Radiochromic film dosimetry of HDR (192)Ir source radiation fields.

Saad Aldelaijan1, Huriyyah Mohammed, Nada Tomic, Li-Heng Liang, Francois Deblois, Arman Sarfehnia, Wamied Abdel-Rahman, Jan Seuntjens, Slobodan Devic.   

Abstract

PURPOSE: A radiochromic film based dosimetry system for high dose rate (HDR) Iridium-192 brachytherapy source was described. A comparison between calibration curves established in water and Solid Water™ was provided.
METHODS: Pieces of EBT-2 model GAFCHROMIC™ film were irradiated in both water and Solid Water™ with HDR (192)Ir brachytherapy source in a dose range from 0 to 50 Gy. Responses of EBT-2 GAFCHROMIC™ film were compared for irradiations in water and Solid Water™ by scaling the dose between media through Monte Carlo calculated conversion factor for both setups. To decrease uncertainty in dose delivery due to positioning of the film piece with respect to the radiation source, traceable calibration irradiations were performed in a parallel-opposed beam setup.
RESULTS: The EBT-2 GAFCHROMIC™ film based dosimetry system described in this work can provide an overall one-sigma dose uncertainty of 4.12% for doses above 1 Gy. The ratio of dose delivered to the sensitive layer of the film in water to the dose delivered to the sensitive layer of the film in Solid Water™ was calculated using Monte Carlo simulations to be 0.9941 ± 0.0007.
CONCLUSIONS: A radiochromic film based dosimetry system using only the green color channel of a flatbed document scanner showed superior precision if used alone in a dose range that extends up to 50 Gy, which greatly decreases the complexity of work. In addition, Solid Water™ material was shown to be a viable alternative to water in performing radiochromic film based dosimetry with HDR (192)Ir brachytherapy sources.

Entities:  

Mesh:

Substances:

Year:  2011        PMID: 22047372     DOI: 10.1118/1.3651482

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  9 in total

1.  Surface dose characterisation of the Varian Ir-192 HDR conical surface applicator set with a vertically orientated source.

Authors:  Konrad Buchauer; Guido Henke; Hans Schiefer; Ludwig Plasswilm
Journal:  Strahlenther Onkol       Date:  2014-08-01       Impact factor: 3.621

2.  Robustness of single-isocenter multiple-metastasis stereotactic radiosurgery end-to-end testing across institutions.

Authors:  Daniel Saenz; Niko Papanikolaou; Emmanouil Zoros; Evangelos Pappas; Michael Reiner; Lip Teck Chew; Hooi Yin Lim; Sam Hancock; Alex Nebelsky; Christopher Njeh; Georgios Anagnostopoulos
Journal:  J Radiosurg SBRT       Date:  2021

3.  Linearization of dose-response curve of the radiochromic film dosimetry system.

Authors:  Slobodan Devic; Nada Tomic; Saad Aldelaijan; Francois Deblois; Jan Seuntjens; Maria F Chan; Dave Lewis
Journal:  Med Phys       Date:  2012-08       Impact factor: 4.506

4.  Dose error from deviation of dwell time and source position for high dose-rate 192Ir in remote afterloading system.

Authors:  Hiroyuki Okamoto; Ako Aikawa; Akihisa Wakita; Kotaro Yoshio; Naoya Murakami; Satoshi Nakamura; Minoru Hamada; Yoshihisa Abe; Jun Itami
Journal:  J Radiat Res       Date:  2014-02-23       Impact factor: 2.724

5.  Film based verification of calculation algorithms used for brachytherapy planning-getting ready for upcoming challenges of MBDCA.

Authors:  Grzegorz Zwierzchowski; Grzegorz Bielęda; Janusz Skowronek; Magdalena Mazur
Journal:  J Contemp Brachytherapy       Date:  2016-08-16

6.  Verification of high-dose-rate brachytherapy treatment planning dose distribution using liquid-filled ionization chamber array.

Authors:  A B Mohamed Yoosuf; Prakash Jeevanandam; Glenn Whitten; Geraldine Workman; Conor K McGarry
Journal:  J Contemp Brachytherapy       Date:  2018-04-30

7.  Imaging Cherenkov emission for quality assurance of high-dose-rate brachytherapy.

Authors:  Katsunori Yogo; Akihiro Matsushita; Yuya Tatsuno; Takahiro Shimo; Seiko Hirota; Marika Nozawa; Shuichi Ozawa; Hiromichi Ishiyama; Hiroshi Yasuda; Yasushi Nagata; Kazushige Hayakawa
Journal:  Sci Rep       Date:  2020-02-27       Impact factor: 4.379

8.  Evaluating dosimetric accuracy of the 6 MV calibration on EBT3 film in the use of Ir-192 high dose rate brachytherapy.

Authors:  Lyu Huang; Hani Gaballa; Jenghwa Chang
Journal:  J Appl Clin Med Phys       Date:  2022-02-28       Impact factor: 2.243

9.  Physics-aspects of dose accuracy in high dose rate (HDR) brachytherapy: source dosimetry, treatment planning, equipment performance and in vivo verification techniques.

Authors:  Antony Palmer; David Bradley; Andrew Nisbet
Journal:  J Contemp Brachytherapy       Date:  2012-06-30
  9 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.