Thitirat Mantim1,2,3, Korbua Chaisiwamongkhol1,4,5, Kanchana Uraisin1,3, Peter C Hauser6, Prapin Wilairat1,7, Duangjai Nacapricha1,3.
Abstract
This work presents a new dual-purpose detector for photometric and conductivity measurements in flow-based analysis. The photometric detector is a paired emitter-detector diode (PEDD) device, whilst the conductivity detection employs a capacitively coupled contactless conductivity detector (C4D). The flow-through detection cell is a rectangular acrylic block (ca. 2 × 2 × 1.5 cm) with cylindrical channels in Z-configuration. For the PEDD detector, the LED light source and detector are installed inside the acrylic block. The two electrodes of the C4D are silver conducting ink painted on the PEEK inlet and outlet tubing of the Z-flow cell. The dual-purpose detector is coupled with a sequential injection analysis (SIA) system for simultaneous detection of the absorbance of the orange dye and conductivity of the dissolved oral rehydration salt powder. The detector was also used for sequential measurements of creatinine and the conductivity of human urine samples. The creatinine analysis is based on colorimetric detection of the Jaffé reaction using the PEDD detector, and the conductivity of the urine, as measured by the C4D detector, is expressed in millisiemens (mS cm-1).Entities:
Keywords: contactless conductivity detector; flow-based analysis; paired emitter–detector diode detector; sequential detection; simultaneous detection
Mesh:
Year: 2020 PMID: 32414012 PMCID: PMC7287826 DOI: 10.3390/molecules25102284
Source DB: PubMed Journal: Molecules ISSN: 1420-3049 Impact factor: 4.411
Figure 1(a) Schematics of C4D-PEDD flow-through detection cell for the “2-in-1” detector for simultaneous detection. (b) Photograph of the “2-in-1” C4D-PEDD flow-through detection cell and the pre-amplifier board.
Figure 2Schematic diagram of the SIA system coupled with the “2-in-1” flow cell for dual C4D-PEDD detection. Note: For the analysis of oral rehydration salt (ORS) samples, port 4 (labelled R) is not used. Port 4 is used as the inlet for the alkaline picrate reagent (R) in the sequential measurements of creatinine and conductivity of urine samples. SV—switching valve, SP—syringe pump, HC—holding coil, W—waste.
Figure 3The series of signals recorded simultaneously by the C4D (top) and PEDD (bottom) detectors for triplicate injections of aqueous standard solutions in increasing order of concentrations and then in decreasing order, as carried out using SIA system (see Figure 2). The concentration of NaCl and the absorbance at 525 nm of orange dye in each standard solution: (i) 5 mM and 0.1124, (ii) 10 mM and 0.2228, (iii) 20 mM and 0.4466, (iv) 30 mM and 0.6657, (v) 40 mM and 0.8844 and (vi) 50 mM and 1.0332.
Figure 4Bar graphs of the mean conductivity (mS cm−1) of each sachet (dose) and the mean absorbances of 150.0 mL solutions for Brand A or 120.0 mL solutions of Brand B, with error bars (n = 3). The red text shows the means and standard deviations for samples A1–A8 and for samples B1–B8, respectively. The red dotted lines indicate the range for the mean ± 2SD of these two products. (a) Eight samples from “Brand A” for adults, labelled net weight of 3.645 g. (b) Eight samples from “Brand B” for children, labelled net weight of 3.3 g.
Figure 5Measurements of urine conductivity and determination of creatinine using SIA equipped with the “2-in-1” C4D-PEDD detector. (a) Zone sequence of sample and reagent segments with aspirated volumes in microliters (D: detector). (b) Peak profile of urine conductivity (Con) from C4D. The large peak is due to the conductivity of the ions in the mixed zones of Sequence 2. (c) Peak profile of creatinine signal (Cre) from PEDD.
Analysis of conductivity and creatinine of human urine samples.
| Sample | Conductivity (mS cm−1) | Creatinine (mg L−1) * | |
|---|---|---|---|
| C4D Detector ( | PEDD Detector ( | Spectrophometer ( | |
| S1 | 19.18 ± 0.03 | (1.06 ± 0.02) × 103 | (1.08 ± 0.01) × 103 |
| S2 | 27.10 ± 0.12 | (1.40 ± 0.02) × 103 | (1.41 ± 0.03) × 103 |
| S3 | 26.57 ± 0.05 | (1.54 ± 0.01) × 103 | (1.56 ± 0.02) × 103 |
| S4 | 31.60 ± 0.01 | (1.17 ± 0.02) × 103 | (1.14 ± 0.02) × 103 |
| S5 | 12.42 ± 0.02 | (7.36 ± 0.05) × 102 | (7.13 ± 0.07) × 102 |
| S6 | 10.07 ± 0.02 | (5.60 ± 0.05) × 102 | (5.06 ± 0.05) × 102 |
| S7 | 4.99 ± 0.02 | (1.52 ± 0.04) × 102 | (1.22 ± 0.01) × 102 |
| S8 | 11.96 ± 0.06 | (4.39 ± 0.06) × 102 | (4.09 ± 0.06) × 102 |
* Paired t-test: tstat = 2.09, tcrit = 2.45.