Influence of length, diameter and position of the implant in its fracture incidence: A Systematic Review.

Marcelo Coelho Goiato1, Agda Marobo Andreotti2, Daniela Micheline Dos Santos3, Adhara Smith Nobrega2, Fernanda Pereira de Caxias2, Lisiane Cristina Bannwart3.   

Abstract

Background. Implant fractures can cause difficult problems for patients and dentists. This systematic review aimed to determine the influence of some implant parameters on the occurrence of their fracture and to determine the incidence of fractures reported in recent years. Methods. A search was conducted in Pubmed database, from which 12 studies published in the last 12 years were selected. Results. This review reported a 2% incidence of implant fracture. Most implants had been in function between 3 and 4 years until fracture. The studies did not provide necessary information to establish a relationship between the different parameters of implants and the incidence of fractures. Conclusion. Thus, the indication of type, diameter and length of an implant and the bone quality in the region receiving it should be studied and accurately examined for each individual case in order to avoid future failures.
© 2019 Coelho Goiato et al.

Entities:  

Keywords:  Dental implants; dental restoration failure; mouth rehabilitation; periprosthetic fractures.

Year:  2019        PMID: 31592306      PMCID: PMC6773918          DOI: 10.15171/joddd.2019.017

Source DB:  PubMed          Journal:  J Dent Res Dent Clin Dent Prospects        ISSN: 2008-210X


Introduction

Since the beginning of implantology, implant-supported prostheses have been a highly predictable treatment for total or partial edentulous patients.[1,2] However, complications affecting implant osseointegration can occur in specific situations, and professionals should be aware of treatment limitations to avoid risky situations that could lead to implant and prosthesis failures.[2,3] Implantologyis constantly seeking improvements in and development of new materials and implants. Titanium implants have been used for decades and the implants fabricated with zirconia were developed as an alternative to some problems caused by titanium.[4] Complications related to dental implants can be classified as biological or mechanical. Although mechanical complications are rare, they can lead to serious clinical consequences.[5] These complications might involve loosening or fracture of the prosthetic screw, loosening or fracture of the abutment screw, and also implant fracture.[2] Recent studies have investigated the rates, pattern and potential risk of implant fractures.[6-8] Lee et al found lower risk of fractures in implants with wide‐diameter microthreads, placed in patients with a history of bone graft, as well as those positioned in the mandibular anterior area.[6] Karl, Scherg and Grobecker-Karl reported that torque might be a risk factor for fracture in zirconia implants.[8] However, the literature reports a low incidence of implant fractures, but when it occurs, it can cause difficult problems for patients and dentists. Determining the etiology of fractures can be a challenge for dentists.[9] A great number of factors should be considered when analyzing the possible causes for dental implant fractures. One of the main causes is the biomechanical overloading that can occur due to parafunctional activities such as bruxism, malocclusion, presence of cantilevers and lack of passive fit of implant-supported prostheses, resulting in fatigue.[10-12] Also, the implant location, insufficient quantity of implants supporting the prosthesis, implant material, implant diameter and other factors must also be considered.[13] The treatment of a fractured implant can be a challenge for the clinician due to surgical, rehabilitative and emotional implications.[12] Many factors can influence implant fractures and the complexity of a failure on this level, regarding both the unsuccessful treatment and the failure resolution. This systematic review aimed to evaluate the effect of some implant parameters, such as length, diameter and position, on the occurrence of fracture and to determine the incidence of fractures reported in recent years.

Methods

Search Strategy and Screening of Articles

An electronic survey was conducted in Pubmed database using the filters: Species (Humans) and Languages (English). Two examiners using the following term “dental implant fracture” conducted the search process independently. The search period was the interval from 07/31/2004 to 03/02/2016. Study selection was initially directed to title and abstract analysis. Given the existence of few randomized controlled studies, it included prospective and retrospective studies. Subsequently, the eligible studies were analyzed and included or excluded from the total sample. Thus, the population, intervention, comparison and outcome (PICO), as recommended by PRISMA,[13] were determined as questioning criteria to organize a clear clinical question and an appropriate inclusion approach where the "Population" corresponds to patients rehabilitated with dental implants. An "Intervention" is the occurrence of fractures and the different characteristics of fractured implants. The "Comparison" corresponds to not fractured implants. Finally, the "Outcome" was the influence of the implant characteristics on fracture incidence. Inclusion criteria were as follows: English, prospective, retrospective randomized controlled clinical trials; the study should report at least one case of implant fracture, even if their aim was not to evaluate the fracture incidence. Exclusion criteria consisted of duplicated, animal, cadaver and in vitro studies, isolated case reports, reviews, and studies that reported no cases of implant fractures. The following information was collected from selected articles: number, gender and age of patients evaluated; number, material, diameter, length and installation region of both installed and fractured implants, and the loading period until implant fracture.

