| Literature DB >> 31409385 |
Ting Wan1, Houfa Yin1, Yi Yang1, Fang Wu1, Zhiyi Wu1, Yabo Yang2.
Abstract
BACKGROUND: To compare and correlate anterior segment measurements of myopic eyes implanted with Implantable Collamer lens (ICL V4c) by using anterior segment optical coherence tomography (AS-OCT), Pentacam and ultrasound biomicroscopy (UBM).Entities:
Keywords: Anterior chamber depth; Implantable Collamer lens; Optical coherence tomography; Pentacam; Ultrasound biomicroscopy; Vault
Mesh:
Year: 2019 PMID: 31409385 PMCID: PMC6693247 DOI: 10.1186/s12886-019-1194-y
Source DB: PubMed Journal: BMC Ophthalmol ISSN: 1471-2415 Impact factor: 2.209
Fig. 1Anterior segment measured by AS-OCT (a), Pentacam (b) and UBM (c) in phakic eyes after ICL implantation. ACD: Anterior chamber depth; C-ICL: Distance between the corneal endothelium and anterior surface of ICL; Vault: Distance between the back surface of ICL and the front surface of the lens
Repeatability of Pentacam, UBM and AS-OCT in measuring ACD, C-ICL and Vault
| Device | Parameter | Pearson (r) | Difference(mm) | t-value | ||
|---|---|---|---|---|---|---|
| AS-OCT | ACD | 0.98 | 0.000 | 0 ± 0.04 | −1.20 | 0.23 |
| C-ICL | 0.99 | 0.000 | 0 ± 0.04 | −0.48 | 0.64 | |
| Vault | 0.99 | 0.000 | 0 ± 0.03 | 1.44 | 0.15 | |
| Pentacam | ACD | 0.98 | 0.000 | 0 ± 0.04 | −0.85 | 0.40 |
| C-ICL | 0.99 | 0.000 | 0. ± 0.04 | −0.90 | 0.37 | |
| Vault | 0.99 | 0.000 | 0 ± 0.03 | −0.77 | 0.45 | |
| UBM | ACD | 0.98 | 0.000 | 0 ± 0.05 | 0.36 | 0.72 |
| C-ICL | 0.99 | 0.000 | 0 ± 0.05 | 0.83 | 0.41 | |
| Vault | 0.99 | 0.000 | 0 ± 0.04 | 0.22 | 0.83 |
AS-OCT: Anterior segment optical coherence tomography; UBM: Ultrasound biomicroscopy; Pearson (r): Pearson correlation coefficient; ACD: Anterior chamber depth; C-ICL: Distance between the corneal endothelium and anterior surface of ICL; Vault: Distance between the back surface of ICL and the front surface of the lens. P values less than 0.05 were considered statistically significant
Measurements of anterior section parameters using AS-OCT, Pentacam and UBM
| Parameters(Mean ± SD) | AS-OCT(mm) | Pentacam(mm) | UBM(mm) |
|---|---|---|---|
| ACD | 3.18 ± 0.23 | 2.93 ± 0.22 | 3.04 ± 0.22 |
| C-ICL | 2.32 ± 0.28 | 2.22 ± 0.27 | 2.24 ± 0.29 |
| Vault | 0.64 ± 0.25 | 0.43 ± 0.24 | 0.56 ± 0.27 |
Comparison of anterior section parameters measurements using AS-OCT, Pentacam and UBM
| Parameter | AS-OCT vs Pentacam | AS-OCT vs UBM | Pentacam vs UBM | ||||||
|---|---|---|---|---|---|---|---|---|---|
| ACD | C-ICL | Vault | ACD | C-ICL | Vault | ACD | C-ICL | Vault | |
| t | 35.3 | 10.64 | 27.67 | 14.37 | 5.47 | 7.81 | −10.87 | −2.19 | −12.67 |
|
| 0.000 | 0.000 | 0.000 | 0.000 | 0.000 | 0.000 | 0.000 | 0.031 | 0.000 |
Paired t-tests were used to compare measurements. P values less than 0.05 were considered statistically significant
Results of ICC and Pearson correlation coefficient (r) for each parameter between different devices
| Parameter | AS-OCT vs Pentacam | AS-OCT vs UBM | Pentacam vs UBM | ||||||
|---|---|---|---|---|---|---|---|---|---|
| ACD | C-ICL | Vault | ACD | C-ICL | Vault | ACD | C-ICL | Vault | |
| ICC | 0.98 | 0.98 | 0.98 | 0.96 | 0.95 | 0.97 | 0.96 | 0.96 | 0.97 |
| 0.000 | 0.000 | 0.000 | 0.000 | 0.000 | 0.000 | 0.000 | 0.000 | 0.000 | |
| Pearson (r) | 0.96 | 0.95 | 0.96 | 0.92 | 0.91 | 0.94 | 0.91 | 0.92 | 0.94 |
| 0.000 | 0.000 | 0.000 | 0.000 | 0.000 | 0.000 | 0.000 | 0.000 | 0.000 | |
ICC: Intraclass correlation coefficient. P values less than 0.05 were considered statistically significant
Fig. 2Bland and Altman plots comparing the level of agreement between the 3 instruments for ACD (a), C-ICL (b), Vault (c) measurements. The vertical axis represents the difference between these measurements and the horizontal axis shows their mean. Dashed lines represent the 95% confidence intervals. Solid line represents the mean difference. All scales in mm. ACD: Anterior chamber depth; C-ICL: Distance between the corneal endothelium and anterior surface of ICL; Vault: Distance between the back surface of ICL and the front surface of the lens
Agreement of parameters measured by AS-OCT, Pentacam and UBM
| Parameter(mm) | AS-OCT vs Pentacam | AS-OCT vs UBM | Pentacam vs UBM |
|---|---|---|---|
| ACD | 0.25 (0.13, 0.38) | 0.15 (−0.03, 0.33) | −0.11 (− 0.29, 0.07) |
| C-ICL | 0.10 (− 0.07, 0.27) | 0.07 (− 0.16, 0.31) | −0.03 (− 0.25, 0.20) |
| vault | 0.21 (0.08, 0.34) | 0.08 (− 0.10, 0.26) | −0.13 (− 0.31, 0.05) |
Mean (95%): mean value of measurement difference between 2 methods (95% limits of agreement)