Photoacoustic tomography (PAT) has become increasingly popular for molecular imaging due to its unique optical absorption contrast, high spatial resolution, deep imaging depth, and high imaging speed. Yet, the strong optical attenuation of biological tissues has traditionally prevented PAT from penetrating more than a few centimeters and limited its application for studying deeply seated targets. A variety of PAT technologies have been developed to extend the imaging depth, including employing deep-penetrating microwaves and X-ray photons as excitation sources, delivering the light to the inside of the organ, reshaping the light wavefront to better focus into scattering medium, as well as improving the sensitivity of ultrasonic transducers. At the same time, novel optical fluence mapping algorithms and image reconstruction methods have been developed to improve the quantitative accuracy of PAT, which is crucial to recover weak molecular signals at larger depths. The development of highly-absorbing near-infrared PA molecular probes has also flourished to provide high sensitivity and specificity in studying cellular processes. This review aims to introduce the recent developments in deep PA molecular imaging, including novel imaging systems, image processing methods and molecular probes, as well as their representative biomedical applications. Existing challenges and future directions are also discussed.
Photoacoustic tomography (PAT) has become increasingly popular for molecular imaging due to its unique optical absorption contrast, high spatial resolution, deep imaging depth, and high imaging speed. Yet, the strong optical attenuation of biological tissues has traditionally prevented PAT from penetrating more than a few centimeters and limited its application for studying deeply seated targets. A variety of PAT technologies have been developed to extend the imaging depth, including employing deep-penetrating microwaves and X-ray photons as excitation sources, delivering the light to the inside of the organ, reshaping the light wavefront to better focus into scattering medium, as well as improving the sensitivity of ultrasonic transducers. At the same time, novel optical fluence mapping algorithms and image reconstruction methods have been developed to improve the quantitative accuracy of PAT, which is crucial to recover weak molecular signals at larger depths. The development of highly-absorbing near-infrared PA molecular probes has also flourished to provide high sensitivity and specificity in studying cellular processes. This review aims to introduce the recent developments in deep PA molecular imaging, including novel imaging systems, image processing methods and molecular probes, as well as their representative biomedical applications. Existing challenges and future directions are also discussed.
Molecular imaging provides unique opportunities for visualizing, characterizing, and
quantifying biological processes at the molecular level.[1-4] With the aid of state-of-the art imaging technologies, the working mechanisms
of the biological systems have been better understood in the context of living
tissues, which leads to numerous advancements in disease diagnosis, drug design, and
therapy assessment. Although the methodology originated from nuclear medicine
(e.g., positron emission tomography), which uses radiolabeled
tracers that generate gamma signals from radioactive decay, molecular imaging has
been generalized so that signaling from molecules is not restricted to only
radioactive atoms.[1,5] X-ray computed tomography (CT),[6-8] magnetic resonance imaging (MRI),[6,9,10] ultrasound imaging (US),[11-13] optical imaging,[14,15] and photoacoustic tomography (PAT)[16-18] have all been utilized for molecular imaging. X-ray CT has high acquisition
speed, high spatial resolution, whole-body penetration, and relatively low cost.
However, X-ray CT uses ionizing radiation, which limits the long-term observation.
MRI uses non-ionizing radiation, and has a large penetration and excellent imaging
contrast of soft tissues. However, MRI has poor sensitivity to molecular probes,
which leads to a relatively long acquisition time and a large dose of contrast
agent. Ultrasound imaging has high spatial and temporal resolution, relatively large
penetration depth, and high portability; but it also has poor molecular sensitivity.
Optical imaging has excellent spatial resolution and superior molecular sensitivity,
and has been widely used in basic research to observe intracellular events in live
intact cells, but it suffers from poor penetration depth due to strong optical
scattering in biological tissues.PAT, also referred to as optoacoustic tomography, is a hybrid imaging modality that
combines the high contrast of optical imaging with the high spatial resolution and
penetration depth of ultrasound imaging.[16,19-21] In PAT, the target is illuminated by nanosecond laser pulses, and the
absorbed optical energy is partially or completely converted into heat, leading to a
transient local temperature rise that generates a local pressure rise and
subsequently propagating acoustic waves. PAT detects the acoustic signals produced
by optical absorption from either endogenous chromophores, such as hemoglobin,
melanin and lipid, or exogenous contrast agents, including nanoparticles, organic
dyes and reporter gene products. Since the scattering of ultrasound waves in tissue
is 2-3 orders of magnitude less than the optical equivalent, PAT can generate
high-resolution images in optical diffusive regime.[20] PAT is also extremely sensitive to optical absorption. A small change in
optical absorption coefficient leads to the same fractional change in photoacoustic
signal. Therefore, PAT has a 100% sensitivity to optical absorption contrast and is
inherently capable of functional and molecular imaging. With these merits, PAT has
been widely applied in diverse fundamental research and clinical practice.[18,22-25]There are 2 major implementations in PAT: photoacoustic microscopy (PAM) and
photoacoustic computed tomography (PACT).[26-30] PAM targets the quasi-ballistic and quasi-diffusive regimes with
tightly/weakly focused excitation light and tightly focused single-element
ultrasonic detection for direct image formation,[20,26,31,32] while PACT uses wide-field light illumination and parallel acoustic detection
to image targets mainly in the diffusive regime.[33-36] PACT can be configured by using ultrasonic transducer arrays such as a linear
array, ring array and 2D matrix array.[33,34,37,38] The axial resolution of PACT is determined by the detection bandwidth of the
ultrasonic transducer array, and the lateral resolution is determined by the central
frequency and numerical aperture. Compared with PAM, PACT typically has lower
spatial resolution (∼400 µm for a 5 MHz linear array) but higher imaging speed and
penetration depth (up to a few centimeters). So far, PACT is the most promising for
clinical translations, such as breast cancer imaging and functional brain imaging,
due to its deep penetration and high compatibility with commercial ultrasound
imaging systems. Thus, in this review, we will focus on the advances and
applications of PACT technologies for deep molecular imaging. Readers are referred
to other review articles about the PAM technologies.[28,39] In the rest of the paper, we use PAT and PACT interchangeably if not
otherwise noted.Despite the above merits, PAT also faces certain challenges. Although PAT can achieve
a penetration depth of several centimeters, which far exceeds the penetration of
high-resolution optical imaging modalities, it still cannot match the penetration
depths of ultrasound imaging, X-ray CT and MRI. The strong optical attenuation of
biological tissues is the major limiting factor of PAT’s penetration depth. In
addition, PAT is more sensitive to molecular contrast agents than ultrasound imaging
or MRI, but it is less sensitive than fluorescence imaging and PET.[20] Moreover, many PACT implementations suffer from 2 types of imaging artifacts:
limited-view and limited-bandwidth artifacts, which often deteriorate the image
quality of deep seated targets with weak signals.[40-43] Numerous efforts have been made by various research groups to address the
aforementioned challenges. Through careful selection of excitation wavelength,[44-46] optimization of light delivery,[47-50] and sophisticated design and fabrication of the ultrasonic transducer,[51-55] the penetration depth of PAT has been substantially improved. Moreover, by
increasing the imaging speed or frame rate, PAT can reduce the motion artifacts and
improve the detection sensitivity with signal averaging, leading to larger imaging depth.[56-58] More advanced imaging reconstruction methods, such as model-based and
deep-learning-based reconstruction algorithms, have been developed to further
mitigate image artifacts and improve the image quality.[40,41,59-63] Moreover, the enhanced optical absorption of molecular probes has boosted the
sensitivity of PAT considerably.In this review, we describe the recent advances in improving the penetration depth
and sensitivity of PA molecular imaging, focusing on 3 groups of efforts: innovative
imaging systems, advanced reconstruction algorithms, and novel molecular contrast
agents. In the following sessions, we first introduce various PAT technological
innovations to upgrade penetration with more efficient optical excitation and
acoustic detection. Next, we present advanced image reconstruction and data analysis
methods to compensate for optical fluence attenuation and eliminate image artifacts.
