| Literature DB >> 34307913 |
Lachlan Garrett1,2, Nicholas Hardcastle2,3, Adam Yeo2, Peta Lonski2, Rick Franich1, Tomas Kron1,2,3.
Abstract
BACKGROUND ANDEntities:
Keywords: High dose rate; Out-of-field dose; Paediatric; Peripheral dose; SABR
Year: 2021 PMID: 34307913 PMCID: PMC8295843 DOI: 10.1016/j.phro.2021.05.006
Source DB: PubMed Journal: Phys Imaging Radiat Oncol ISSN: 2405-6316
Thermoluminescence dosimeter locations (organs) and distances from the central axis for all four radiotherapy field sites.
| Distance from the field centre (cm) | ||||
|---|---|---|---|---|
| Measurement location | Brain | Neck | Thorax | Abdomen |
| Brain | – | 15.2 | 25.2 | 39.9 |
| Thyroid | 15.2 | – | 10.2 | 24.7 |
| Heart | 25.0 | 9.9 | 2.7 | 14.8 |
| Lung | 25.9 | 11.8 | 9.1 | 16.1 |
| Abdomen | 39.9 | 24.7 | 15.1 | – |
| Gonads | 52.5 | 37.3 | 27.6 | 12.6 |
Fig. 1Schematic representing the four different field locations used in the present study: (a) brain, (b) neck, (c) thorax, and (d) abdomen. Shown are the 2 × 2 cm2 fields. Additional field sizes of 5 × 5 cm2 and 10 × 10 cm2 were also used. TLDs were placed at each labelled anatomical location. (e) shows two measurement points at equal distances from the beam (i) and (ii) but with different radiological path lengths due to inhomogeneities.
Fig. 2Out-of-field doses for two field sizes using a 6MV X-ray beam in flattened and flattening filter free (FFF) mode as a function of the distance from the field centre.
Fig. 3Comparison of out-of-field dose behind lung and tissue for a 2 × 2 cm2 field of 6MV X-rays with and without flattening filter. The out-of-field doses in the FFF beams were lower than for the flattened beams shown in the left panel.
Fig. 4Variation of the ratio between out-of-field dose in FFF beams compared to flattened beams as a function of field size and distance from the field centre. While the data show a lower out-of-field dose for the FFF beam (ratio < 1), there was a significant reduction of dose ratio with distance from the field centre only in case of the smallest field size.
Fig. 5Out-of-field doses from abdominal 5 × 5 cm2 fields at various distances from the field edge using a flattened beam and different collimator angles. The location of the MLC bank in case of 90° collimator rotation is indicated.