Intravascular ultrasonic assessment of stent diameters derived from manufacturer's compliance charts.

Jose de Ribamar Costa1, Gary S Mintz, Stéphane G Carlier, Ricardo A Costa, Kenichi Fujii, Koichi Sano, Masashi Kimura, Joanna Lui, Giora Weisz, Issam Moussa, George Dangas, Roxana Mehran, Alexandra J Lansky, Edward M Kreps, Michael Collins, Gregg W Stone, Jeffrey W Moses, Martin B Leon.   

Abstract

We used intravascular ultrasound (IVUS) to assess the accuracy of manufacturers' stent balloon compliance charts. Many interventional cardiologists rely on manufacturers' compliance charts to select stent size and optimize stent diameters according to inflation pressures during percutaneous procedures. We randomly selected 212 patients who had de novo coronary lesions that had been treated with a single, bare metal, > or =3.0-mm stent (Bx velocity, NIR, TETRA/PENTA, S660/S670/S7) under IVUS guidance. Cases of stent overlap and postdilatation with another balloon were excluded. Predicted stent diameters were derived from each manufacturer's compliance charts, and stent size and final maximal deployment pressures were derived from each physician's report. IVUS-measured minimum stent diameters (range 1.4 to 4.0 mm, mean 2.79 +/- 0.48) were smaller than predicted diameters (range 3.1 to 4.57 mm, mean 3.79 +/- 0.44). The ratio of IVUS to predicted diameters ranged from 44% to 97% (mean 74 +/- 10%). This finding was common to all 3 stent sizes: 74 +/- 12% for 3.0 mm, 73 +/- 9% for 3.5 mm, and 74 +/- 9% for 4.0-mm stents (p = 0.9). This finding was also common to all 4 stent manufacturers, 72 +/- 8% for Boston Scientific, 76 +/- 11% for Guidant, 73 +/- 9% for Cordis, and 74 +/- 11% for Medtronic (p = 0.13), and to different stent lengths. Only 3.8% of the stents achieved 90% of the predicted minimum stent diameters, and only 24.6% achieved 80% of the predicted minimum stent diameters. In conclusion, in human coronary arteries, minimal stent diameter measured by IVUS is significantly smaller than that predicted by in vitro compliance charts. These differences are independent of stent manufacturer, length, diameter, and deployment pressure.

Entities:  

Mesh:

Year:  2005        PMID: 15979438     DOI: 10.1016/j.amjcard.2005.02.049

Source DB:  PubMed          Journal:  Am J Cardiol        ISSN: 0002-9149            Impact factor:   2.778


  11 in total

1.  Finite Element Analysis of the Implantation Process of Overlapping Stents.

Authors:  Jiang Xu; Jie Yang; Salman Sohrabi; Yihua Zhou; Yaling Liu
Journal:  J Med Device       Date:  2017-05-03       Impact factor: 0.582

2.  Compliance charts to guide non-complex small artery stenting: validation by quantitative coronary angiography.

Authors:  Yohan Chacko; J Kimberly Haladyn; Debbie M Smith; Richard Lim
Journal:  Heart Asia       Date:  2013-05-31

3.  Expert Opinion: Optimising Stent Deployment in Contemporary Practice: The Role of Intracoronary Imaging and Non-compliant Balloons.

Authors:  Ashok Seth; Sajal Gupta; Vivudh Pratap Singh; Vijay Kumar
Journal:  Interv Cardiol       Date:  2017-09

4.  Two-in-one aortic valve sizing and valvuloplasty conductance balloon catheter.

Authors:  Mark C Svendsen; Anjan K Sinha; Zachary C Berwick; William Combs; Shawn D Teague; Thierry Lefevre; Vasilis Babaliaros; Ghassan Kassab
Journal:  Catheter Cardiovasc Interv       Date:  2015-01-30       Impact factor: 2.692

5.  Polymer Coating Integrity, Thrombogenicity and Computational Fluid Dynamics Analysis of Provisional Stenting Technique in the Left Main Bifurcation Setting: Insights from an In-Vitro Model.

Authors:  Marek Milewski; Chen Koon Jaryl Ng; Pawel Gąsior; Shaoliang Shawn Lian; Su Xiao Qian; Shengjie Lu; Nicolas Foin; Elvin Kedhi; Wojciech Wojakowski; Hui Ying Ang
Journal:  Polymers (Basel)       Date:  2022-04-22       Impact factor: 4.967

6.  Longitudinal stent elongation or shortening after deployment in the coronary arteries: which is dominant?

Authors:  Magdy Algowhary; Mohammed Aboel-Kassem F Abdelmegid
Journal:  Egypt Heart J       Date:  2021-05-17

7.  Effect of stent inflation pressure and post-dilatation on the outcome of coronary artery intervention. A report of more than 90,000 stent implantations.

Authors:  Ole Fröbert; Giovanna Sarno; Stefan K James; Nawsad Saleh; Bo Lagerqvist
Journal:  PLoS One       Date:  2013-02-13       Impact factor: 3.240

8.  Expansion of the Multi-Link Frontier™ coronary bifurcation stent: micro-computed tomographic assessment in human autopsy and porcine heart samples.

Authors:  Stefan Kralev; Benjamin Haag; Jens Spannenberger; Siegfried Lang; Marc A Brockmann; Soenke Bartling; Alexander Marx; Karl-Konstantin Haase; Martin Borggrefe; Tim Süselbeck
Journal:  PLoS One       Date:  2011-07-21       Impact factor: 3.240

9.  In vivo measurement of stent length by using intravascular ultrasound.

Authors:  Magdy Algowhary; Salma Taha; Hosam Hasan-Ali; Akihiko Matsumura
Journal:  Egypt Heart J       Date:  2019-12-19

10.  Efficacy of non-compliant balloon post-dilation in optimization of contemporary stents: A digital stent enhancement study.

Authors:  Jaya Chandrasekhar; Christopher Allada; Simon O'Connor; Moyazur Rahman; Bruce Shadbolt; Ahmad Farshid
Journal:  Int J Cardiol Heart Vessel       Date:  2014-03-19
View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.