Literature DB >> 24320414

Source position verification and dosimetry in HDR brachytherapy using an EPID.

R L Smith1, M L Taylor, L N McDermott, A Haworth, J L Millar, R D Franich.   

Abstract

PURPOSE: Accurate treatment delivery in high dose rate (HDR) brachytherapy requires correct source dwell positions and dwell times to be administered relative to each other and to the surrounding anatomy. Treatment delivery inaccuracies predominantly occur for two reasons: (i) anatomical movement or (ii) as a result of human errors that are usually related to incorrect implementation of the planned treatment. Electronic portal imaging devices (EPIDs) were originally developed for patient position verification in external beam radiotherapy and their application has been extended to provide dosimetric information. The authors have characterized the response of an EPID for use with an (192)Ir brachytherapy source to demonstrate its use as a verification device, providing both source position and dosimetric information.
METHODS: Characterization of the EPID response using an (192)Ir brachytherapy source included investigations of reproducibility, linearity with dose rate, photon energy dependence, and charge build-up effects associated with exposure time and image acquisition time. Source position resolution in three dimensions was determined. To illustrate treatment verification, a simple treatment plan was delivered to a phantom and the measured EPID dose distribution compared with the planned dose.
RESULTS: The mean absolute source position error in the plane parallel to the EPID, for dwells measured at 50, 100, and 150 mm source to detector distances (SDD), was determined to be 0.26 mm. The resolution of the z coordinate (perpendicular distance from detector plane) is SDD dependent with 95% confidence intervals of ± 0.1, ± 0.5, and ± 2.0 mm at SDDs of 50, 100, and 150 mm, respectively. The response of the EPID is highly linear to dose rate. The EPID exhibits an over-response to low energy incident photons and this nonlinearity is incorporated into the dose calibration procedure. A distance (spectral) dependent dose rate calibration procedure has been developed. The difference between measured and planned dose is less than 2% for 98.0% of pixels in a two-dimensional plane at an SDD of 100 mm.
CONCLUSIONS: Our application of EPID dosimetry to HDR brachytherapy provides a quality assurance measure of the geometrical distribution of the delivered dose as well as the source positions, which is not possible with any current HDR brachytherapy verification system.

Entities:  

Mesh:

Substances:

Year:  2013        PMID: 24320414     DOI: 10.1118/1.4823758

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  4 in total

Review 1.  In vivo dosimetry: trends and prospects for brachytherapy.

Authors:  G Kertzscher; A Rosenfeld; S Beddar; K Tanderup; J E Cygler
Journal:  Br J Radiol       Date:  2014-07-08       Impact factor: 3.039

2.  Camera selection for real-time in vivo radiation treatment verification systems using Cherenkov imaging.

Authors:  Jacqueline M Andreozzi; Rongxiao Zhang; Adam K Glaser; Lesley A Jarvis; Brian W Pogue; David J Gladstone
Journal:  Med Phys       Date:  2015-02       Impact factor: 4.071

3.  Imaging Cherenkov emission for quality assurance of high-dose-rate brachytherapy.

Authors:  Katsunori Yogo; Akihiro Matsushita; Yuya Tatsuno; Takahiro Shimo; Seiko Hirota; Marika Nozawa; Shuichi Ozawa; Hiromichi Ishiyama; Hiroshi Yasuda; Yasushi Nagata; Kazushige Hayakawa
Journal:  Sci Rep       Date:  2020-02-27       Impact factor: 4.379

Review 4.  In vivo dosimetry in brachytherapy: Requirements and future directions for research, development, and clinical practice.

Authors:  Gabriel P Fonseca; Jacob G Johansen; Ryan L Smith; Luc Beaulieu; Sam Beddar; Gustavo Kertzscher; Frank Verhaegen; Kari Tanderup
Journal:  Phys Imaging Radiat Oncol       Date:  2020-09-28
  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.