Then, we summarize the new PA-sensitive molecular probes that have superior imaging
performance and functionality. Representative applications in biomedical research
are also discussed. Finally, we conclude the review with a brief discussion about
future improvements in molecular PAT.
Technical Advancements in Molecular PAT
Novel PAT Systems to Improve Penetration Depth
A deep-penetrating PAT system should achieve deep optical excitation and
sensitive acoustic detection. To deliver more light into deeper tissues, efforts
have been reported with using longer excitation wavelengths to reduce the
optical attenuation, illuminating the target from the inside of the tissues, and
applying wavefront shaping to focus the light in the scattering medium. To
detect weak signals from deeper tissues, new fabrication procedures of
piezoelectric transducers and various optical detectors have been explored to
improve the detection sensitivity.
PAT with innovative light source and delivery
In conventional PAT systems, laser pulses are delivered either via free space
or an optical guide to form a dark- or bright-field illumination pattern on
the tissue surface.[16,20,21] Yet due to the severe wavelength-dependent attenuation of photons
(e.g., effective attenuation coefficient of human
breast tissue is ∼4.0 cm-1 at 730 nm[64]), it remains a great challenge for PAT to penetrate beyond a few centimeters.[19] Novel light sources and delivery methods are highly desired to break
the penetration limit.
Long-wavelength photoacoustic tomography
As light in the near-infrared (NIR) wavelength range has relatively low
attenuation, applying longer-wavelength excitation light is often the
first step in PAT to improve the penetration depth.[65] In the early days of PAT, short-wavelength light (<700 nm) was
often used for imaging hemoglobin due to its strong absorption in the
visible wavelength range,[19,65,66] but a depth of only a few millimeters could be imaged due to the
overwhelming optical scattering at short wavelengths.[67,68] Later, NIR light (700 nm to 1350 nm) was explored due to its
larger penetration depth with relatively low absorption by water.[19,44,65,69,70] For example, a penetration depth of 4 cm has been achieved
in vivo in the human breast at 800 nm.[71] With highly-absorbing exogenous contrast agents, a penetration
depth of 11 cm was achieved in chicken breast tissue.[72]
Figure 1A to D
compare the PA images of a mouse at 4 different wavelengths of 532nm,
700 nm, 850 nm, and 1064 nm. While the PA image at 532 nm provides
mostly superficial blood vessels, organs such as spleen and cecum can be
better observed at longer wavelengths.
Figure 1.
Deep PAT and TAT with long excitation wavelengths. (a-d)
Noninvasive PA imaging of a whole-body mouse in
vivo.
[44] Depth-encoded PA maximum amplitude projection (MAP)
images acquired at optical wavelengths of 532 nm, 700 nm, 850
nm, and 1064 nm. (e) TAT image of excised normal breast tissue
of an ewe.[83] (f) TAT image of a breast tumor embedded in the excised
breast tissue. (g) X-ray image of the tumor phantom. (h)
Normalized image contrast of normal tissue, tumor, and
background in the TAT and X-ray images. Reproduced with
permission from.[44,83]
Deep PAT and TAT with long excitation wavelengths. (a-d)
Noninvasive PA imaging of a whole-body mouse in
vivo.
[44] Depth-encoded PA maximum amplitude projection (MAP)
images acquired at optical wavelengths of 532 nm, 700 nm, 850
nm, and 1064 nm. (e) TAT image of excised normal breast tissue
of an ewe.[83] (f) TAT image of a breast tumor embedded in the excised
breast tissue. (g) X-ray image of the tumor phantom. (h)
Normalized image contrast of normal tissue, tumor, and
background in the TAT and X-ray images. Reproduced with
permission from.[44,83]Long wavelengths in the microwave spectrum have been employed as
radiation sources in PAT to achieve greater penetration depth, a
technique also referred to as thermoacoustic tomography (TAT).[73] Unlike PAT, which relies on the optical absorption contrast, the
contrast in TAT stems from tissue’s dielectric properties.[74,75] In the frequency range from 0.1 to 10 GHz, the relative
dielectric constant in soft tissues ranges from 5 to 70, and the
conductivity ranges from 0.02 to 3 S/m, leading to a penetration depth
of TAT up to 15 cm,[73,74] which is sufficient for imaging whole-body small animals and many
deep organs in humans. Great progress has been made in TAT for imaging
breast and prostate cancer tissues with >10 cm penetration depth.
Moreover, because the dielectric properties of malignant tissues are
larger than the normal tissues, the cancer tissues often have relatively
strong microwave absorption than healthy tissues.[76-79] For example, breast cancer tissue has up to 10 times higher
microwave absorption than normal tissues, providing high imaging
contrast for TAT.[73]Conventional TAT systems usually use high-power pulse-modulated
carrier-frequency generators, and the resultant transmitted microwaves
operate at frequencies ranging from 434 MHz to 3 GHz with a pulse width
of a few microseconds.[80-82] The excitation pulse is prolonged to deposit more energy at the
expense of spatial resolution.[46,74,76,81] To reduce the microwave pulsewidth and improve the spatial
resolution, a high-voltage discharge mode has been used to generate
ultrashort microwave pulses. The microwave generator uses an oscillation
circuit and a Tesla transformer to generate high-voltage electrical
pulses, which propagate through the antenna and radiate microwaves with
a pulse energy of hundreds of millijoules[45,75,83] and a pulsewidth of several nanoseconds. A spatial resolution of
tens of microns can thus be achieved. Figure 1E to G show the TAT and
corresponding X-ray images of the excised normal breast tissue and tumor
tissue, respectively. The location of the tumor can be clearly imaged by
TAT, with higher contrast between normal tissue and tumor tissue than
that in the X-ray images (Figure 1H). Nevertheless, the
microwaves used in TAT often have a low frequency (100–900 MHz), which
can penetrate deeper in tissue, but may result in a lower
signal-to-noise ratio (SNR). Meanwhile, it is challenging to direct
microwave propagation direction.[73]
Internal light delivery
In many clinical applications, the deeply-seated organs of interest are
relatively close to some body cavities. For example, the posterior
boundary of an adult’s kidney can be more than 10 cm from the skin
surface, but the kidney is connected to the urinary tract that can be
directly accessed from outside.[84] For these organs, it is therefore possible to deliver the light
to the intracorporeal targets using a small-diameter optical fiber via
the body cavities, largely avoiding the otherwise strong optical
attenuation by the intervening tissues underneath the skin surface. Such
light delivery strategy is termed as internal light illumination.
