael Taylor; Alicia Phipps; Rick Franich; Boon Chua
Journal:  Phys Imaging Radiat Oncol       Date:  2018-03-26
View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.