| Literature DB >> 21897562 |
A Manikandan1, Sarkar Biplab, Perianayagam A David, R Holla, T R Vivek, N Sujatha.
Abstract
For high dose rate (HDR) brachytherapy, independent treatment verification is needed to ensure that the treatment is performed as per prescription. This study demonstrates dosimetric quality assurance of the HDR brachytherapy using a commercially available two-dimensional ion chamber array called IMatriXX, which has a detector separation of 0.7619 cm. The reference isodose length, step size, and source dwell positional accuracy were verified. A total of 24 dwell positions, which were verified for positional accuracy gave a total error (systematic and random) of -0.45 mm, with a standard deviation of 1.01 mm and maximum error of 1.8 mm. Using a step size of 5 mm, reference isodose length (the length of 100% isodose line) was verified for single and multiple catheters of same and different source loadings. An error ≤1 mm was measured in 57% of tests analyzed. Step size verification for 2, 3, 4, and 5 cm was performed and 70% of the step size errors were below 1 mm, with maximum of 1.2 mm. The step size ≤1 cm could not be verified by the IMatriXX as it could not resolve the peaks in dose profile.Entities:
Keywords: 2D array; IMatriXX; brachy
Year: 2011 PMID: 21897562 PMCID: PMC3159223 DOI: 10.4103/0971-6203.83491
Source DB: PubMed Journal: J Med Phys ISSN: 0971-6203
Figure 1Position of catheters over the detector array
Comparison of planned and measured dwell positions
Figure 2The position of the point where the dose was considered as 100%
Figure 3Reference isodose length verification
Reference isodose length verification for same source loading of 7 cm
Reference Isodose length verification for different source loadings
Figure 4Planned versus measured dwell position
Figure 5Planned versus measured reference isodose length
Figure 6The step size analysis for 2 cm step size
Figure 7Dose profile of 1 mm shifted dwell positions. It is clear from the graph that peaks were differentiated with each other