Results

A total of 632 studies were found with the search terms. By reviewing the titles and abstracts of each study, 15 were selected, of which one was excluded because it was not available for download at our institution and 2 were excluded after reading their full texts, for not fitting the inclusion criteria. Thus, 12 studies were selected to carry out this systematic review (Figure 1).
Figure 1
PRISMA flowchart. From the selected studies, 4 were prospective and 8 were retrospective. The number of patients evaluated was 594, with 354 women and 240 men, aged 17‒94 years, considering that one of the included studies did not provide such information.[14] The follow-up of cases ranged from 1 to 20 years (Table 1).
Table 1

Demographic data of studies selected

Study Type of study Follow-up Number of patients evaluated Female Male Age Range (Average Age)
Gargallo Albiol et al, 2008 RetrospectiveMinimum of 5 yearsNRNRNRNR
Antoun et al, 2012 Retrospective17.6 months44321251 a 94 - (70)
Cha et al, 2013 Retrospective5 years120635718.8 a 81.1 - (47)
Eccellente et al, 2011 Prospective26.7 months45182743 a 76 - (60)
Eliasson et al, 2010 Prospective5 years291316NR - (65)
Gahlert et al, 2013 RetrospectiveUp to 3 years79433619 a 85 - (52,2)
Grant et al, 2009 Retrospective2 years and 1 month124893518 a 80 - (56)
Lekholm et al, 2006 Retrospective20 years178943 a 87 - (68)
Misje et al, 2013 Retrospective12 to 15 years1811717 a 41 - (NR)
Šćepanović et al, 2012 Prospective1 year30161445 a 63 - (NR)
Wahlstrom et al, 2010 RetrospectiveAverage of 5 years (40 to 84 months)46331336 a 84 - (59)
Preoteasa et al, 2014 Prospective3 years23131052 a 76 – (62)
Total 1 to 20 years59435424017 a 94

NR: Not reported

NR: Not reported Tables 2 and 3 show data regarding the installed implants and implants fractured for each study selected: number of implants, material, diameter, length, region (maxillary/mandibular) and loading time until fracture. It can be observed that not all studies provided full information, which was designated with "NR" (Not Reported).
Table 2

Number, material, diameter and length of implants installed and fractured

Study Implants installed Implants fractured Implant material Diameter of implants installed (number of implants) Diameter of implants fractured (number of implants) Length of implants installed (number of implants) Length of implants fractured (number of implants)
Gargallo Albiol et al, 2008 150021TitaniumNR 3.75 mm (20);4 mm (1) NR10 to 15 mm
Antoun et al, 2012(2) 205 1 Titanium 3.35 mm (2)3.75 mm (15)4 mm (185)5 mm (3) NR 10 mm (3);11.5 mm (13);13 mm (86);15 mm (103) NR
Cha et al, 2013 13611Titanium4 mm (136)4 mm (11)11/13 mm (124)NR
Eccellente et al, 2011 180 1 Titanium 3.5 mm (153);4.5 mm (27) 3.5 mm (1) 9.5 mm (3);11 mm (18);14 mm (92);17 mm (1) 17 mm (1)
Eliasson et al, 2010 168 3 Titanium 3.75 mm (168) 3.75 mm (3) 10 mm (4);13 mm (33);16 mm (131) NR
Gahlert et al, 2013 17013Zirconia 3.25 mm (59);4 mm (82);5 mm (29) 3.25 mm (12);4 mm (1) 10 to 13 mm NR
Grant et al, 2009 335 1 Titanium 3.5 mm (42);4.3 mm (212);5 mm (75);6 mm (6) NR 8 mm (335) 8 mm (1)
Lekholm et al, 2006 69 4 Titanium 4 mm (69) 4 mm (4) 7 mm (38%);10 mm (33%);13/15 mm (29%) NR
Misje et al, 2013 221Titanium 3.75 mm (21)3.25 (1) 3.75 mm (1) 15 mm (13);13 mm (8)10 mm (1) 10 mm (1)
Šćepanović et al, 2012 1233Titanium1.8 mm (123)1.8 mm (3)13 mm (123)13 mm (3)
Wahlstrom et al, 2010 1164TitaniumNRNRNRNR
Preoteasa et al, 2014 1102Titanium 1.8 mm (NR)2.1 mm (NR)2.4 mm (NR) NR 10 mm (NR)13 mm (NR)15 mm (NR)18 mm (NR) NR
Total 313465