Various internal-illumination designs have been employed in PAT to
deliver light to the deep organs of interest, and can be roughly
classified into side-viewing illumination and forward-viewing
illumination. In side-viewing illumination, the laser beam is coupled
into a multi-mode optical fiber and the fiber is inserted into the
vicinity region of the target. To deflect the light sideward, a
45-degree-angled prism is attached to the fiber tip or the fiber tip
itself is polished with a 45-degree angle. The tip is sealed within a
transparent air chamber.[49,85] In contrast, in forward-viewing illumination, the end of the
optical fiber is usually sanded until the tip is beveled to a desired
conical shape. Light emits from the conical fiber tip and forms a
spherical illumination pattern; in other cases, light emits directly
from the flat fiber tip to form a less diverged illumination.[86,87]
Figure 2A and B
show example designs of forward- and side-viewing fiber tips,
respectively. Generally speaking, side-viewing illumination is more
suitable for imaging tunnel-shaped organs, such as the gastrointestinal
tract, and front-viewing illumination works better for imaging
planar-surfaced organs, such as the cervix. However, both types of
illumination typically cover only a small field of view, mostly because
of the limited beam-diverging distance between the fiber tip and the
tissue surface.
Figure 2.
Internal-illumination PAT. (a) A representative forward-viewing fiber.[87] (b) A representative side-viewing fiber.[49] (c) A representative radial-emission fiber diffuser.[50] (d) A representative frontal-emission diffuser fiber
diffuser. (e-f) B-mode ultrasound image (e) and PA image (f) of
the blood-filled tube on an ex vivo pig kidney
overlaid by an 8-cm-thick layer of fresh chicken breast tissue.[92] Reproduced with permission from.[49,50,87,92]
Internal-illumination PAT. (a) A representative forward-viewing fiber.[87] (b) A representative side-viewing fiber.[49] (c) A representative radial-emission fiber diffuser.[50] (d) A representative frontal-emission diffuser fiber
diffuser. (e-f) B-mode ultrasound image (e) and PA image (f) of
the blood-filled tube on an ex vivo pig kidney
overlaid by an 8-cm-thick layer of fresh chicken breast tissue.[92] Reproduced with permission from.[49,50,87,92]To illuminate a larger field of view, such as the adult kidney, Li et al
have proposed to use an optical fiber diffuser for internal light
delivery in PAT.[49] Two types of fiber diffusers have been explored for this purpose:
radial-emission and frontal-emission diffusers. To make a
radial-emission fiber diffuser, the jacket and cladding of an optical
fiber are removed, and the surface of the fiber core is roughened to
scatter the light out.[47,48,88] The fiber core can also be coated with a layer of optically
scattering particles to further homogenize the light emission.[50,89,90] In contrast, a frontal diffuser is typically separable from the
optical fiber. The diffuser is made of scattering medium, such as
intralipid and TiO2, with increasing concentrations from the
fiber tip so as to generate a homogeneous light distribution in the
nearby tissues.[91,92]
Figure 2C and D
show the representative designs of radial-emission and frontal-emission
diffusers. As an example, Figure 2D shows a
frontal-emission diffuser made of segmented agar mixed with intralipid.
By using this fiber diffuser, the internal illumination PAT has
demonstrated deep imaging on an ex vivo pig kidney at a
depth of 8 cm, with the fiber diffuser inserted into the pig kidney
through the urinary tract, as shown Figure 2E and F. Such an imaging
depth is typically not attainable by traditional PAT.
Internal-illumination PAT needs to use optical fibers with high damage
threshold to deliver sufficient energy to the target. Therefore, there
is often a tradeoff between the fiber’s diameter and its flexibility. A
larger-diameter fiber is less flexible and thus may not fit for certain
applications with limited space. In addition, internal-illumination PAT
still suffers from the acoustic attenuation from the deep targets,
especially the high frequency components, which may lead to deteriorated
spatial resolutions.
X-ray acoustic imaging
Another way to increase the penetration depth of PAT is to use X-ray
photons as the radiation source, which is termed as X-ray induced
acoustic imaging (XAI). X-ray photon scattering in tissue is much lower
than that of visible-NIR photons.[93] Several groups have reported XAI systems that include a linear
accelerator (LINAC), an ultrasonic transducer (array), and a data
acquisition system.[94-96] The X-ray pulses generated by the LINAC typically consist of
picoseconds pulse trains and has a total pulsewidth of a few microseconds.[95] The amplitude of an X-ray-acoustic (XA) wave is proportional to
the X-ray photon absorption, which is more prominent in high-density
materials such as lead blocks, bones, and microcalcifications.[96] An ultrasonic transducer, which typically has the central
frequency of less than 1 MHz, is used to acquire the generated XA signals.[94,97,98] One common application of XAI is to provide dosimetric
information during radiation therapy.[95-97]
Figure 3A and B
show the XAI map of a 6 cm×3 cm radiation field at 10 cm depth in water,
demonstrating that the XA intensity is closely related to the deposited
X-ray dose.[96] Compared with traditional PAT using visible-NIR light, XAI has
superior penetration depth in tissues due to the weak X-ray scattering.[95] As shown in Figure 3C, while the PA signals rapidly decreased with the
depth, XA signals only slighted decreased up to 10 cm. However, compared
to visible-NIR laser pulses, X-ray pulses are much longer, which leads
to a lower X-ray acoustic conversion efficiency and degraded spatial resolution.[21,99] To improve the resolution and SNR of XAI, Xiang et al used an
X-ray tube to produce nanosecond pulses and achieved an ∼300 µm resolution.[99] Ultrasonic transducer arrays were also used to achieve high-speed
3D XAI.[98] Nevertheless, one major concern in XAI is the ionizing radiation exposure.[100] It has been reported that the minimum dose required to obtain an
XAI field image in water is 11.6 mGy, with an SNR of 6.4,[97] which is comparable to a typical chest X-ray CT dose of 12-18
mGy. It is promising that XAI can provide clinically meaningful images
at low doses.
Figure 3.
X-ray acoustic tomography. (a) XAI image of a 6 cm × 3 cm
radiation field at 10 cm depth.[96] (b) Comparison of lateral profiles of the radiation field
extracted from simulated and experimental XAI images and ion
chamber measured profiles. (c) SNRs of XA and PA signals of a
lead sample beneath various thicknesses of chicken breast tissues.[95] (d) Photograph and XAI image of a metal object embedded
in tissue at 40.5 mm depth.[94] Reproduced with permission from.[94-96]
X-ray acoustic tomography. (a) XAI image of a 6 cm × 3 cm
radiation field at 10 cm depth.[96] (b) Comparison of lateral profiles of the radiation field
extracted from simulated and experimental XAI images and ion
chamber measured profiles. (c) SNRs of XA and PA signals of a
lead sample beneath various thicknesses of chicken breast tissues.[95] (d) Photograph and XAI image of a metal object embedded
in tissue at 40.5 mm depth.[94] Reproduced with permission from.[94-96]
Wavefront shaping enhanced light delivery
PAT’s penetration depth in tissues can be enhanced by employing wavefront
shaping techniques. Wavefront engineering technologies have enabled
refocusing light inside scattering medium by compensating for or
reversing the scattering-induced phase scrambling, based on the fact
that optical scattering is microscopically deterministic and
time-reversible in nature.[101-105] In a typical wavefront shaping implementation, a spatial light
modulator (SLM) is used to optimize the phase of input light by using a
guide star as the signal feedback,[106-108] as shown in Figure 4A and B. Doing so, the light intensity in deep
scattering medium can be enhanced by more than 1000 times (Figure 4C and D).
Many types of guide stars have been reported, including fluorescent labels,[109,110] nonlinear optical particles,[111] as well as photoacoustic and ultrasound targets.[102,103,112,113]
Figure 4.
