Weena Siangproh1, Norio Teshima, Tadao Sakai, Shuji Katoh, Orawon Chailapakul.
Abstract
A simple, automatic and practical system for successive determination of albumin and creatinine has been developed by combining sequential injection analysis (SIA) and highly sensitive dye-binding assays. Albumin detection was based on the increase in the absorbance due to complex formation between albumin and eosin Y in acidic media. The absorbance of the complex was monitored at 547 nm. For the creatinine assay, the concentration of creatinine was measured by reaction with alkaline picrate to form a colored product which absorbs at 500 nm. The influences of experimental variables such as effects of pH, reagent concentration, standard/sample volume and interferences were investigated. Under optimal conditions, the automated method showed linearity up to 20 mg L(-1) for albumin and 100 mg L(-1) for creatinine. The 3 sigma detection limits were 0.6 and 3.5 mg L(-1) for albumin and creatinine, respectively, and the relative standard deviations (n=10) were 2.49% for 20 mg L(-1) albumin, and 3.14% for 20 mg L(-1) creatinine. Application of the proposed method to the direct analysis of urinary samples yielded results which agreed with those obtained from the Bradford protein assay and a creatinine enzymatic assay according to a paired t-test. The results obtained should be a step towards developing a fully automated and reliable analytical system for clinical research, which requires direct determination of albumin and creatinine and/or its ratios.Entities:
Mesh:
Substances:
Year: 2009 PMID: 19615518 DOI: 10.1016/j.talanta.2008.12.068
Source DB: PubMed Journal: Talanta ISSN: 0039-9140 Impact factor: 6.057