NR: Not reported

Table 3

Region (maxillary/mandibular) of implants installed and fractured, and loading period until fracture occurred

Study Number of implants installed (Maxilla/Mandible) Number of fractures per region (Maxilla/Mandible) Loading period until fracture
Gargallo Albiol et al, 2008 NR (Maxilla)NR (Mandible) NR (Maxilla)NR (Mandible) 2 years (2)3-4 years (16)5-6 years (1)9 years (2)
Antoun et al, 2012(2) 78 (Maxilla)124 (Mandible) 0 (Maxilla)1 (Mandible) 21 months
Cha et al, 2013 70 (Maxilla)66 (Mandible) NR (Maxilla)NR (Mandible) 6 months (1);10 months (1);1 year (2);2 years (7)
Eccellente et al, 2011 180 (Maxilla)0 (Mandible) 1 (Maxilla)0 (Mandible) 6 months
Eliasson et al, 2010 0 (Maxilla)168 (Mandible) 0 (Maxilla)3 (Mandible) 3 years (1);4 years (2)
Gahlert et al, 2013 90 (Maxilla)80 (Mandible) NR (Maxilla)NR (Mandible) NR
Grant et al, 2009 0 (Maxilla)335 (Mandible) 0 (Maxilla)1 (Mandible) 10 months
Lekholm et al, 2006 NR (Maxilla)NR (Mandible) 2 (Maxilla)2 (Mandible) 2 years (1);7 years (1);17 years (2)
Misje et al, 2013 22 (Maxilla)0 (Mandible) 1 (Maxilla)0 (Mandible) 10 years
Šćepanović et al, 2012 0 (Maxilla)123 (Mandible) 0 (Maxilla)3 (Mandible) Not loaded.
Wahlstrom et al, 2010 116 (Maxilla)0 (Mandible) 4 (Maxilla)0 (Mandible) 6.5 years (3);3 years (1)
Preoteasa et al, 2014 36 (Maxilla)74 (Mandible) NR (Maxilla)NR (Mandible) 1 to 2 years
Total 760 (Maxilla)970 (Mandible)

NR: Not reported

NR: Not reported NR: Not reported Twelve studies reported 3134 implants installed, of which 94.6% were titanium and 5.4% were zirconia implants. Of the implants placed, 65 were lost by fracture, representing an incidence of 2.0%. Among the 65 fractured implants, 20% were zirconia and 80% were titanium implants (Table 2). Considering the studies that provided complete information about installation region, 760 implants were installed in the maxilla and 970 in the mandible. The implant loading period until fracture ranged from "not loaded" to 17 years (Table 3). Data about fractured implants were also evaluated considering the studies that provided information about both installed and fractured implants for the different variables evaluated, allowing a proper comparison (Tables 4 to 6).
Table 4

Fracture incidence for each diameter, according to the studies that provided information about installed and fractured diameters

Diameters installed Total number of implants Implants fractured (%)
1.81233 (2.4)
3.256012 (20)
3.51531 (0.6)
3.7518924 (12.7)
428717 (5.9)
4.5270 (0)
5290 (0)
Total86857
Table 6

Fracture incidence in maxilla and mandible, according to the studies that provided information about regions of installed and fractured implants

Region (Maxilla/Mandible) Installed Fractured
Maxilla396 (38.7%)6 (1.5%)
Mandible627 (61.3%)8 (0.8%)
Total102314
Diameter of Implants Installed vs. Diameter of Im-plants Fractured Seven studies provided information about diameters of both the installed and fractured implants (Table 2). Thus, 868 implants were reported with seven different diameters. The most commonly used diameter was 4 mm ‒ 287 implants, of which 17 fractured, corresponding to a fracture incidence of 5.9% (Table 4). The installation of implants with diameters of 4.5 and 5 mm was less frequent (27 and 29 implants, respectively), followed by implants 3.25 mm in diameter (60 implants). The first two diameters showed no fracture cases, while the 3.25-mm implants presented a higher incidence of fracture compared to the total implants of the same diameter and to fracture incidence in other diameters (Table 4).