PAT enhanced by wavefront shaping. (a) Principle of a
phase-unmodified coherent beam of light travels into tissue.[107] (b) Principle of beam refocusing by wavefront-shaping the
incident light field with an SLM.[115] (c) Representative transmission micrograph with an
unshaped incident beam. (d) Representative transmission after
optimization for beam focusing. (e) PA imaging of 2 capillary
samples without and with applying wavefront shaping. (f) PA
profiles across the capillary tubes with and without wavefront shaping.[116] Reproduced with permission.[107,115,116]
PAT enhanced by wavefront shaping. (a) Principle of a
phase-unmodified coherent beam of light travels into tissue.[107] (b) Principle of beam refocusing by wavefront-shaping the
incident light field with an SLM.[115] (c) Representative transmission micrograph with an
unshaped incident beam. (d) Representative transmission after
optimization for beam focusing. (e) PA imaging of 2 capillary
samples without and with applying wavefront shaping. (f) PA
profiles across the capillary tubes with and without wavefront shaping.[116] Reproduced with permission.[107,115,116]PAT is advantageous over fluorescence imaging for providing wavefront
shaping feedback signals, since PAT can penetrate relatively deep. Via
wavefront shaping, maximizing the PA signal strength will lead to
maximizing the local optical fluence within the ultrasound detection zone,[102,103,114] which in turn increases the penetration depth of PAT. An
enhancement of 5∼10 fold in the PA signal strength has been reported by
Kong et al after wavefront optimization.[114]
Figures 4E and F
show the enhancement of photoacoustic imaging with wavefront shaping
(PAWS), in which both the SNR and spatial resolution were substantially
improved. However, so far, most wavefront shaping technologies are still
too slow for in vivo PAT applications, due to the fast
optical speckle decorrelation induced by the tissue motion and blood
flow. Further improvement in the wavefront shaping speed is still
needed, such as by accelerating the phase optimization process.
PAT with ultrasensitive and fast ultrasonic detection
Another important factor that affects the penetration depth of PAT is the
sensitivity of the ultrasonic detection. A highly sensitive ultrasonic
transducer is able to detect weak signals from deep-seated chromophores near
which light has been severely attenuated. The sensitivity of an ultrasonic
transducer is usually assessed by its noise equivalent pressure (NEP) with
units of Pa, which reflects the minimum detectable
photoacoustic signal pressure.[39,117] A lower NEP is desired for a more sensitive ultrasonic
transducer.
Sensitive PZT transducers
Piezoelectric transducers are most widely used for acoustic detection in
PAT, and have demonstrated the best sensitivity. For example, a
piezoelectric transducer with a detection area of 30 mm2 and
a bandwidth of 50 MHz has achieved an NEP of ∼77 Pa.[118] The detection sensitivity of a traditional piezoelectric
transducer can be improved by further increasing the active element area
and decreasing the detection bandwidth.[39] However, both measures lead to reduced PAT image quality. First,
it is technically challenging to fabricate ultrasonic transducer arrays
with large elements, which also suffer from reduced directivity and
strong grating lobes.[119] Moreover, miniaturized ultrasonic transducers are often needed
for photoacoustic endoscopy in order to access the intracorporeal organs.[120-124] Second, a broadband detection bandwidth is preferred in PAT to
achieve high spatial resolution and reduce limited-bandwidth artifacts.[16,125] Although high-frequency ultrasonic waves are attenuated more
strongly by tissues, a broadband transducer is still highly desired to
detect the deep-seated small targets.[19]Recent advances in ultrasonic transducer fabrication has shown that the
loss of sensitivity with a small element size can be mitigated in
capacitive micromachined ultrasonic transducers (CMUTs) and
piezoelectric micromachined ultrasonic transducers (PMUTs).[126-130] The basic building block of CUMT is a capacitor cell, which
consists of a thin, movable plate suspended over a vacuum gap on top of
silicon substrates. CMUT uses semiconductor fabrication techniques, such
as the wafer bonding and sacrificial release processes. The metal layer
on the plate or the plate itself serves as the top electrode of the
capacitor, and the conductive substrate acts as the bottom electrode.
After connecting to a constant bias voltage, an electrical current is
generated when ultrasound pressure waves hit the movable top plate that
leads to a change in capacitance.[128,129,131] PMUTs have similar structures with CMUTs, in which an array of
sensitive micro-piezoelectric membranes are formed on the silicon substrates.[55,126,127] It has been experimentally demonstrated that row-column-addressed
2D CMUTs have wider reception fractional bandwidth than classic
piezoelectric transducers with similar acoustic features
(e.g., dimension, central frequency, packaging).
Compared with traditional piezoelectric transducers (PZT), one strength
of CMUTs and PMUTs is the potential high-throughput batch manufacturing,
which can drastically reduce the unit cost and benefit the fabrication
of inexpensive 2D arrays. The sensitivity of CMUTs, however, is not
necessarily higher than conventional PZT transducers, and it can be
possibly improved by optimizing the structure, plate design, layout, and
driving conditions.[54,130]
Optical detection of ultrasound waves
Optical techniques are becoming increasingly popular for PA signal
detection, since they have several advantages over traditional
piezoelectric transducers. The sensitivity of optical sensors is
independent of the detector size, and their bandwidths are usually much wider.[125,132,133] As a result, with the same element size, optical detectors
typically provide greater sensitivity over a significant wider frequency
range than its piezoelectric counterpart, which is particularly
attractive for PA endoscopy or 3D PAT.[51,134,135] In addition, the transparent nature of optical detectors
simplifies the light delivery in PAT.[51] However, when the transducer element size is not a constraining
factor, piezoelectric transducers still provide a higher sensitivity.
Here we introduce several representative optical detection of ultrasound
waves, including Fabry-Pérot (FP) interferometers,[132,136] micro-ring resonators,[52,134,137] and fiber Bragg gratings.[53,138]Fabry-Pérot interferometers (FPI) use transparent planar FP polymer films
as acoustically sensitive elements. The PA excitation light can transmit
through the films into the underlying tissue. The resultant ultrasound
waves modulate the thickness of the FPI. A CW interrogation beam reads
the FPI thickness and records the PA signals.[136] FPI sensors have much higher sensitivity than PVDF receivers with
a similar element size: a 75-µm diameter PVDF receiver has a typical NEP
of 55 kPa, while an FPI with a 64-µm diameter has a NEP
of only 0.31 kPa. A -3 dB detection bandwidth of a
22-µm-thick FPI ranges from 0 to 39 MHz, which is sufficient to detect
deeply located targets. Zhang et al have reported an improved FPI
sensor using a plano-concave optical microresonator, which comprises a
solid plano-concave polymer microcavity formed between 2 highly
reflective mirrors, as shown in Figure 5A.[132,139] The interrogation CW laser beam is then tightly focused by the
top mirror curvature, which is precisely designed to perfectly match the
beam divergence and avoid the beam from walking off as it does in the
planar FPI.[132] A high degree of optical confinement can then be achieved and
eventually lead to a much improved detection sensitivity: the NEP of a
100-µm plano-concave microresonator is 13.75 Pa, which
is comparable to the NEP of a 2 mm PVDF needle hydrophone. Figure 5D and E
show PA images of mouse abdominal skin microvasculature using an
endoscopic FPI probe with a diameter of 3.2 mm. The high image quality
has demonstrated the possibility of all-optics methods for deep PAT.
