Failure analysis of fractured dental zirconia implants.

M Gahlert1, D Burtscher, I Grunert, H Kniha, E Steinhauser.   

Abstract

OBJECTIVES: The purpose of the present study was the macroscopic and microscopic failure analysis of fractured zirconia dental implants.
METHODS: Thirteen fractured one-piece zirconia implants (Z-Look3) out of 170 inserted implants with an average in situ period of 36.75±5.34 months (range from 20 to 56 months, median 38 months) were prepared for macroscopic and microscopic (scanning electron microscopy [SEM]) failure analysis. These 170 implants were inserted in 79 patients. The patient histories were compared with fracture incidences to identify the reasons for the failure of the implants.
RESULTS: Twelve of these fractured implants had a diameter of 3.25 mm and one implant had a diameter of 4 mm. All fractured implants were located in the anterior side of the maxilla and mandibula. The patient with the fracture of the 4 mm diameter implant was adversely affected by strong bruxism. By failure analysis (SEM), it could be demonstrated that in all cases, mechanical overloading caused the fracture of the implants. Inhomogeneities and internal defects of the ceramic material could be excluded, but notches and scratches due to sandblasting of the surface led to local stress concentrations that led to the mentioned mechanical overloading by bending loads.
CONCLUSIONS: The present study identified a fracture rate of nearly 10% within a follow-up period of 36.75 months after prosthetic loading. Ninety-two per cent of the fractured implants were so-called diameter reduced implants (diameter 3.25 mm). These diameter reduced implants cannot be recommended for further clinical use. Improvement of the ceramic material and modification of the implant geometry has to be carried out to reduce the failure rate of small-sized ceramic implants. Nevertheless, due to the lack of appropriate laboratory testing, only clinical studies will demonstrate clearly whether and how far the failure rate can be reduced.
© 2011 John Wiley & Sons A/S.

Entities:  

Mesh:

Substances:

Year:  2011        PMID: 21545535     DOI: 10.1111/j.1600-0501.2011.02206.x

Source DB:  PubMed          Journal:  Clin Oral Implants Res        ISSN: 0905-7161            Impact factor:   5.977


  9 in total

1.  The osteoinduction of RGD and Mg ion functionalized bioactive zirconia coating.

Authors:  Zhengfei Huang; Zhifeng Wang; Chuanhua Li; Ning Zhou; Fei Liu; Jing Lan
Journal:  J Mater Sci Mater Med       Date:  2019-08-14       Impact factor: 3.896

2.  The Effects of Syndecan on Osteoblastic Cell Adhesion Onto Nano-Zirconia Surface.

Authors:  Lu Sun; Guang Hong; Hiroyuki Matsui; Yun-Jia Song; Keiichi Sasaki
Journal:  Int J Nanomedicine       Date:  2020-07-14

3.  The Influence of Surface Preparation, Chewing Simulation, and Thermal Cycling on the Phase Composition of Dental Zirconia.

Authors:  Markus Wertz; Florian Fuchs; Hieronymus Hoelzig; Julia Maria Wertz; Gert Kloess; Sebastian Hahnel; Martin Rosentritt; Andreas Koenig
Journal:  Materials (Basel)       Date:  2021-04-22       Impact factor: 3.623

4.  Clinical management of implant prostheses in patients with bruxism.

Authors:  Osamu Komiyama; Frank Lobbezoo; Antoon De Laat; Takashi Iida; Tsuyoshi Kitagawa; Hiroshi Murakami; Takao Kato; Misao Kawara
Journal:  Int J Biomater       Date:  2012-06-04

5.  An in vitro evaluation of the biological effects of carbon nanotube-coated dental zirconia.

Authors:  Wen Kou; Tsukasa Akasaka; Fumio Watari; Göran Sjögren
Journal:  ISRN Dent       Date:  2013-08-20

6.  Biomechanical and histological evaluation of the osseointegration capacity of two types of zirconia implant.

Authors:  Jian-Min Han; Guang Hong; Hong Lin; Yoshinaka Shimizu; Yuhan Wu; Gang Zheng; Hongyu Zhang; Keiichi Sasaki
Journal:  Int J Nanomedicine       Date:  2016-12-07

7.  A Novel Technique for the Connection of Ceramic and Titanium Implant Components Using Glass Solder Bonding.

Authors:  Enrico Mick; Joachim Tinschert; Aurica Mitrovic; Rainer Bader
Journal:  Materials (Basel)       Date:  2015-07-14       Impact factor: 3.623

Review 8.  A Critical Review of Dental Implant Materials with an Emphasis on Titanium versus Zirconia.

Authors:  Reham B Osman; Michael V Swain
Journal:  Materials (Basel)       Date:  2015-03-05       Impact factor: 3.623

9.  Influence of length, diameter and position of the implant in its fracture incidence: A Systematic Review.

Authors:  Marcelo Coelho Goiato; Agda Marobo Andreotti; Daniela Micheline Dos Santos; Adhara Smith Nobrega; Fernanda Pereira de Caxias; Lisiane Cristina Bannwart
Journal:  J Dent Res Dent Clin Dent Prospects       Date:  2019-08-14
  9 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.