| Literature DB >> 32108157 |
Katsunori Yogo1,2, Akihiro Matsushita3, Yuya Tatsuno3, Takahiro Shimo4, Seiko Hirota5, Marika Nozawa6, Shuichi Ozawa7,8, Hiromichi Ishiyama3,6, Hiroshi Yasuda5, Yasushi Nagata7,8, Kazushige Hayakawa3,6.
Abstract
With advances in high-dose-rate (HDR) brachytherapy, the importance of quality assurance (QA) is increasing to ensure safe delivery of the treatment by measuring dose distribution and positioning the source with much closer intervals for highly active sources. However, conventional QA is time-consuming, involving the use of several different measurement tools. Here, we developed simple QA method for HDR brachytherapy based on the imaging of Cherenkov emission and evaluated its performance. Light emission from pure water irradiated by an 192Ir γ-ray source was captured using a charge-coupled device camera. Monte Carlo calculations showed that the observed light was primarily Cherenkov emissions produced by Compton-scattered electrons from the γ-rays. The uncorrected Cherenkov light distribution, which was 5% on average except near the source (within 7 mm from the centre), agreed with the dose distribution calculated using the treatment planning system. The accuracy was attributed to isotropic radiation and short-range Compton electrons. The source positional interval, as measured from the light images, was comparable to the expected intervals, yielding spatial resolution similar to that permitted by conventional film measurements. The method should be highly suitable for quick and easy QA investigations of HDR brachytherapy as it allows simultaneous measurements of dose distribution, source strength, and source position using a single image.Entities:
Year: 2020 PMID: 32108157 PMCID: PMC7046619 DOI: 10.1038/s41598-020-60519-z
Source DB: PubMed Journal: Sci Rep ISSN: 2045-2322 Impact factor: 4.379
Figure 1(a) Schematic (a) and photograph (b) of experimental setup. The dashed line on the lead shield shows the aperture of the lead block.
Figure 2Images of light emission from water irradiated with 192Ir source. (a) Image of light from pure water and catheter irradiated with source. (b) Image of light from the catheter in air when irradiated with source. (c) Image of light from pure water irradiated with source after the catheter was covered with tape. Scale bar: 5 mm. Exposure time was 58 s. Images are expressed in grey values in 16-bit scales as in colour bars.
Figure 3Representative profiles of light emissions from water irradiated with 192Ir source. (a) Light from pure water irradiated with source after catheter was covered with black tape (water). Data are compared with those from light from uncovered catheter in water (all) and light from catheter in air when irradiated with source (catheter). (b) Measured light emission data compared with those from Monte Carlo (MC) calculations. Data are normalised relative to those at the position = 10 mm. The data are zoomed in the inset.
Figure 4Characteristics of measured light intensity of water. (a) Light intensity as a function of dose. (b) Percentage error of intensity as a function of dose. Percentage errors are calculated relative to the averaged values for four repeated measurements, i.e., (standard deviation)/(averaged value). Dose presented are accumulated considering water thickness, t = 100 mm. Water thickness is the perpendicular distance between the source and the water tank wall near the camera.
Figure 5Comparison of light intensity profiles calibrated to dose with dose distributions calculated using TPS along directions (a) perpendicular and (b) parallel to source movement. Calibration to dose are performed by a fit of Cherenkov vs dose at r = 10 mm as in Fig. 4(a). Dose presented are accumulated considering water thickness, t = 100 mm. Water thickness is the perpendicular distance between the source and the water tank wall near the camera. The data are zoomed in the insets.
Figure 6Histogram of range of electrons produced by γ-rays and of β-rays from Ir source. Calculations were performed using Monte Carlo methods (GEANT4).
Figure 7Source strength measured as light intensity (a) Decay curve of measured Cherenkov light intensity. A decay curve of 192Ir (half-life T1/2 = 73.83 days) is presented for comparison. The reference source strength was measured using a well-type chamber (chamber). (b) Relationship between light intensity and measured dose rate of brachytherapy source. Error bars represent standard deviations for four repeated readings. All measurements are normalised with respect to averaged intensity over all data (n = 12).
Figure 8Source position measurements. Profiles of light from pure water compared to those from film measurements: (a) preset interval = 40 mm and (b) preset interval = 2.5 mm. Note that the range of position is expanded in (b) for clarification.