Figure 5.
PAT with optical detection of ultrasound waves. (a) The schematic
of a plano-concave FPI microresonator.[132] (b) Scanning electron micrograph of the MRR (left) and
one cross-section of the ring (right).[52] (c) Geometry of fiber Bragg gating (FBG) detector.[138] (d-e) PA images of mouse abdominal skin microvasculature
using a forward-viewing FPI probe.[51] (f) Photograph of a hair phantom. (g) Volumetric
rendering of the PA imaging of the hair phantom by an MRR
detector. The cross-sectional view of the dashed region is shown
in the bottom panel.[134] Reproduced with permissions from.[51,52,132,134,138]
Micro-ring resonators (MRR), which consist of a bus waveguide and a
ring-shaped resonator separated by a low-dielectric gap, are also useful
for optical detection of ultrasound waves. The light transmitted through
the bus waveguide can be evanescently coupled into the ring resonator
across the gap to form strong optical resonance inside the ring, in
which the stop band or zero transmission is achieved at the resonant
frequency. The deformation of the ring resonator by the ultrasound wave
can cause changes in the optical path length and thus a shift in the
resonant frequency.[52] The design of an MRR detector is shown in Figure 5B. Li et al have
demonstrated that the NEP of a 60-µm-diameter MRR can achieve as low as
6.8 Pa with a 140 MHz bandwidth. Figure 5F and G show the
volumetric PA imaging of human hairs sandwiched by 2 tube walls,
acquired by an MRR sensor. However, MRR typically provides a broad
bandwidth only with a small acceptance angle, which is not ideal for
parallel acoustic detection in PAT.[125]PAT with optical detection of ultrasound waves. (a) The schematic
of a plano-concave FPI microresonator.[132] (b) Scanning electron micrograph of the MRR (left) and
one cross-section of the ring (right).[52] (c) Geometry of fiber Bragg gating (FBG) detector.[138] (d-e) PA images of mouse abdominal skin microvasculature
using a forward-viewing FPI probe.[51] (f) Photograph of a hair phantom. (g) Volumetric
rendering of the PA imaging of the hair phantom by an MRR
detector. The cross-sectional view of the dashed region is shown
in the bottom panel.[134] Reproduced with permissions from.[51,52,132,134,138]Fiber Bragg grating (FBG) detectors fabricate Bragg gratings inside the
core of an optical fiber.[53] When the wavelength of the incident light satisfies the Bragg
condition, the reflection from all gratings interfere constructively,
resulting in a narrow dip in the transmission optical spectrum. The
ultrasonic wave-induced mechanical perturbation creates changes in the
grating period and leads to a shift in the Bragg wavelength and thus
change in the light transmission. The geometry of an FBG detector is
shown in Figure
5C. By detecting the frequency shift of the beating signal
between the 2 orthogonal polarization modes induced by photoacoustic
waves in the FBG based detectors, Liang et al have reported that the FBG
detector can achieve a NEP of 40 Pa and a bandwidth of
50 MHz, with a fiber diameter of 65 µm.[140]
High-speed PAT
High-speed PAT is advantageous for (pre)clinical applications, due to reduced
motion artifacts, real-time feedback, and possible signal averaging. The
imaging speed of traditional PAT systems is often limited by the low pulse
repetition rate of powerful Nd: YAG lasers. With a kilohertz pulse
repetition rate, laser diodes have been explored as the excitation light
source in high-speed PAT, and a large number of frames can be acquired in a
short time.[141,142] However, the low pulse energy is currently the major drawback of
laser diodes. Furthermore, with a large number of transducer elements,
traditional PAT suffers from the time-consuming data transfer and image
reconstruction, preventing real-time imaging. To address this issue,
compress-sensing based PAT has been developed by using only a small number
of transducer elements and incorporating total variation in the image reconstruction.[58,143] Real-time 3D PAT has been achieved by Ozbek et al, with a highest
frame rate of 1.6 kHz and a spatial resolution of 75 µm.[57]
Novel Imaging Reconstruction and Signal Processing Methods
Reconstruction algorithms are essential for PA images with high SNR, high
resolution, and low artifacts, especially when reconstructing PA images with
weak signals from the large depth. In this section, we discuss innovations in PA
molecular imaging as it pertains to penetration depth, reconstruction accuracy,
and detection sensitivity.
Fluence compensation
The initial PA pressure amplitude is directly proportional to the absorbed
optical energy, which is a product of optical fluence and optical absorption
coefficient in tissue. Therefore, the generated PA image is heavily
dependent on the local optical fluence, especially for deep tissue imaging
applications. Biological tissues are highly scattering media for visible-NIR
photons. As the imaging depth increases, optical fluence tends to decrease
exponentially even in a homogenous medium. Due to the heterogeneous nature
of biological tissues, modeling light fluence in tissue is complicated. Some
groups have developed PAT and diffuse optical tomography (DOT) dual-modality
systems to compensate for light fluence variation and achieve more accurate
reconstruction of optical absorption coefficients.[144-146] However, these systems are bulky as they require different
illumination and detection equipment for each modality. Fluence compensation
has been a major challenge in quantitative PAT of blood oxygenation and
other exogenous molecules. In this section, we introduce the most recent
advances in this field. Earlier developments in the area can be found in
review articles.[24,147]Hussain et al have proposed a method for fluence compensation using
ultrasound-enabled, non-invasive tagging of light at the region of interest.[148] As ultrasound scattering in tissue is negligible in comparison to
optical scattering, the ultrasound beam can be well focused at the region of
interest. When tissue is illuminated with a coherent light beam, the photons
are scattered out of the tissue and form a speckle pattern. The fluence map
can be obtained by scanning the focused ultrasound to measure the speckle
contrast. The fluence measurement is directly integrated into a PAT system
such that the fluence maps and PA images can be obtained simultaneously. The
feasibility of this method was assessed in agar and intralipid models,
porcine tissue samples, and sO2 measurements. Imaging results
from sacrificed mice indicate that there was an improvement in fluence
compensated images compared to images without fluence compensation. However,
this method has yet to be tested for in vivo applications
in which complex tissue dynamics come into play. Furthermore, this technique
requires the use of 2 separate lasers and point-by-point scanning, which
makes the system more complex and increases the imaging time (approximately
29 min for an area of 530 mm2).Kim et al have developed a real-time, spectroscopic PA/US system with
simultaneous fluence compensation and motion correction.[149] For PA acquisition, a single image is formed by sequential scanning
using a narrow laser beam through fiber sweeping (called “fast-sweep”).
Light from different fibers propagates different distances in tissue. The
partial PA image reconstructed from a single fiber illumination is used to
estimate the signal loss due to light attenuation (Figure 6). This method allows the
estimation of laser fluence for the targeted absorber. This procedure is
repeated over 10 wavelengths in the range of 700-900 nm. The final image is
formed by a coherent summation of partial images at a video rate of 50 Hz.
The method was tested on in vivo mouse models for
monitoring drug delivery using gold nanorods. Using this “fast-sweep”
approach, the system footprint and cost are reduced. However, a major
drawback of this system is the slow imaging speed, which suffer from motion
artifacts for in vivo cases.
Figure 6.
