Literature DB >> 30672029

Implant fracture failure rate and potential associated risk indicators: An up to 12-year retrospective study of implants in 5,124 patients.

Dong-Woon Lee1, Na-Hong Kim2, Young Lee3, Yeon-Ah Oh1, Jae-Hong Lee4, Hyung-Keun You5.   

Abstract

OBJECTIVES: This study investigated fracture rates and risk indicators for fractures in internal connection dental implants.
MATERIAL AND METHODS: We performed a retrospective analysis of 19,006 internal connection implants used in fixed restoration in 5,124 patients (4,570 males, 554 females) at the Dental Hospital of Veterans Health Service Medical Center between 2006 and 2015. Patients were followed through June 2018 (0.03-12.39 years post-installation). Clinical factors (age, sex, implant diameter, implant length, placement site, bone graft, fixture material, cervical feature, abutment connection, microthread, and platform switching) were recorded. Kaplan-Meier survival analysis identified risk indicators associated with an implant fracture. Cox regression models elucidated potential fracture risks.
RESULTS: One hundred and seventy-four implants fractured in 135 patients, for an incidence rate of 0.92% after an average of 4.95 ± 2.14 years of use. Kaplan-Meier estimates showed that the 3-, 5-, and 10-year survival rates of implants were 99.8%, 99.2%, and 97.7%, respectively. In the multivariable Cox regression model, the diameter, location, history of bone graft, and microthread presence were significantly correlated with implant fractures. Wide-diameter implants had a reduced fracture risk within 90 months, after which the diameter did not correlate with fractures. Implants placed in the anterior mandible had a lower fracture risk within 90 months; mandibular premolar implants corresponded with a lower risk after 90 months. Implants without a history of bone graft or microthreads were more likely to fracture throughout the follow-up time.
CONCLUSIONS: These results elucidate risk indicators for implant fractures and facilitate their reduction in clinical practice.
© 2019 John Wiley & Sons A/S. Published by John Wiley & Sons Ltd.

Entities:  

Keywords:  cervical feature; dental implant; diameter; implant fracture; late implant failure

Mesh:

Substances:

Year:  2019        PMID: 30672029     DOI: 10.1111/clr.13407

Source DB:  PubMed          Journal:  Clin Oral Implants Res        ISSN: 0905-7161            Impact factor:   5.977


  5 in total

1.  Osteotomy combined with the trephine technique for invisible implant fracture: A case report.

Authors:  Liang-Wen Chen; Min Wang; Hai-Bin Xia; Dong Chen
Journal:  World J Clin Cases       Date:  2022-06-06       Impact factor: 1.534

2.  Influence of length, diameter and position of the implant in its fracture incidence: A Systematic Review.

Authors:  Marcelo Coelho Goiato; Agda Marobo Andreotti; Daniela Micheline Dos Santos; Adhara Smith Nobrega; Fernanda Pereira de Caxias; Lisiane Cristina Bannwart
Journal:  J Dent Res Dent Clin Dent Prospects       Date:  2019-08-14

3.  Fractures of implant fixtures: a retroscccpective clinical study.

Authors:  Han-Chang Yu; Young-Kyun Kim
Journal:  Maxillofac Plast Reconstr Surg       Date:  2020-04-25

4.  Artificial Intelligence in Fractured Dental Implant Detection and Classification: Evaluation Using Dataset from Two Dental Hospitals.

Authors:  Dong-Woon Lee; Sung-Yong Kim; Seong-Nyum Jeong; Jae-Hong Lee
Journal:  Diagnostics (Basel)       Date:  2021-02-03

5.  Deep learning improves implant classification by dental professionals: a multi-center evaluation of accuracy and efficiency.

Authors:  Jae-Hong Lee; Young-Taek Kim; Jong-Bin Lee; Seong-Nyum Jeong
Journal:  J Periodontal Implant Sci       Date:  2022-06       Impact factor: 2.086