Lengths of Implants Installed vs. Lengths of Implants Fractured

From the 12 selected studies, only 4 provided the lengths of both installed and fractured implants (Table 2). In these 4 studies, 594 implants presented 8 different lengths. The most common was the 8-mm-long implant (56,4%); however, only one 8-mm-long implant fractured, corresponding to 0.3% of fracture in relation to total implants of the same length (Table 5). Regarding implants 10 and 17 mm in length, one implant was installed in each category and both fractured, corresponding to 100% fracture of implants with such lengths (Table 5).
Table 5

Fracture incidence for each length, according to the studies that provided information about installed and fractured lengths

Lengths installed Implants installed Implants fractured
83351 (0.3%)
9.530
1011 (100%)
11180
131313 (2.3%)
14920
15130
1711 (100%)
Total5946

Regions of Installed and Fractured Implants

Six studies allowed comparison of regions of installation and fracture (Table 3). Most implants were installed in the mandible (61.3%) (Table 6). However, the highest incidence of fractures occurred in implants installed in the maxilla (1.5%) (Table 6). The implant loading period until fracture was reported by 11 studies (Table 3). Of these, 50 fractures were observed, with 38.5% occurring between 3 to 4 years of implants in function (Table 7).
Table 7

Loading period until fracture and number of implants fractured in each period

Loading period until fracture Implants fractured
Before loading 3
6 months 2
10 months 2
1–2 years 15
3–4 years 20
5–6 years 1
6.5 years 3
7 years 1
9–10 years 3
17 years 2
Total52

Discussion

Of all the mechanical complications, the implant fracture is considered the most frustrating and might occur after a certain period in function. Literature reports a large variation (0% to 3.45%) in the incidence of implant fractures,,[5,15] although the studies that have reported 0% of fracture incidence exhibit relatively small sample sizes and short periods of follow-up.[16,17] A 10-year follow-up study, with 1618 implants, reported an 0.8% fracture incidence, with only 13 implants fractured.[18] In contrast, Adell et al[19] evaluated 1997 Branemark implants and reported a relatively high fracture rate of 3.45% over a period of 15 years of follow-up. Balshi[9] published data from 4045 implants and reported a 0.2% fracture rate over 5 years. This systematic review reported a fracture incidence of 2% in the studies selected, a value that fits the range demonstrated in the literature. From the 65 fractured implants found in the studies selected for this review, 20% were zirconia (n=13) and 80% (n=52) were titanium implants. The reasons for titanium implants’ fracture are well described in the literature,[9,20-22] with overload identified as the major cause,[5,23] being first attributed to a progressive fatigue until implant loses the appropriate strength to maintain its integrity, culminating in a catastrophic failure.[3] Factors such as implant design, manufacturing defects and lack of a passive fit of prosthesis can also be associated with fractures.[20] The absence of periodontal ligament which is present in natural teeth and direct bone apposition to the implant do not allow its movement when subjected to occlusal loads and may result in excessive stress, leading to different mechanical failures, among them, the implant fracture.[24] Regarding zirconia implants, Gahlert et al[20] assessed 119 implants and reported 13 lost by fracture. Macroscopic and microscopic analyses of these fractures have shown that overload, implant design, and surface microcracks due to manufacturing process were also the main reasons for zirconia implants’ fracture.[25] Zirconia milling process might result in imperfections and microcracks[26,27] that can influence the fracture resistance and material reliability.[20,26-28] Regarding zirconia implants, Gahlert et al[20] assessed 119 implants and reported 13 lost by fracture. Macroscopic and microscopic analyses of these fractures have shown that overload, implant design, and surface microcracks due to manufacturing process were also the main reasons for zirconia implants’ fracture.[25] Zirconia milling process might result in imperfections and microcracks[26,27] that can influence the fracture resistance and material reliability.[20,26-28]