Spectroscopic PAT with simultaneous fluence compensation and motion
correction. (a) Fluence compensation through a “fast-sweep”
sequential illumination technique. Light emitted from different
fibers propagates different distances to the target located within
tissue. (b) Sequential illumination in PA/US system. These
measurements are used for robust estimation of the laser fluence in
the medium independent of the wavelength. The procedure is repeated
for all wavelengths used in the experiment. Reproduced with permission.[149]
Spectroscopic PAT with simultaneous fluence compensation and motion
correction. (a) Fluence compensation through a “fast-sweep”
sequential illumination technique. Light emitted from different
fibers propagates different distances to the target located within
tissue. (b) Sequential illumination in PA/US system. These
measurements are used for robust estimation of the laser fluence in
the medium independent of the wavelength. The procedure is repeated
for all wavelengths used in the experiment. Reproduced with permission.[149]An integrated PAT system with diffuse reflectance imaging has also been
developed by Jin et al.[150] This system utilizes a single piezoelectric transducer that receives
both PA waves and passive ultrasound (PU) waves. The absorbed photons by the
tissue generate PA signals, while the reflected photons generate PU waves.
For fluence compensation, the relationship between the PU wave amplitude and
light penetration is calibrated. After calibration, the absorption
differences caused by optical fluence variation can be corrected. The
compact design of this system allows integration of various deep imaging
applications. However, the calibration method used to estimate the
relationship between PU amplitude and light fluence may not be available for
all applications.The fluence compensation can also be realized by matching the spectra
profiles of PA images acquired at different wavelengths.[151] The key assumption in this method is that optical diffusion affects
the power spectra of PA signals. When there is no optical diffusion, the PA
spectra follow the same frequency distribution (Figure 7A), regardless of the light
wavelength (Figure
7B). Adding the scattering material leads to a broadening in PA
signal spectrum (Figure
7C), which is also light wavelength dependent (Figure 7D). Frequency
filters are used to correct the frequency domain signal and thus the optical
fluence. Although this work was developed to quantify sO2, this
concept may be extended for molecular imaging as well. However, this
approach was validated only using a very thin phantom and murine muscle
model; therefore, its applications for deeper tissues is unclear. Moreover,
the linear fit for the power spectra ratio may not be valid for certain
geometries.
Figure 7.
Fluence compensation using the PA signal power spectra. (a, c) PA
signal spectra of a gold-plated microscope slide, acquired at 720 nm
and 870 nm, in clear medium and scattering medium, respectively. (b,
d) The ratio of the 2 signal power spectra in (a, c), respectively.
Reproduced with permission from.[151]
Fluence compensation using the PA signal power spectra. (a, c) PA
signal spectra of a gold-plated microscope slide, acquired at 720 nm
and 870 nm, in clear medium and scattering medium, respectively. (b,
d) The ratio of the 2 signal power spectra in (a, c), respectively.
Reproduced with permission from.[151]
Eigenspectrum methods for signal unmixing
In PAT, to estimate different chromophore concentrations, whether it is for
endogenous or exogenous contrast agents, a linear spectral unmixing method
is commonly used to separate the signals from different chromophores.
However, in deep tissue, the linear spectral unmixing method fails due to
spectral coloring induced by wavelength-dependent optical attenuation.[147,152,153] Eigenspectrum-based methods have been recently proposed to address
this issue by converting the fluence correction problem from the spatial
domain to the spectral domain. By modeling light fluence as a linear
combination of eigenspectra, the fluence can be corrected with improved
accuracy.The eigenspectra method was initially proposed in multispectral optoacoustic
tomography (eMSOT).[154] The method is based on the fact that light fluence spectra at
different locations in tissue is a cumulative result of light absorption
from various tissue chromophores. Thus, the fluence spectrum can be
expressed as the linear combination of some base spectra (one mean fluence
spectrum and 3 eigenspectra), determined through principal component
analysis (Figure 8).
It was observed that the first eigenspectrum was associated with the
“spectral shape” of light fluence that related to the average blood
oxygenation of the surrounding tissue, while the second eigenspectrum was
associated with the changes in depth of light fluence and the average
optical properties of the surrounding tissue.[154]
In vivo and postmortem studies were used
to validate these observations. The sO2 estimation can be made
independent of the tissue’s optical properties. However, the accuracy of
eMSOT is prone to noise and artifacts. Based on Bayesian inversion, Olefir
et al described a method to improve eMSOT for spectral unmixing in noisy environments.[155] With this method, spectra can be weighted automatically based on
their reliability. However, this new method still needs improvement in speed
as it is currently ∼12 times slower than eMSOT. Also, this new method needs
to be tested for absorbers other than hemoglobin for molecular imaging
applications.
Figure 8.
eMSOT for optical fluence correction. (a-d) The eigenspectra model
composed of a mean fluence spectrum (a) and 3 fluence eigenspectra
(b-d). (e) Application of a circular grid (red points) for eMSOT
inversion on an area of a simulated PA image. (f) Maps of model
parameters used to spectrally correct the original PA image. (g)
sO2 estimation using linear unmixing (left), eMSOT
(middle), and the gold standard (right) of the selected region.
Reproduced with permission from.[154]
eMSOT for optical fluence correction. (a-d) The eigenspectra model
composed of a mean fluence spectrum (a) and 3 fluence eigenspectra
(b-d). (e) Application of a circular grid (red points) for eMSOT
inversion on an area of a simulated PA image. (f) Maps of model
parameters used to spectrally correct the original PA image. (g)
sO2 estimation using linear unmixing (left), eMSOT
(middle), and the gold standard (right) of the selected region.
Reproduced with permission from.[154]
Model-based 3D reconstruction methods
Although filtered-back projection (FBP) has been widely used in PAT,[156] it does not account for many important factors of the imaging system,
such as the impulse and spatial response of the ultrasonic transducer,
limited-view detection, as well as thermal and electronic noise. Also, in
deep tissues, there exist spectrally derived artifacts. All these factors
impact the quantitative accuracy of the molecular imaging.Many groups have been developing model-based methods to address the above issues.[157] Prakash et al have developed a logarithmic, regularization-based,
entropy maximization algorithm to improve the quantitative performance of PAT.[158] Entropy maximization is used prior to PA image reconstruction,[159,160] with an implicit non-negativity constraint. This method is superior
than conventional l2-norm minimization.[161] However, entropy evaluation with limited independent data still needs
to be investigated for clinical translation, as most clinical systems use
linear ultrasound transducer arrays with relatively small detection
angles.Mastanduno and Gambhir have proposed a model-based tomographic method to map
tissue absorption based on the acoustic data only, with no prior knowledge
of optical properties, initial pressure distributions, or tissue structure.[162] Using the optical diffusion model and k-space acoustic model,
absolute absorption coefficients are reconstructed iteratively.[163] As shown in Figure
9, this method was able to recover the target’s absorption with
an error of up to 22% at 4 cm depth. Although this method showed remarkable
improvement over FBP, its performance remains to be tested on real
biological tissues.
Figure 9.
