A quality assurance tool for high-dose-rate brachytherapy.

Daniel W Rickey1, David Sasaki, Jeff Bews.   

Abstract

PURPOSE: The purpose of this work was to develop a quality assurance (QA) tool for high-dose-rate (HDR) brachytherapy that would quickly and easily verify both source positioning (dwell positions) and durations (dwell times).
METHODS: The authors constructed a QA tool that combined radiochromic film to verify position with four photodiode detectors to verify dwell times. To characterize the temporal accuracy of the tool, a function generator powered four red light-emitting diodes that were optically coupled to the four photodiode detectors. The QA tool was used to verify the dwell positions and times of a commercial brachytherapy afterloader. Measurements of dwell time were independently verified by a one-dimensional optical camera that acquired 1000 lines/s.
RESULTS: The temporal accuracy of the QA tool was found to be about 1 ms. For visual assessment, the source position could be located within about 0.5 mm. Evaluating the accuracy and precision of an HDR brachytherapy afterloader, the authors found that the bias in dwell time can exceed 60 ms and the dwell time associated with the first dwell position had an unexpectedly large standard deviation of 30 ms. They found that the source locations were much easier to locate on the film if a plastic catheter was used instead of a metal treatment tube. Scanning the films enabled the dwell positions to be determined within about 0.2 mm.
CONCLUSIONS: For pretreatment QA, the authors found that this tool allowed verification of dwell positions and dwell times in about 6 min.

Mesh:

Year:  2010        PMID: 20632564     DOI: 10.1118/1.3425786

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  9 in total

1.  Relative dosimetrical verification in high dose rate brachytherapy using two-dimensional detector array IMatriXX.

Authors:  A Manikandan; Sarkar Biplab; Perianayagam A David; R Holla; T R Vivek; N Sujatha
Journal:  J Med Phys       Date:  2011-07

2.  A new approach to measure dwell position inaccuracy in HDR ring applicators - quantification and corrective QA.

Authors:  Abdul Qadir Jangda; Sherali Hussein; Zaka Rehman
Journal:  J Appl Clin Med Phys       Date:  2010-09-09       Impact factor: 2.102

3.  Imaging Cherenkov emission for quality assurance of high-dose-rate brachytherapy.

Authors:  Katsunori Yogo; Akihiro Matsushita; Yuya Tatsuno; Takahiro Shimo; Seiko Hirota; Marika Nozawa; Shuichi Ozawa; Hiromichi Ishiyama; Hiroshi Yasuda; Yasushi Nagata; Kazushige Hayakawa
Journal:  Sci Rep       Date:  2020-02-27       Impact factor: 4.379

4.  Development of an HDR-BT QA tool for source position verification.

Authors:  Yu Kumazaki; Ryuta Hirai; Mitsunobu Igari; Nao Kobayashi; Shohei Okazaki; Takanori Abe; Tomoaki Tamaki; Shin-Ei Noda; Shingo Kato
Journal:  J Appl Clin Med Phys       Date:  2020-11-02       Impact factor: 2.102

5.  In Vivo Verification of Treatment Source Dwell Times in Brachytherapy of Postoperative Endometrial Carcinoma: A Feasibility Study.

Authors:  Antonio Herreros; José Pérez-Calatayud; Facundo Ballester; Jose Barrera-Gómez; Rosa Abellana; Joana Melo; Luis Moutinho; Luca Tagliaferri; Ángeles Rovirosa
Journal:  J Pers Med       Date:  2022-05-31

6.  Automated multi-parameter high-dose-rate brachytherapy quality assurance via radioluminescence imaging.

Authors:  Mengyu Jia; Tae Jin Kim; Yong Yang; Lei Xing; Paul De Jean; Elliot Grafil; Cesare H Jenkins; Benjamin P Fahimian
Journal:  Phys Med Biol       Date:  2020-11-17       Impact factor: 3.609

Review 7.  Review of clinical brachytherapy uncertainties: analysis guidelines of GEC-ESTRO and the AAPM.

Authors:  Christian Kirisits; Mark J Rivard; Dimos Baltas; Facundo Ballester; Marisol De Brabandere; Rob van der Laarse; Yury Niatsetski; Panagiotis Papagiannis; Taran Paulsen Hellebust; Jose Perez-Calatayud; Kari Tanderup; Jack L M Venselaar; Frank-André Siebert
Journal:  Radiother Oncol       Date:  2013-11-30       Impact factor: 6.280

8.  COMP report: CPQR technical quality control guidelines for brachytherapy remote afterloaders.

Authors:  Normand Frenière
Journal:  J Appl Clin Med Phys       Date:  2018-02-07       Impact factor: 2.102

9.  End-to-end test for computed tomography-based high-dose-rate brachytherapy.

Authors:  Fabian Krause; Franziska Risske; Susann Bohn; Marc Delaperriere; Jürgen Dunst; Frank-André Siebert
Journal:  J Contemp Brachytherapy       Date:  2018-12-28
  9 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.