The feasibility study and characterization of a two-dimensional diode array in "magic phantom" for high dose rate brachytherapy quality assurance.

A Espinoza1, B Beeksma, M Petasecca, I Fuduli, C Porumb, D Cutajar, S Corde, M Jackson, M L F Lerch, A B Rosenfeld.   

Abstract

PURPOSE: High dose rate (HDR) brachytherapy is a radiation treatment technique capable of delivering large dose rates to the tumor. Radiation is delivered using remote afterloaders to drive highly active sources (commonly (192)Ir with an air KERMA strength range between 20,000 and 40,000 U, where 1 U = 1 μGy m(2)/h in air) through applicators directly into the patient's prescribed region of treatment. Due to the obvious ramifications of incorrect treatment while using such an active source, it is essential that there are methods for quality assurance (QA) that can directly and accurately verify the treatment plan and the functionality of the remote afterloader. This paper describes the feasibility study of a QA system for HDR brachytherapy using a phantom based two-dimensional 11 × 11 epitaxial diode array, named "magic phantom."
METHODS: The HDR brachytherapy treatment plan is translated to the phantom with two rows of 10 (20 in total) HDR source flexible catheters, arranged above and below the diode array "magic plate" (MP). Four-dimensional source tracking in each catheter is based upon a developed fast iterative algorithm, utilizing the response of the diodes in close proximity to the (192)Ir source, sampled at 100 ms intervals by a fast data acquisition (DAQ) system. Using a (192)Ir source in a solid water phantom, the angular response of the developed epitaxial diodes utilized in the MP and also the variation of the MP response as a function of the source-to-detector distance (SDD) were investigated. These response data are then used by an iterative algorithm for source dwelling position determination. A measurement of the average transit speed between dwell positions was performed using the diodes and a fast DAQ.
RESULTS: The angular response of the epitaxial diode showed a variation of 15% within 360°, with two flat regions above and below the detector face with less than 5% variation. For SDD distances of between 5 and 30 mm the relative response of the epitaxial diodes used in the MP is in good agreement (within 8%) with radial dose function measurements found within the TG-43 protocol, with SDD of up to 70 mm showing a 40% over response. A method for four-dimensional localization of the HDR source was developed, allowing the source dwell position to be derived within 0.50 mm of the expected position. An estimation of the average transit speed for varying step sizes was determined and was found to increase from (12.8 ± 0.3) up to (38.6 ± 0.4) cm/s for a step size of 2.5 and 50 mm, respectively.
CONCLUSIONS: Our characterization of the designed QA "magic phantom" with MP in realistic HDR photon fields demonstrates the promising performance for real-time source position tracking in four dimensions and measurements of transit times. Further development of this system will allow a full suite for QA in HDR brachytherapy and analysis, and for future in vivo tracking.

Entities:  

Mesh:

Substances:

Year:  2013        PMID: 24320410     DOI: 10.1118/1.4822736

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  6 in total

Review 1.  In vivo dosimetry: trends and prospects for brachytherapy.

Authors:  G Kertzscher; A Rosenfeld; S Beddar; K Tanderup; J E Cygler
Journal:  Br J Radiol       Date:  2014-07-08       Impact factor: 3.039

2.  Verification of high-dose-rate brachytherapy treatment planning dose distribution using liquid-filled ionization chamber array.

Authors:  A B Mohamed Yoosuf; Prakash Jeevanandam; Glenn Whitten; Geraldine Workman; Conor K McGarry
Journal:  J Contemp Brachytherapy       Date:  2018-04-30

3.  HDR brachytherapy in vivo source position verification using a 2D diode array: A Monte Carlo study.

Authors:  Joel Poder; Dean Cutajar; Susanna Guatelli; Marco Petasecca; Andrew Howie; Joseph Bucci; Anatoly Rosenfeld
Journal:  J Appl Clin Med Phys       Date:  2018-06-01       Impact factor: 2.102

4.  Imaging Cherenkov emission for quality assurance of high-dose-rate brachytherapy.

Authors:  Katsunori Yogo; Akihiro Matsushita; Yuya Tatsuno; Takahiro Shimo; Seiko Hirota; Marika Nozawa; Shuichi Ozawa; Hiromichi Ishiyama; Hiroshi Yasuda; Yasushi Nagata; Kazushige Hayakawa
Journal:  Sci Rep       Date:  2020-02-27       Impact factor: 4.379

Review 5.  In vivo dosimetry in brachytherapy: Requirements and future directions for research, development, and clinical practice.

Authors:  Gabriel P Fonseca; Jacob G Johansen; Ryan L Smith; Luc Beaulieu; Sam Beddar; Gustavo Kertzscher; Frank Verhaegen; Kari Tanderup
Journal:  Phys Imaging Radiat Oncol       Date:  2020-09-28

6.  End-to-end test for computed tomography-based high-dose-rate brachytherapy.

Authors:  Fabian Krause; Franziska Risske; Susann Bohn; Marc Delaperriere; Jürgen Dunst; Frank-André Siebert
Journal:  J Contemp Brachytherapy       Date:  2018-12-28
  6 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.