Model-based image reconstruction of absolute optical absorption on a
numerical phantom with a 5 mm-diameter target. The optical source is
located at the bottom of the circle (red), and a ring-shaped
ultrasound transducer array is around the entire phantom (blue). The
target depth is varied from 10 mm to 40 mm, and the PAT images are
reconstructed by the FBP method (top row) and model-based method
(bottom row). Reproduced with permission from.[162]
Model-based image reconstruction of absolute optical absorption on a
numerical phantom with a 5 mm-diameter target. The optical source is
located at the bottom of the circle (red), and a ring-shaped
ultrasound transducer array is around the entire phantom (blue). The
target depth is varied from 10 mm to 40 mm, and the PAT images are
reconstructed by the FBP method (top row) and model-based method
(bottom row). Reproduced with permission from.[162]
Deep learning-based reconstruction methods
Model-based reconstruction methods, although more accurate than the FBP
method, are computationally expensive for practical use.[18,23] Deep learning, a rapidly emerging research area focusing on pattern
recognition and machine learning techniques, has recently found applications
for biomedical imaging.[164-166] In a typical deep-learning algorithm, the artificial neural network
represents the non-linear relationship between the input and output images,
and the algorithm performs various optimization processes at different
layers based on the training datasets.[167] Recently, several groups have adapted deep learning methods to
address the limited-view and under-sampling issues in PAT.[40,59,168-170] Here, we will focus on several examples for quantitative PAT
reconstruction.Bench et al have demonstrated the feasibility of 3D convolutional neural
networks to generate accurate sO2 maps, by using the full spatial
information contained in the 3D PA images.[171] Two convolutional encoder-decoder networks with skip connections
(EDS) were trained to output 3D sO2 and vessel segmentation maps.
However, the network accuracy decreased with depth, due to decreased SNR
and/or limited view.A residual learning framework (ResU-net) was also employed for PAT.[172] ResU-net takes PAT images acquired at different wavelengths as the
input and generates sO2 maps or molecular concentration images as
the output. The residual learning mechanism can prevent accuracy degradation
that commonly happens with increasing network depth. Figure 10A to C show the estimated
indocyanine green concentration using ResU-net, demonstrating great promise
for molecular imaging applications. However, its performance on animals
remains to be investigated.
Figure 10.
PAT image reconstruction based on deep learning. (a, b) ResU-net is
used to estimate the absolute concentrations of indocyanine green.
The reconstruction error relative to the true concentration is
<5%. (c) Normalized true molar extinction coefficient and the
measured PA spectrum of the rightmost target in (a-b).[172] (d-f) U-net based network is used to estimate the absorption
coefficients of the targets.[173] Reproduced with permission from.[172,173]
An U-net based architecture was developed for PAT to directly recover the
optical absorption from the deposited energy distribution.[173] The training data was generated from Monte Carlo simulation of light
transport with wide-field illumination. This method has been validated on
circular targets of different sizes and positions (Figure 10D-F). The relative error
with different absorption backgrounds was less than 10%. Again, more complex
experiments are necessary to demonstrate its feasibility for biological
applications.PAT image reconstruction based on deep learning. (a, b) ResU-net is
used to estimate the absolute concentrations of indocyanine green.
The reconstruction error relative to the true concentration is
<5%. (c) Normalized true molar extinction coefficient and the
measured PA spectrum of the rightmost target in (a-b).[172] (d-f) U-net based network is used to estimate the absorption
coefficients of the targets.[173] Reproduced with permission from.[172,173]
Novel Molecular Probes for Deep PAT
The use of molecular probes with high optical absorption is another way of
extending the penetration depth of PAT. Recently there have been significant
advancements in the formulation and evaluation of organic dyes and nanoparticles
that absorb light between 1000-1700 nm, which is known as the NIR-II region. The
NIR-II region is noteworthy due to it deeper tissue penetration with PAT, with
minimal interference from endogenous tissue components such as hemoglobin and
water. Several interesting PAT contrast agents have been developed with NIR-II
activity. Additionally, novel genetic probes and activation sensing probes have
shown their potential for PAT applications.
Inorganic probes
Inorganic probes have the advantage of being highly adjustable in size and
absorption bandwidth, and are typically formed from metals or unnatural
molecular structures. These customizations allow for alterations that can
modulate the optical properties and pharmacokinetic behaviors. Historically,
inorganic contrast agents have been used extensively in PAT. In particular,
gold and carbon nanotubes have shown great potential for strong PA contrast.
One application involved targeting the lymphatic system in mice with an
antibody conjugation to the gold plated nanotubes.[174] Another study conjugated a tumor specific peptide to the gold
nanorods on the surface of carbon nanotubes to target gastric cancer cells.[175] This technique is potentially generalizable to other types of cancer
cells.Gold nanoparticles have historically been used in the NIR-I region.[176] In the longer NIR-II region, improved signal to background ratio
(SBR) improves PA imaging.[177] One study with small gold nanorods that absorbed light in the NIR-II
region led to strong PA signal in tumors in mice as well as higher thermal
stability relative to larger counterparts. With small gold nanorods that
absorbed light in the NIR-II region, PAT can image tumors more effectively.
Figure 11 shows
that peptide targeting in combination with small gold nanorods can be used
to effectively image tumors with high contrast.
Figure 11.
PAT of targeted small and large gold nanorods (AuNRs) in a murine
model of prostate cancer. (a-d) Photographs (left) and PAT images
(right) of tumor-bearing mice with non-targeted large (a) and small
(b) AuNRs and with peptide targeted large (c) and small (d) AuNRs.
The PA signal intensities are displayed in color, overlaid with the
ultrasound images in gray for anatomical information. Reproduced
with permission from.[177]
PAT of targeted small and large gold nanorods (AuNRs) in a murine
model of prostate cancer. (a-d) Photographs (left) and PAT images
(right) of tumor-bearing mice with non-targeted large (a) and small
(b) AuNRs and with peptide targeted large (c) and small (d) AuNRs.
The PA signal intensities are displayed in color, overlaid with the
ultrasound images in gray for anatomical information. Reproduced
with permission from.[177]The customizability of inorganic probes makes them useful for mapping tumor
vasculature using quantum dots comprised of polymers and silver[178-180] as well as different polymer coated metal mixtures: Cd, Te, Mn, and Hg.[181,182] The main PAT application is imaging tumor angiogenesis. Quantum dots
can absorb in both the NIR-I and NIR-II regions, which presents the
potential for multispectral PA imaging.[179,180] The advantages of quantum dots for PA molecular imaging include high
stability, stronger absorption, and longer circulation times due to the
polymer coating and custom sizing to evade renal clearance. The main
disadvantage of quantum dots is their toxic side effects. Efforts to
minimize tissue accumulation must be realized before clinical translation.[183]
Organic dyes and nanoparticles
Organic dyes and nanoparticles are noteworthy probes because they tend to be
biocompatible, with the possibility of biodegradation. One formulation of
cyanine fluoroalkylphosphate salt (CyFaP), an NIR-II dye that achieves
imaging in deep tissues, is a micelle formed with surfactant stripping. This
formulation was successful used in PAT at a 12 cm depth through chicken
breast tissue as well as imaging through 5 cm of human breast tissue in
adult volunteers,[184] as shown in Figure
12.
Figure 12.
PA and US overlaid images of tube containing ss-CyFaP placed beneath
the breasts of 3 different human adult female volunteers with
indicated cup sizes. Reproduced with permission from.[184]
PA and US overlaid images of tube containing ss-CyFaP placed beneath
the breasts of 3 different human adult female volunteers with
indicated cup sizes. Reproduced with permission from.[184]New NIR-II organic dyes are rapidly developed in order to exploit the
advantages of deeper penetration, less tissue scattering, and permissible
use of higher laser powers of longer wavelength. Often, organic dyes need to
be solubilized with a surfactant of polymer matrix due to their hydrophobic nature.[185] Bright NIR-II signal is achievable using organic dyes,[186,187] some of which have fast renal clearance.[188] Modifications can be made to the dyes to narrow the absorbing
spectrum and modulate the biodistribution, which can minimize the background
signals in PAT. Another type of modification is the use of polymers in the
formation of J-aggregates, which can improve the PA signal generation.[189] Other dyes have been formulated with Pluronic micelles surfactant
stripping, which results in concentrated dye suspensions for strong PA
signal generation. For example, these micelles were used for PAT of mouse
gut following oral administration, as shown in Figure 13.[190]
Figure 13.
