| Literature DB >> 31858288 |
Magdy Algowhary1, Salma Taha2, Hosam Hasan-Ali2, Akihiko Matsumura3.
Abstract
BACKGROUND: What happens to stent length when deployed in a coronary artery? It is the aim of this study.Entities:
Keywords: IVUS; Stent length; Stent shortening
Year: 2019 PMID: 31858288 PMCID: PMC6923296 DOI: 10.1186/s43044-019-0036-9
Source DB: PubMed Journal: Egypt Heart J ISSN: 1110-2608
Clinical data
| Patients ( | ||
|---|---|---|
| Age (years) | 66.6 ± 10.0 | |
| Sex (%) | Men | 74 (82.2) |
| Women | 16 (17.8) | |
| Smokers (%) | 51 (56.7) | |
| Hypertension (%) | 60 (66.7) | |
| Diabetes mellitus (%) | 32 (35.6) | |
| Dyslipidemia (%) | 61 (67.8) | |
| Family history (%) | 14 (15.6) | |
| Chronic stable angina pectoris (%) | 39 (43.3) | |
| Unstable angina pectoris (%) | 29 (32.2) | |
| Acute myocardial infarction (%) | 22 (24.4) | |
| Ejection fraction | 0.62 ± 0.12 | |
| Statin (%) | 52 (57.8) | |
| β-Blocker (%) | 27 (30) | |
| Calcium antagonist (%) | 39 (43.3) | |
| ACE inhibitor (%) | 21 (23.3) | |
| ARB (%) | 32 (35.5) | |
| Cholesterol, mg/dl | 190.06 ± 35.94 | |
| LDL cholesterol, mg/dl | 120.28 ± 28.34 | |
| HDL cholesterol, mg/dl | 48.44 ± 13.34 |
Data are given as mean ± SD or percentage
ACE angiotensin-converting enzyme, ARB angiotensin receptor blocker, β-blocker beta-blocker, HDL cholesterol high-density lipoprotein cholesterol, LDL cholesterol low-density lipoprotein cholesterol
Stenting data
| Lesions ( | ||
|---|---|---|
| Lesion site (%) | LAD | 55 (57.9) |
| LCX | 18 (18.9) | |
| RCA | 22 (23.2) | |
| Lesion typea (%) | Low risk | 41 (43.2) |
| Moderate risk | 46 (48.4) | |
| High risk | 8 (8.4) | |
| Lesion length, mm | 9.09 (7.0–13.25) | |
| Reference diameter, mm | 3.13 ± 0.64 | |
| Prestent MLD, mm | 0.82 ± 0.32 | |
| Prestent diameter stenosis (%) | 63.06 ± 14.87 | |
| Poststent MLD, mm | 2.56 ± 0.44† | |
| Poststent diameter stenosis (%) | 10.49 ± 5.50† | |
| Drug-eluting stent (%) | 38 (40) | |
| Bare metal stent (%) | 57 (60) | |
| Stent diameter, mm (%) | 2.5 mm | 14 (14.74) |
| 3.0 mm | 54 (56.84) | |
| 3.5 mm | 23 (24.21) | |
| 4.0 mm | 4 (4.21) | |
| Stent length, mm (%) | 8.0 mm | 2 (2.1) |
| 9.0 mm | 2 (2.1) | |
| 12.0 mm | 5 (5.3) | |
| 13.0 mm | 4 (4.2) | |
| 15.0 mm | 18 (18.9) | |
| 18.0 mm | 32 (33.7) | |
| 20.0 mm | 3 (3.2) | |
| 23.0 mm | 18 (18.9) | |
| 24.0 mm | 3 (3.2) | |
| 25.0 mm | 1 (1.1) | |
| 28.0 mm | 6 (6.3) | |
| 30.0 mm | 1 (1.1) | |
| Stent strut (wire) thickness, mm | 0.14 (13.0–16.0) | |
| Stent deployment pressure, atm | 16.0 (14.0–18.0) | |
| Poststent balloon dilatation (%) | 19 (20%) |
Data are given as mean ± SD, percentage, or median (IQR)
LAD left anterior descending artery, LCX left circumflex artery, MLD minimal lumen diameter, RCA right coronary artery
aAmerican College of Cardiology/American Heart Association (ACC/AHA) classification [10]
†p < 0.0001 (compared with the prestent measurements)
IVUS data
| Lesion segment | Prestenting | Poststenting |
|---|---|---|
| Calcified plaque-media (%) | 41 (43.16) | |
| Lumen area, mm2 | 4.33 ± 2.00 | 7.91 ± 2.48* |
| Minimal lumen diameter, mm | 2.05 ± 0.47 | 2.91 ± 0.46* |
| Maximal lumen diameter, mm | 2.56 ± 0.57 | 3.45 ± 0.52* |
| EEM area, mm2 | 14.06 ± 4.37 | 16.84 ± 4.61* |
| Minimal EEM diameter, mm | 3.94 ± 0.17 | 4.32 ± 0.58* |
| Maximal EEM diameter, mm | 4.52 ± 0.66 | 4.94 ± 0.68* |
| Plaque-media area, mm2 | 9.74 ± 3.76 | 8.93 ± 3.21† |
| Proximal reference segment | ||
| Calcified plaque-media (%) | 32 (33.68) | |
| Lumen area, mm2 | 9.03 ± 3.29 | 8.77 ± 2.99 |
| EEM area, mm2 | 17.29 ± 5.05 | 17.82 ± 5.01† |
| Plaque-media area, mm2 | 8.26 ± 3.37 | 9.05 ± 3.56* |
| Distal reference segment | ||
| Calcified plaque-media (%) | 41 (43.16) | |
| Lumen area, mm2 | 6.30 ± 2.07 | 6.91 ± 2.13* |
| EEM area, mm2 | 11.89 ± 3.84 | 12.78 ± 3.90* |
| Plaque-media area, mm2 | 5.59 ± 2.93 | 5.87 ± 2.69 |
Data are given as mean ± SD or percentage
EEM external elastic membrane
*p < 0.001 (compared with the prestent measurements);
†p < 0.01 (compared with the prestent measurements)
IVUS stent length measurements and manufacturer stent length (manufacturer-stated length and foreshortened length)
| Measurements, mm | |
|---|---|
| E-E length* | 18.76 (15.65–23.60) |
| A-A length† | 18.36 (15.19–23.47) |
| M-L length | 18.0 (15.0–23.0) |
| Calculated foreshortened stent length | 17.84 (14.59–22.79) |
Data are given as median (IQR)
E-E length edge-to-edge stent length, A-A length area-to-area stent length, M-L length manufacturer-stated length
*p < 0.0001 is for E-E length versus M-L length, calculated foreshortened stent length and A-A length
†p < 0.0001 is for A-A length versus M-L length and calculated foreshortened stent length
Fig. 1Comparison between IVUS-measured stent length and manufacturer stent length according to the stent diameter. E-E length = edge-to-edge stent length by IVUS, M-L length = manufacturer-stated length
Predictors of difference in stent length
| Lesion length | 0.01 |
| Prestenting minimal EEM diameter | 0.03 |
| Lesions present in the LAD vessel | 0.03 |
| Manufacturer-stated length (M-L) | 0.04 |
EEM external elastic membrane, LAD left anterior descending vessel