Diameter

In this systematic review, the highest incidence of fractures in terms of diameter occurred in implants 3.25 mm in diameter (20%) (Table 4), which are considered small-diameter implants. In a recent systematic review, Klein et al[29] assessed the success of narrow-diameter implants. The authors showed survival rates between 90.9% and 100% for implants <3 mm in diameter, while for diameters between 3.0 and 3.25 mm, survival rates ranged from 93.8% to 100%.[30,31] It has been shown that a reduced implant diameter might reduce the osseointegration surface and compromise the mechanical conditions in the implant body, abutment and screw components.[32,33] Moreover, narrow-diameter implant assemblies are more prone to mechanical failure because of their compromised fatigue, with the effects of magnitude of force and angulation being of greater clinical significance when narrow-diameter implants are used.[5,33] Although narrow-diameter implants show greater propensity to failures, a study by Šćepanović et al[34] showed that 1.8-mm-diameter implants exhibited a low fracture incidence (2.4%) (Tables 2 and 4), which could be justified by the fact that most of the implants evaluated were splinted to receive the prosthesis. On the other hand, it is observed that in this review, 12 fractured implants with 3.25 mm belong to the study of Gahlert et al,[20] (Table 2), which, in addition to being zirconia implants, were installed to support one-unit prostheses. A study showed better sharing of occlusal loads and distribution of stress with splinted versus individually restored implant designs.[35] Literature describes that narrow-diameter implants should be used with specific indications.[30] These implants are offered by almost all implant manufacturers and are designed specifically for restricted interdental spaces, mandibular incisors and maxillary lateral teeth.[32,36] In contrast to what is described, a long survival period might be expected for reduced diameter implants, as long as the number of implants used is sufficient to support well planned prostheses.

Length

The literature is heterogeneous about the survival rates regarding implant length, as well as the definition of short implants, which varies from 4 to 11 mm, with some studies considering standard-length implants as implants with lengths of ≥10 mm.[37] Winkler et al[38] demonstrated a survival rate of 66.7% for short implants (7 mm) and 96.4% for long implants (16 mm) after 6 months in situ.[1] Another recent study detected similar survival rates of short and long implants.[39] Annibali et al[40] published a systematic review including clinical studies of short implants (<10 mm) placed in the maxilla and mandible (6193 implants in 3848 participants) and reported an overall cumulative survival rate of 99.1%, while Srinivasan et al[37] reported survival rates ranging from 92.2% to 100% for implants <8 mm in length. In this review, short implants (8 mm) showed only 0.3% of fractures compared to the total number of implants (Table 5), while longer implants (10 and 17 mm) presented 100% of fracture. Although many studies have demonstrated higher failure rates for short implants, recent reports show survival rates of these implants similar to longer implants.[41] In 2006, Misch et al[41] published a literature review of failure rates associated with dental implants <10 mm in length in the posterior regions of partially edentulous patients undergoing placement from 1991 to 2003. They reported that among 2837 short implants, survival rate was 85.3%. Furthermore, they and other authors have shown that failures are independent of implant length, with no clear linear relationship between the length and implant survival.[40,42] It should be noted that data extracted from studies on this topic are limited and inconclusive. Since, in this systematic review, only one implant in both the 10- and 17-mm lengths was found, this is not sufficient to draw conclusions regarding the success/failure of long implants. Still, few studies provided data about the length of implants installed and fractured. Only 4 of the selected studies provided this information, where from 594 implants placed, we have data from only 6 of the fractured implants (Table 5).

Installation Region (Maxilla/Mandible)