PAT of the mouse intestine after oral administration of frozen
micelles of naphthalocyanine dye. The depth was encoded in color.
Reproduced with permission from.[190]
PAT of the mouse intestine after oral administration of frozen
micelles of naphthalocyanine dye. The depth was encoded in color.
Reproduced with permission from.[190]
Genetic reporters
Genetically-encoded probes are usually developed for fluorescence imaging,
but have become increasingly attractive in PAT as biomarkers of tumors or
healthy cells.[191] With directed evolution, these probes can be modified for higher
photostability, larger absorption coefficient in the NIR region, and more
distinct spectra.[192,193] Genetically-encoded probes have good biocompatibility with much
reduced toxicity concerns. However, wavelength tunability and manipulation
of the absorption bandwidth is challenging. They often have relatively low
concentration in cells and high photobleaching rate. Some genetic reporters
can absorb strongly in the NIR-I region, but few if any have been developed
with NIR-II absorption.[194]One recent study reported PAT with red fluorescent protein (RFP) activated in
the presence of biliverdin, which is typically found in the liver and
spleen. It was evident this cofactor could also be directly injected into
the tumor for localized PA signal enhancement.[195] A similar concept has been developed that uses a viral delivery
system with cell receptor tyrosine that triggers the human cancer cells to
permanently express eumelanin. Using eumelanin as the endogenous contrast in
the NIR-I region, PAT can be used to image tumor growth over time, as shown
in Figure 14.[17]
Figure 14.
In vivo PAT of Tyr-expressing 293T cells acquired
at different times post-inoculation, illustrating cell population
growth. (a-d) Maximum amplitude projections of 3D image acquired on
day 7 (a), day 14 (b), day 22 (c) and day 26 post-inoculation (d).
Reproduced with permission from.[17]
In vivo PAT of Tyr-expressing 293T cells acquired
at different times post-inoculation, illustrating cell population
growth. (a-d) Maximum amplitude projections of 3D image acquired on
day 7 (a), day 14 (b), day 22 (c) and day 26 post-inoculation (d).
Reproduced with permission from.[17]One major hurdle in PAT of genetically encoded probes is the overwhelming
background signals from the abundant hemoglobin. Recently,
bacteriophytochromes (BphPs) have been reported as promising contrast for
molecular PAT that can drastically improve the detection sensitivity. BphPs
can be reversibly photoswitched between the Pfr (ON) and Pr (OFF) states,
which is induced by the photoisomerization of biliverdin IX located inside
the chromophore-binding pocket. The molar extinction coefficient of BphPs in
the Pfr state at 780 nm is ∼70-folds higher that of oxy-hemoglobin. By
repeatedly switching the BphPs between the ON and OFF states, the PA signals
from the BphPs can be reliably extracted from the non-switching background
hemoglobin signals.[196-200] Different photoswitchable probes can also be identified by using a
machine-learning-based unmixing method.[197]
Activatable probe
To improve the detection specificity, a useful strategy in PA molecular
imaging is the ability to activate a molecular probe under certain
biological and physiological conditions. For example, Ju et al have
demonstrated a nanoparticle formulation with a melanin core, whose PA signal
can be amplified by 8 times in mildly acid conditions.[201] This nanoparticle is useful for cancer imaging due to the lower pH
microenvironment inside a tumor. Wang et al have reported a new NIR probe
1-RGD that was activated by caspase-3, one of the chain events in apoptosis.
Activation of 1-RGD lead to monodispersed nanoparticles with strong PA
signal generation, which allowed for imaging tumor cell apoptosis.[202] Similarly, using plasmonic nanosensors activatable by epidermal
growth factor receptor, Luke et al was able to image metastatic cancer cells.[22]Another type of activatable PA probes can be used to record physiological
signals. For example, Zhang et al have developed a voltage-sensitive probe
that responds to the voltage stimulation via fluorescence quenching. This
probe has the potential to record action potentials in the brain using PAT.[203] Similarly, another synthesized PA probe can be activated by
Ca+2 and was able to measure electric potentials in the
heart, brain, and skeletal muscles.[204]
Conclusion and Outlook
PAT has been demonstrated as a powerful technology to provide reliable molecular
imaging. Various approaches have been explored to further improve PAT’s penetration
depth for molecular applications. Using microwave or X-ray as the radiation source
can greatly increase the penetration depth, but the clinical application is limited
to certain targets. NIR light can be applied in internal-illumination PAT or PA
endoscopy to image organs closely located to body cavities. Wavefront shaping
combined with ultrasensitive ultrasound transducers can further extend penetration
depth. In addition, fluence compensation and artifact reduction reconstruction
algorithms are actively applied to enhance PA images with weak signals.
Deep-learning-based reconstruction methods are growing rapidly and have shown great
promise to further improve the image quality at depths. Moreover, many types of
novel molecular probes have been extensively studied as PAT contrast, and their
strong optical absorption in the NIR region have resulted in improved penetration
depth.Future work in PA molecular imaging will likely include all of the above endeavors in
a more integrated, rational, and clinically-oriented manner. For instance,
internal-illumination PAT can be combined with advanced image reconstruction methods
and high-absorbing molecular probes to future improve the SNR and image quality.
Moreover, while many promising PA contrast agents have been developed, most of these
studies were carried out in small animal models. There is a strong need to test
these imaging paradigms in large animals. The costs for required sterile
manufacturing and toxicity testing are substantially higher than the small animal
applications, so rational selection of target indication is important. The clinical
translation of PA molecular imaging is still progressing slowly but the potential
impact is not beyond reach. For example, by using the FDA-approved indocyanine
green, it is possible to apply PAT for detecting tumor malignancy during surgery.[205] Another solution is the repurposing of natural materials for PA molecular
imaging. For example, roasted barley was found to have strong NIR-II PA signal and
can be used to monitor swallowing in humans.[206] These natural materials have the potential to be used for human imaging and
can be readily translated with minimized safety risk.
Authors: Haifeng Wang; Terry B Huff; Daniel A Zweifel; Wei He; Philip S Low; Alexander Wei; Ji-Xin Cheng Journal: Proc Natl Acad Sci U S A Date: 2005-10-20 Impact factor: 11.205
Authors: Nikolaos C Deliolanis; Angelique Ale; Stefan Morscher; Neal C Burton; Karin Schaefer; Karin Radrich; Daniel Razansky; Vasilis Ntziachristos Journal: Mol Imaging Biol Date: 2014-10 Impact factor: 3.488
Authors: Mucong Li; Nathan Beaumont; Chenshuo Ma; Juan Rojas; Tri Vu; Max Harlacher; Graeme O'Connell; Ryan C Gessner; Hailey Kilian; Ludmila Kasatkina; Yong Chen; Qiang Huang; Xiling Shen; Jonathan F Lovell; Vladislav V Verkhusha; Tomek Czernuszewicz; Junjie Yao Journal: IEEE Trans Med Imaging Date: 2022-09-30 Impact factor: 11.037