In this systematic review, failures were interestingly observed more frequently in the maxilla than in the mandible (1.5% to 0.8%, respectively) (Table 6), similar to a study by Srinivasan.[38] In a long-term multicenter study, Adell et al[19] reported a cumulative implant fracture rate of less than 5% in both the maxilla and mandible in a period of 10 to 15 days, except for one study group where fracture rate reached 16% in the maxilla. It can be explained by the fact that shape and bone density are important factors for implants survival.[33] The mandible is a cortical bone, while the maxillary bone is trabecular and less mineralized, which might compromise primary implant stability and lead to future failures.[32] A systematic review reported an implant fracture rate of 0.5% after 5 years (Jung et al, 2008).[15] In the study of Adell et al,[19] an incidence rate of 1.0‒3.5% for implant fracture was observed, with most fractures also occurring after 5 years of clinical function. However, in this systematic review, the largest number of fractures was reported in a range of 3‒4 years (20 fractures, Table 7), which can be attributed to the fact that from the 12 selected studies, 6 underwent follow-ups shorter than 5 years (Table 1). The studies selected exhibited a broad diversity in terms of implant length and diameter, location of installation, study design and observation period. Furthermore, the studies showed variations related to unspecified dropouts, specific time of fracture and method of statistical analysis. These factors deemed it impossible to systematically compare the reviewed publications with one another; which was a similar finding in an earlier published review. Hence, a descriptive, but nevertheless structured and methodologically solid analysis was performed in this review.[37] As could be observed, some studies in the literature do not provide necessary information to establish a relationship between the different parameters of implants and the incidence of fractures. However, from the above, we consider that although the incidence of implant fractures is relatively low, this failure can be avoided taking into consideration the different implant characteristics. The indication of type, diameter and length of an implant and bone quality in the region that will receive it should be studied and evaluated precisely for each specific case.

Authors’ Contributions

AMA and ASN were responsible for data collection and interpretation, preparation of the English manuscript and data interpretation. DMS and MCG were responsible for the project and interpretation and provided advice and revised important intellectual content. FPC was responsible for data interpretation, revision of important intellectual contente and final review.

Funding

This study had no funding source.

Conflict of Interests

The authors declare no conflict(s) of interest related to the publication of this work.
  41 in total

1.  A prospective clinical study evaluating the safety and effectiveness of narrow-diameter threaded implants in the anterior region of the maxilla.

Authors:  E Andersen; E Saxegaard; B M Knutsen; H R Haanaes
Journal:  Int J Oral Maxillofac Implants       Date:  2001 Mar-Apr       Impact factor: 2.804

2.  Failures and complications in 127 consecutively placed fixed partial prostheses supported by Brånemark implants: from prosthetic treatment to first annual checkup.

Authors:  T Jemt; B Lindén; U Lekholm
Journal:  Int J Oral Maxillofac Implants       Date:  1992       Impact factor: 2.804

3.  Fatigue and damage tolerance of Y-TZP ceramics in layered biomechanical systems.

Authors:  Yu Zhang; Antonia Pajares; Brian R Lawn
Journal:  J Biomed Mater Res B Appl Biomater       Date:  2004-10-15       Impact factor: 3.368

4.  The longitudinal clinical effectiveness of osseointegrated dental implants: the Toronto study. Part III: Problems and complications encountered.

Authors:  G A Zarb; A Schmitt
Journal:  J Prosthet Dent       Date:  1990-08       Impact factor: 3.426

5.  Effects of implant diameter, insertion depth, and loading angle on stress/strain fields in implant/jawbone systems: finite element analysis.

Authors:  Lihe Qian; Mitsugu Todo; Yasuyuki Matsushita; Kiyoshi Koyano
Journal:  Int J Oral Maxillofac Implants       Date:  2009 Sep-Oct       Impact factor: 2.804

6.  Implant fracture failure rate and potential associated risk indicators: An up to 12-year retrospective study of implants in 5,124 patients.

Authors:  Dong-Woon Lee; Na-Hong Kim; Young Lee; Yeon-Ah Oh; Jae-Hong Lee; Hyung-Keun You
Journal:  Clin Oral Implants Res       Date:  2019-02-18       Impact factor: 5.977

7.  Analysis of incidence and associated factors with fractured implants: a retrospective study.

Authors:  S E Eckert; S J Meraw; E Cal; R K Ow
Journal:  Int J Oral Maxillofac Implants       Date:  2000 Sep-Oct       Impact factor: 2.804

8.  Failure analysis of fractured dental zirconia implants.

Authors:  M Gahlert; D Burtscher; I Grunert; H Kniha; E Steinhauser
Journal:  Clin Oral Implants Res       Date:  2011-05-05       Impact factor: 5.977

9.  Incidence and pattern of implant fractures: A long-term follow-up multicenter study.

Authors:  Jae-Hong Lee; Yeon-Tae Kim; Seong-Nyum Jeong; Na-Hong Kim; Dong-Woon Lee
Journal:  Clin Implant Dent Relat Res       Date:  2018-05-15       Impact factor: 3.932